Skip to content

Olympus A22251ARecalls & reported problems

OLYMPUS Winter & Ibe GmbH · HF-resection electrode, roller · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04042761021562), received between Aug 14, 2018 and Apr 18, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Fracture2
  • Material Fragmentation2
  • Mechanical Problem2
  • Material Twisted/Bent1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 18, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED, THE HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30° DISPLAYED A BENT ELECTRODE TIP. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR A TRANSCERVICAL RESECTION PROCEDURE. THE PROCEDURE WAS COMPLETED USING THE SAME SET OF EQUIPMENT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19134392 ↗
  • MalfunctionOct 19, 2022Break; Fracture; Material Fragmentation; Mechanical Problem; Break; Fracture; Material Fragmentation; Mechanical Problem

    OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION (TUR) PROCEDURE THE BALL PART OF THE ROLLER ELECTRODE WAS MISSING WHEN THE ELECTRODE WAS PULLED OUT THROUGH THE URETHRA. THE STAFF COULD NOT FOUND THE MISSING PART AND ASSUMES THAT THE ROLLER MAY HAVE BEEN ASPIRATED ALONG WITH THE TISSUE. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED USING A SIMILAR DEVICE AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.

    Patient problems: Foreign Body In Patient

    MAUDE report 15630844 ↗
  • MalfunctionAug 14, 2018Break; Fracture; Material Fragmentation; Mechanical Problem; Break; Fracture; Material Fragmentation; Mechanical Problem

    OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE, THE ROLLER AT THE DISTAL END OF THE HF RESECTION ELECTRODE DETACHED AND FELL INSIDE THE PATIENT¿S BLADDER. HOWEVER, NO FRAGMENT REMAINED INSIDE THE PATIENT SINCE IT WAS REPORTEDLY RETRIEVED. NO FURTHER INFORMATION WAS PROVIDED BUT THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7780434 ↗

Other Hysteroscope (And Accessories) devices

All Hysteroscope (And Accessories) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.