Skip to content

Olympus WA58365SRecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Trocar spike, 11 x 60 mm, blunt · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04042761030779), received between Jan 26, 2021 and Oct 14, 2024 — classified as 7 injuries.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K923982); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation3
  • Detachment of Device or Device Component2
  • Material Separation2
  • Break1
  • Fracture1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Code Available1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A THORACOSCOPY, THE FIRST THORACOSCOPY PAN THE PHYSICIAN OPENED WAS MISSING THE 10 MM METAL OBTURATOR FOR THE METAL TROCAR. A SECOND PAN WAS OPENED. UPON USING THE 10 MM TROCAR, WHEN THE PHYSICIAN WENT TO MOVE THE OBTURATOR, ONLY THE PLASTIC TOP OF THE OBTURATOR CAME OUT. THE METAL END OF THE OBTURATOR FELL THROUGH THE TROCAR INTO THE PATIENT¿S CHEST. IT TOOK THE PHYSICIAN 45-60 MINUTES TO FIND AND REMOVE THE FOREIGN METAL OBJECT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20446216 ↗
  • InjuryOct 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A THORACOSCOPY, THE FIRST THORACOSCOPY PAN THE PHYSICIAN OPENED WAS MISSING THE 10 MM METAL OBTURATOR FOR THE METAL TROCAR. A SECOND PAN WAS OPENED. UPON USING THE 10 MM TROCAR, WHEN THE PHYSICIAN WENT TO MOVE THE OBTURATOR, ONLY THE PLASTIC TOP OF THE OBTURATOR CAME OUT. THE METAL END OF THE OBTURATOR FELL THROUGH THE TROCAR INTO THE PATIENT¿S CHEST. IT TOOK THE PHYSICIAN 45-60 MINUTES TO FIND AND REMOVE THE FOREIGN METAL OBJECT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20445849 ↗
  • InjuryMay 26, 2023Material Separation; Material Separation

    THE CUSTOMER REPORTED TO OLYMPUS THAT DURING A THERAPEUTIC THORACIC SURGERY, THIS LAPAROSCOPIC TROCAR BLADE CAME APART AND FELL INTO THE PATIENT'S CHEST CAVITY. THE OBTURATOR WAS RETRIEVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17012351 ↗
  • InjuryMay 26, 2023Material Separation; Material Separation

    THE CUSTOMER REPORTED TO OLYMPUS THAT DURING A THERAPEUTIC THORACIC SURGERY, THIS LAPAROSCOPIC TROCAR BLADE CAME APART AND FELL INTO THE PATIENT'S CHEST CAVITY. THE OBTURATOR WAS RETRIEVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17012071 ↗
  • InjuryMay 8, 2023Material Fragmentation; Material Fragmentation

    OLYMPUS RECEIVED A MEDWATCH REPORT (2301650000-2023-8005) THROUGH THE MAIL WHICH STATING "WHILE SCRUBBED INTO DOCTOR'S VIDEO-ASSISTED THORACIC SURGERY (VATS) CASE AND I HANDED HIM AN 11MM TROCAR OBTURATOR. AFTER HE INSERTED IT, HE ONLY HANDED ME BACK A PIECE OF PLASTIC, INSTEAD OF THE OBTURATOR. THE PLASTIC WAS PART OF THE OBTURATOR. THE METAL PORTION OF THE OBTURATOR WAS NOT RETURNED TO ME, I INFORMED HIM THAT I DID NOT GET IT. ON FURTHER EXAMINATION, WE FOUND THAT THE OBTURATOR CAME APART. THEY FEEL INTO THE PATIENT'S CHEST CAVITY. THE OBTURATOR WAS RETRIEVED FROM THE PATIENT. THE ENTIRE SILVER PORTION DROPPED INTO THE BODY CAVITY; ONLY THE BLACK HANDLE WAS GIVEN TO THE SCRUB TECH."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16885765 ↗
  • InjuryMay 5, 2023Material Fragmentation; Material Fragmentation

    OLYMPUS RECEIVED A MEDWATCH REPORT (B)(4) THROUGH THE MAIL WHICH STATING "WHILE SCRUBBED INTO DOCTOR'S VIDEO-ASSISTED THORACIC SURGERY (VATS) CASE AND I HANDED HIM AN 11MM TROCAR OBTURATOR. AFTER HE INSERTED IT, HE ONLY HANDED ME BACK A PIECE OF PLASTIC, INSTEAD OF THE OBTURATOR. THE PLASTIC WAS PART OF THE OBTURATOR. THE METAL PORTION OF THE OBTURATOR WAS NOT RETURNED TO ME, I INFORMED HIM THAT I DID NOT GET IT. ON FURTHER EXAMINATION, WE FOUND THAT THE OBTURATOR CAME APART. THEY FEEL INTO THE PATIENT'S CHEST CAVITY. THE OBTURATOR WAS RETRIEVED FROM THE PATIENT. THE ENTIRE SILVER PORTION DROPPED INTO THE BODY CAVITY; ONLY THE BLACK HANDLE WAS GIVEN TO THE SCRUB TECH."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16879835 ↗
  • InjuryJan 26, 2021Break; Fracture; Material Fragmentation; Mechanical Problem; Break; Fracture; Material Fragmentation; Mechanical Problem

    OLYMPUS WAS INFORMED THAT DURING AN UNSPECIFIED THORACOSCOPY PROCEDURE, WHEN INSERTING THE TROCAR, THE MANDREL OF THE TROCAR SPIKE BROKE AND THE ROD SLID ALONG THE TROCAR SLEEVE TO PROTRUDE INTO THE THORAX. THE SURGEON MANAGED TO RECOVER THE METAL ROD BUT THE OPERATION/ANESTHESIA WAS PROLONGED AS A RESULT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Code Available

    MAUDE report 11228549 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K923982. One submission often covers many models, so check the product and lot details in each record.

  • Z-1163-2010 · Feb 8, 2010 · Terminated
    Olympus Trocar Spikes, Model numbers: A5821, A5948, A5949, WA58341L, and WA58343L Used in conjunction with a compatible trocar tube, trocar tubes are used for intraperitoneal acce

Other Laparoscope, General & Plastic Surgery devices

All Laparoscope, General & Plastic Surgery devices →