Olympus WA50021BRecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus rigid video laparoscope is a surgical scope used in minimally invasive operations. The reports mostly describe damage found during cleaning or servicing: cracked, scratched or dented lens tips, broken fiber bundles, bent outer tubes, and images that come out blurry or blank, including one where the camera broke. All 54 reports are malfunctions with no patient involvement, and no deaths or injuries were reported. There are no recalls naming this device, and the report count stayed the same over the last12 months at 15.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 04042761066815), received between Apr 17, 2024 and Jun 9, 2026 — classified as 54 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 5 reportsAug 2024: 4 reportsSep 2024: 6 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 3 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 5 reportsApr 2026: 4 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image19
  • Dent in Material12
  • Break10
  • No Display/Image10
  • Scratched Material7
  • Crack6
  • Display or Visual Feedback Problem4
  • Material Twisted/Bent3

Patient problems

  • No Clinical Signs, Symptoms or Conditions54

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2026Break; Poor Quality Image; Dent in Material; Scratched Material

    THE CUSTOMER RETURNED THE RIGID VIDEO LAPAROSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN IDENTIFIED THAT THE DEVICE EXHIBITED DISTAL FIBER BREAKAGE, DENTS AND SCRATCHES ON THE OBJECTIVE FRAME AT THE DISTAL END, AND AN IMAGE THAT WAS OUT OF FIELD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25448456 ↗
  • MalfunctionMay 8, 2026Poor Quality Image; Scratched Material; Material Split, Cut or Torn

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE HAD SCRATCHES ON OBJECTIVE LENS, HOLE ON LIGHT GUIDE CABLE AND DARK IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25128143 ↗
  • MalfunctionMay 5, 2026No Display/Image

    IT WAS REPORTED THAT THE RIGID VIDEO SCOPE EXHIBITS NO IMAGE DURING INSPECTION FOR USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25083547 ↗
  • MalfunctionApr 11, 2026Dent in Material

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24852572 ↗
  • MalfunctionApr 7, 2026Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24801065 ↗
  • MalfunctionApr 3, 2026Break

    IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE HAD A BROKEN CAMARA. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24776607 ↗
  • MalfunctionApr 1, 2026Break; Dent in Material; Material Twisted/Bent

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE PRESENTED BREAKAGE OF DISTAL FIBERS, DENTS ON THE DISTAL EDGE AND BENT ON THE OUTER TUBE PROXIMAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24753919 ↗
  • MalfunctionMar 24, 2026Break; No Display/Image; Dent in Material; Material Twisted/Bent; Scratched Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE HAD FIBER BREAKAGE, DENTS ON THE DISTAL END, A BENT OUTER TUBE, SCRATCHES ON THE DISTAL END AND NO IMAGE WAS DISPLAYED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24674170 ↗
  • MalfunctionMar 23, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE HAD DENTED DISTAL EDGE OF OUTER TUBE AND BROKEN DISTAL FIBER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24661979 ↗
  • MalfunctionMar 13, 2026Connection Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24590924 ↗
  • MalfunctionMar 10, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THE RIGID VIDEO LAPAROSCOPE HAD DIM LIGHT. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24560573 ↗
  • MalfunctionMar 6, 2026No Display/Image

    IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE HAD NO IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24534287 ↗
  • MalfunctionFeb 26, 2026Poor Quality Image; Dent in Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE HAD DENT ON EDGE OF OBJECTIVE FRAME AND HORIZENTAL FLICKERING LINES ON IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24454008 ↗
  • MalfunctionFeb 10, 2026No Display/Image; Poor Quality Image

    IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE HAD NO SCREEN, HAD BLUE SCREEN AND GREEN LINES DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24306225 ↗
  • MalfunctionJan 14, 2026Break; Crack; No Display/Image; Poor Quality Image; Dent in Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE HAD AN OUTER TUBE BENT, FIBER BREAKAGE, BIG DENT ON DISTAL EDGE, NO IMAGE AND IMAGE ONLY SHOW STATIC. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24067577 ↗
  • MalfunctionOct 30, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEOSCOPE EXHIBITED DISTAL END FIBERS BREAKAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23430027 ↗
  • MalfunctionOct 15, 2025Poor Quality Image; Poor Quality Image

    NO OTHER UPDATED INFORMATION FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23292032 ↗
  • MalfunctionSep 16, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD IMAGE ISSUE BLUE TENT. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23063868 ↗
  • MalfunctionSep 14, 2025Break; Break

    IT WAS REPORTED THAT THE SUBJECT DEVICE'S LIGHT GRAY BUTTON KEEPS POPPING OFF. THE EVENT HAD OCCURRED DURING THE SET UP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23047871 ↗
  • MalfunctionSep 8, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD NO IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22994128 ↗
  • MalfunctionAug 13, 2025Dent in Material; Dent in Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE PRODUCT EXHIBITED DENTS ON DISTAL EDGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22793742 ↗
  • MalfunctionAug 12, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THE SUBJECT DEVICE HAD LEAK IN CORD. THE EVENT OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22790591 ↗
  • MalfunctionJul 24, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22594296 ↗
  • MalfunctionMay 15, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THE SUBJECT DEVICE HAD A BUNCH OF BUBBLES IN THE LENS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22026095 ↗
  • MalfunctionMay 5, 2025Crack; Crack

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21959883 ↗

Other Laparoscope, General & Plastic Surgery devices

All Laparoscope, General & Plastic Surgery devices →