Skip to content

Olympus WAIR100ARecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Telescope IR 10 mm, 0° · Class 2

Adverse event reports
12
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 04042761083621), received between Jul 20, 2022 and Mar 24, 2026 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break8
  • Appropriate Device Problem Term/Code Not Available1
  • Detachment of Device or Device Component1
  • Device Markings/Labelling Problem1
  • Fluid/Blood Leak1
  • Loose or Intermittent Connection1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID ENDOSCOPE TELESCOPE HAD A DETACHED DISTAL TERMINAL LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24673300 ↗
  • MalfunctionFeb 3, 2026Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID ENDOSCOPE TELESCOPE EXHIBITED A SHIFTED TUBE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246294 ↗
  • MalfunctionAug 22, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION THAT THE SUBJECT DEVICE HAS ITS COVER GLASS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22876547 ↗
  • MalfunctionOct 14, 2024Break; Break

    IT WAS REPORTED THE OPTICS IN THE RIGID ENDOSCOPE TELESCOPE HAD FALLEN AND THE BLACK PIECE HAD BROKEN. IT IS UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20437772 ↗
  • MalfunctionOct 7, 2024Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    IT WAS REPORTED THE TELESCOPE HAD LENS BREAKAGE AND LIQUID ENTRY DURING STERILIZATION. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20391701 ↗
  • MalfunctionJul 1, 2024Break; Break

    IT WAS REPORTED THAT THE DISTAL TIP GLASS OF THE TELESCOPE IS SHATTERED. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19652716 ↗
  • MalfunctionFeb 29, 2024Loose or Intermittent Connection; Loose or Intermittent Connection

    THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR AN UNKNOWN LAPAROSCOPIC PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18809069 ↗
  • MalfunctionDec 11, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, THE OLYMPUS RIGID SCOPE HAD A DAMAGED TIP. THE ISSUE WAS ASSOCIATED WITH A THERAPEUTIC PROCEDURE THAT WAS COMPLETED WITH A SIMILAR DEVICE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE TIP COVER GLASS WAS DAMAGED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18307485 ↗
  • MalfunctionOct 30, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE LIGHT GUIDE CAP OF THE SCOPE WAS DAMAGED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18035644 ↗
  • MalfunctionFeb 16, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS THAT DURING REPROCESSING, THE TELESCOPE HAD EXCESSIVE DAMAGE AND DENTS IN THE SHAFT. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16390958 ↗
  • MalfunctionJan 24, 2023Break; Break

    OLYMPUS (OMSC) WAS INFORMED THE CUSTOMER IR, AUTOCLAVABLE RIGID TELESCOPE WAS RETURNED TO THE OLYMPUS REPAIR CENTER FOR A REPORTED ¿OBJECTIVE LENS IS DOWN¿. THE CUSTOMER REPORTED EVENT WAS FOUND AT AN UNSPECIFIED EVENT. NO DEATH OR INJURY AND NO IMPACT TO PATIENT OR OTHER HAS BEEN REPORTED TO OLYMPUS. UPON INSPECTING, THE MISALIGNMENT OF THE OBJECTIVE LENS AT THE DISTAL END OF THE SCOPE ATTRIBUTING TO A CROPPED IMAGE. THIS REPORT IS BEING SUBMITTED TO CAPTURE THE MISALIGNMENT OF THE OBJECTIVE LENS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16231400 ↗
  • InjuryJul 20, 2022Device Markings/Labelling Problem; Appropriate Device Problem Term/Code Not Available; Device Markings/Labelling Problem; Appropriate Device Problem Term/Code Not Available

    OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC PROCEDURE FOR DETECTION OF SENTINEL LYMPH NODES, AN INCORRECT DEVICE WAS PROVIDED IN THE INSTRUMENT KIT (A TELESCOPE "ULTRA", 10 MM, 0° (WA4KL100, SN# (B)(4)) WAS IN THE BOX OF THE WAIR100A KIT (TELESCOPE "IR", 10 MM, 0°)) WITH THE EFFECT THAT THE INTENDED INFRARED FLUORESCENCE EXAMINATION COULD NOT BE PERFORMED. INSTEAD, A MORE INVASIVE PROCEDURE WITH A DOUBLE CURAGE LEFT AND RIGHT WAS PERFORMED. HOWEVER, THE PATIENT IS REPORTEDLY WELL.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15063242 ↗

Other Laparoscope, General & Plastic Surgery devices

All Laparoscope, General & Plastic Surgery devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.