Olympus WA50023BRecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus rigid video laparoscope is used for surgery in the abdomen. The 44 reports mostly describe image problems — poor or distorted pictures, no image at all, or an upside-down view — along with cracks, breaks and dents in the scope body found during inspection. None of the reports involve a death or an injury. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
44
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 04042761066501), received between Jan 20, 2024 and May 6, 2026 — classified as 44 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 5 reportsJul 2024: 1 reportsAug 2024: 3 reportsSep 2024: 3 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 3 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image17
  • No Display/Image8
  • Crack6
  • Break5
  • Dent in Material5
  • Communication or Transmission Problem2
  • Failure to Clean Adequately2
  • Image Orientation Incorrect2

Patient problems

  • No Clinical Signs, Symptoms or Conditions44

Recent adverse event reports

The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2026Poor Quality Image

    IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE HAD A CRACKED IMAGE. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25095228 ↗
  • MalfunctionApr 30, 2026Poor Quality Image; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE EXHIBITED DEBRIS UNDER OBJECTIVE LENS, DUST REFLECTION IN IMAGE AND DISTORTED IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25050276 ↗
  • MalfunctionMar 31, 2026Communication or Transmission Problem

    IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE HAD ERROR CODE E311 DISPLAYED. THE ISSUE WAS FOUND DURING SEDATION - DEVICE OUTSIDE PATIENT - TO A THERAPEUTIC LAPAROSCOPIC APPENDECTOMY PROCEDURE THAT WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24743786 ↗
  • MalfunctionMar 5, 2026Poor Quality Image

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE RIGID VIDEO LAPAROSCOPE IMAGE WAS UPSIDE DOWN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24519703 ↗
  • MalfunctionMar 2, 2026Poor Quality Image; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE EXHIBITED DIRT IN OBJECTIVE LENS AND IMAGE OUT OF FIELD OF VIEW. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24490525 ↗
  • MalfunctionFeb 27, 2026Degraded; Dent in Material; Gas/Air Leak; Physical Resistance/Sticking

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE HAD A HANDLE LEAK, A STUCK HANDLE RUBBER BAND, DENTS ON THE DISTAL EDGE, AND DAMAGED EDGES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24469669 ↗
  • MalfunctionFeb 13, 2026Communication or Transmission Problem; Optical Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24331995 ↗
  • MalfunctionJan 9, 2026Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24021347 ↗
  • MalfunctionDec 26, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THE RIGID VIDEO LAPAROSCOPE EXHIBITED A DARK IMAGE AND POOR VIEW DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23898794 ↗
  • MalfunctionDec 8, 2025Poor Quality Image; Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FORM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23750806 ↗
  • MalfunctionDec 4, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23717615 ↗
  • MalfunctionOct 21, 2025Crack; Crack

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23351133 ↗
  • MalfunctionOct 16, 2025Leak/Splash; Leak/Splash

    NO OTHER UPDATED INFORMATION FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310553 ↗
  • MalfunctionJun 24, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD IMAGE PROBLEM. THE EVENT WAS REPORTED DURING SET UP / INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22323343 ↗
  • MalfunctionMay 23, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22081061 ↗
  • MalfunctionApr 25, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THE SUBJECT DEVICE HAD AN IMAGE NOT CLEAR DURING SETUP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21912131 ↗
  • MalfunctionApr 24, 2025Poor Quality Image; Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21905046 ↗
  • MalfunctionMar 31, 2025Crack; Crack

    IT WAS REPORTED THE LENS OF SUBJECT DEVICE WAS CRACKED NEAR ADAPTOR END. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21729217 ↗
  • MalfunctionMar 26, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21697135 ↗
  • MalfunctionMar 26, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD BUBBLES COMING FROM CORD. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21696023 ↗
  • MalfunctionFeb 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE LIGHT GUIDE TIP OF THE SUBJECT DEVICE CAME OFF. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21437032 ↗
  • MalfunctionFeb 12, 2025No Display/Image; Poor Quality Image; No Display/Image; Poor Quality Image

    IT WAS REPORTED THAT THE RIGID VIDEO SCOPE HAD POOR IMAGE QUALITY AND THAT THERE WAS A BLACKOUT. THE ISSUE WAS FOUND DURING SET UP IN THE ROOM/ INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21368281 ↗
  • MalfunctionJan 23, 2025Break; Poor Quality Image; Break; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD POOR IMAGE DUE TO BROKEN DISTAL END LENS. THE EVENT OCCURRED DURING SET UP/ INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21223149 ↗
  • MalfunctionJan 14, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD NO PICTURE. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21155113 ↗
  • MalfunctionNov 27, 2024Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO TELESCOPE EXHIBITED DISTAL FIBER BREAKAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20791465 ↗

Other Laparoscope, General & Plastic Surgery devices

All Laparoscope, General & Plastic Surgery devices →