Skip to content

Olympus A22203CRecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jun 8, 2023Class IIOpen, Classified

    Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip

    Z-2413-2023 · Root cause: Process design · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code GCP (endoscope, ac-powered and accessories) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Olympus A22203C

Has the Olympus A22203C been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2413-2023 (Class II), began on Jun 8, 2023: Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Who makes the Olympus A22203C, and is it still sold?

The FDA's device database lists OLYMPUS Winter & Ibe GmbH as the company responsible for it (brand Olympus, model A22203C). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Endoscope, Ac-Powered And Accessories devices

All Endoscope, Ac-Powered And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page