olympus A42021ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Outer sheath, 8.5 mm / 26 Fr., 2 stopcocks, rotatable · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04042761023665), received between Jul 12, 2023 and Aug 3, 2026 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Break2
- Loose or Intermittent Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Ecchymosis1
- Hemorrhage/Blood Loss/Bleeding1
- Laceration(s)1
- Rupture1
- Unspecified Gastrointestinal Problem1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 3, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50123113 ↗MalfunctionDec 15, 2025Break; Break
IT WAS REPORTED THE OUTER SHEATH EXHIBITED A DAMAGED CERAMIC TIP. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23800236 ↗MalfunctionAug 21, 2024Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT WHILE USING THE SHEATH WITH THE HIGH FREQUENCY UNIT, THE HIGH FREQUENCY UNIT EXHIBITED AN ERROR (INSTRUMENT CANNOT BE RECOGNIZED. UPON CHECKING THE PLASMA ELECTRIC CUTTING INTERFACE, IT WAS FOUND THAT THE INTERFACE WAS LOOSE AND SUSPECTED THAT POOR INTERFACE CONTACT HAD CAUSED THE MALFUNCTION. THE ISSUE OCCURRED DURING AN UNKNOWN THERAPEUTIC SURGERY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE WITHOUT DELAY. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20017247 ↗InjuryJul 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT AN ELECTRICAL FAILURE OCCURRED AND NO IMAGE DURING THE HYSTEROSCOPIC MYOMECTOMY AND THE SAME PROBLEM OCCURRED AFTER REPLACING ALL THE COMBINED DEVICES CAUSING A 30-MINUTE DELAY IN SURGERY AND 30 MINUTES EXTENDED ANESTHESIA TIME. THERE WERE NO ERROR MESSAGES. THE PROCEDURE WAS COMPLETED USING OTHER UROLOGICAL PRODUCTS. THE EVENT DID NOT AFFECT THE DIAGNOSTIC OR TREATMENT OUTCOME. THERE WERE NO REPORTS OF PATIENT HARM. THIS EVENT REQUIRES 12 REPORTS. RELATED PATIENT IDENTIFIERS FOR THE FIRST COMBINATION OF DEVICES USED ARE AS FOLLOWS: (B)(6) (WA00014A / HF-CABLE, BIPOLAR), (B)(6) (WA00014A / HF-CABLE, BIPOLAR) , (B)(6) (A42011A / RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS), (B)(6) (A42021A / OUTER SHEATH, 8.5 MM / 26 FR., 2 STOPCOCKS, ROTATABLE), (B)(6) (WA22366A /WORKING ELEMENT, ACTIVE, FOR RESECTION IN SALINE), (B)(6) (WB50402W / FOOT SWITCH, DOUBLE, FOR, ESG-400). RELATED PATIENT IDENTIFIERS FOR THE SECOND COMBINATION OF DEVICES USED: (B)(6) (WB50402W / FOOT SWITCH, DOUBLE, FOR, ESG-400), (B)(6) (WA22366A / WORKING ELEMENT, ACTIVE, FOR RESECTION IN SALINE), (B)(6) (A42011A / RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17372159 ↗InjuryJul 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT AN ELECTRICAL FAILURE OCCURRED AND NO IMAGE DURING THE HYSTEROSCOPIC MYOMECTOMY AND THE SAME PROBLEM OCCURRED AFTER REPLACING ALL THE COMBINED DEVICES CAUSING A 30-MINUTE DELAY IN SURGERY AND 30 MINUTES EXTENDED ANESTHESIA TIME. THERE WERE NO ERROR MESSAGES. THE PROCEDURE WAS COMPLETED USING OTHER UROLOGICAL PRODUCTS. THE EVENT DID NOT AFFECT THE DIAGNOSTIC OR TREATMENT OUTCOME. THERE WERE NO REPORTS OF PATIENT HARM. THIS EVENT REQUIRES 12 REPORTS. RELATED PATIENT IDENTIFIERS FOR THE FIRST COMBINATION OF DEVICES USED ARE AS FOLLOWS: (B)(6) (WA00014A / HF-CABLE, BIPOLAR) (B)(6) (WA00014A / HF-CABLE, BIPOLAR) (B)(6) (A42011A / RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS) (B)(6) (A42021A / OUTER SHEATH, 8.5 MM / 26 FR., 2 STOPCOCKS, ROTATABLE) (B)(6) (WA22366A /WORKING ELEMENT, ACTIVE, FOR RESECTION IN SALINE) (B)(6) (WB50402W / FOOT SWITCH, DOUBLE, FOR, ESG-400). RELATED PATIENT IDENTIFIERS FOR THE SECOND COMBINATION OF DEVICES USED: (B)(6) (WB50402W / FOOT SWITCH, DOUBLE, FOR, ESG-400) (B)(6) (WA22366A / WORKING ELEMENT, ACTIVE, FOR RESECTION IN SALINE) (B)(6) (A42011A / RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17371127 ↗InjuryJul 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED THAT AT THE END OF ENDOSCOPIC SURGERY FOR TRANSURETHRAL RESECTION FOR PROSTATIC ADENOMA (WEIGHING AROUND 50 GM) A VIOLENT INTRAVESICAL EXPLOSION OF A PROBABLE GASEOUS NATURE OCCURRED DURING THE USE OF CLOT CURRENT. THE PROCEDURE WAS IMMEDIATELY STOPPED, AND EXPLORATORY LAPAROTOMY WAS REQUIRED, REVEALED ENDO PERITONEAL LACERATION OF THE BLADDER. REPORTEDLY, THE BLADDER WAS COMPLETELY RUPTURED WITH BLEEDING AND INTRAVESICAL DISPLACEMENT OF INTESTINAL LOOPS WITH SOME ECCHYMOSIS. AS SUCH, COMPLEX EXPLORATORY SURGERY WAS PERFORMED TO REPAIR THE BLADDER. POST SURGERY, THE PATIENT WAS TRANSFERRED TO THE INTENSIVE CARE UNIT FOR FURTHER CARE. UPON FOLLOW-UP NO ADDITIONAL INFORMATION COULD BE RECEIVED REGARDING THE PATIENT'S CURRENT CONDITION. THERE WAS 6 DEVICES INVOLVED IN THIS EVENT REPORTED UNDER THE FOLLOWING PATIENT IDENTIFIERS: RELATED PATIENT IDENTIFIER # (B)(6), HF UNIT "ESG-400", MODEL- A42021A, SERIAL # (B)(6). RELATED PATIENT IDENTIFIER # (B)(6), OUTER SHEATH, 8.5 MM / 26 FR., 2 STOPCOCKS, ROTATABLE, MODEL- WA2T430A, SERIAL #(B)(6) (THIS REPORT). RELATED PATIENT IDENTIFIER # (B)(6), TELESCOPE "OES ELITE", 4 MM, 30°, HD, AUTOCLAVABLE, MODEL- WA2T430A SERIAL #
Patient problems: Ecchymosis; Hemorrhage/Blood Loss/Bleeding; Laceration(s); Rupture; Unspecified Gastrointestinal Problem
MAUDE report 17313109 ↗
FDA clearance and classification
The FDA regulates product code HIH (hysteroscope (and accessories)) under 21 CFR 884.1690 as a Class II device. 21 CFR 884.1690 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the olympus A42021A
Has the olympus A42021A been recalled?
No FDA recall record in our data names this exact device identifier (DI 04042761023665), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the olympus A42021A?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jul 12, 2023 and Aug 3, 2026. In those reports, the most common specific device problems are “Break” (2) and “Loose or Intermittent Connection” (1); the most common patient problems are “Ecchymosis” (1), “Hemorrhage/Blood Loss/Bleeding” (1) and “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the olympus A42021A?
None of the reports are classified as deaths, 3 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the olympus A42021A, and is it still sold?
The FDA's device database lists OLYMPUS Winter & Ibe GmbH as the company responsible for it (brand olympus, model A42021A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Hysteroscope (And Accessories) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)452 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- VISERA HYSTEROVIDEOSCOPE OLYMPUS HYF TYPE VOlympus Medical Systems Corp. · HYSTEROSCOPE (AND ACCESSORIES)1 recall48 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page