ALPHAMAXX mobile operating table 113322B4Recalls & reported problems
MAQUET GmbH · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046768040007), received between Mar 3, 2023 and Mar 7, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 29, 2024Open, Classified
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.
Z-2271-2024 · Root cause: Component design/selection · FDA record ↗
Most reported problems
Device problems
- Positioning Failure1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 7, 2024Positioning Failure; Positioning Failure
ON 28TH FEBRUARY 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES - 113322B4 - ALPHAMAXX (460 MM LONGIT. SHIFT), EU. AS IT WAS STATED, A SURGEON COULD NOT CONTROL THE TABLE DURING ORTHOPEDIC CASE WITH X-RAY C-ARM ON ANESTHETIZED PATIENT. ERROR LIGHT WAS SHOWN ON LOCK AND UNLOCK BUTTON AND TABLE COULD NOT BE ADJUSTED (HEIGHT, ZERO POSITION, LEFT TILT FUNCTIONS). THE SURGERY WAS DELAYED AND SURGEON HAD TO WAIT FOR 30 MINUTES TO TRANSFER THE PATIENT FROM ALPHAMAXX TO OTHER TABLE IN ANOTHER OPERATING ROOM. THE PROCEDURE WAS CONTINUED AND FINISHED IN DIFFERENT OPERATING ROOM. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY THE INABILITY TO POSITION THE PATIENT AS REQUIRED AND A DELAY RESULTING IN PROLONGED ANESTHESIA TIME, WERE TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18855409 ↗MalfunctionMar 3, 2023Use of Device Problem; Use of Device Problem
MANUFACTURER'S REFERENCE NUMBER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16476452 ↗
Other TABLE, OPERATING-ROOM, AC-POWERED devices
- ORT300 TableIMRIS Imaging, Inc. · Table, Operating-Room, Ac-Powered2 recalls21 reports
- ORT200 TableIMRIS Imaging, Inc. · Table, Operating-Room, Ac-Powered2 recalls12 reports
- MEERA mobile operating table 720001F2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- MEERA mobile operating table 720001B2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall19 reports
- MAGNUS operating table system 118001B2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall17 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.