ALPHAMAXX mobile operating tableRecalls & reported problems across all models
MAQUET GmbH · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
3 open recalls name at least one ALPHAMAXX mobile operating table model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 3
We found 3 models of ALPHAMAXX mobile operating table by MAQUET GmbH in the FDA's device registry (GUDID). Together they are named in 8 adverse event reports in the FDA's MAUDE database, received between Mar 3, 2023 and May 14, 2026 — classified as 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (3 reports).
3 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
ALPHAMAXX mobile operating table recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 29, 2024Open, Classified
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.
Z-2273-2024 · Root cause: Component design/selection · Recall details · FDA record ↗
- May 29, 2024Open, Classified
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.
Z-2272-2024 · Root cause: Component design/selection · Recall details · FDA record ↗
- May 29, 2024Open, Classified
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.
Z-2271-2024 · Root cause: Component design/selection · Recall details · FDA record ↗
Most reported ALPHAMAXX mobile operating table problems
Device problems
- Positioning Failure3
- Use of Device Problem3
- Circuit Failure2
- Failure to Charge2
- Break1
- Failure to Run on Battery1
- Structural Problem1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Unspecified Tissue Injury2
ALPHAMAXX mobile operating table models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- ALPHAMAXX mobile operating table 113322B5TABLE, OPERATING-ROOM, AC-POWERED1 recall5 reports
- ALPHAMAXX mobile operating table 113322B4TABLE, OPERATING-ROOM, AC-POWERED1 recall2 reports
- ALPHAMAXX mobile operating table 113322F5TABLE, OPERATING-ROOM, AC-POWERED1 recall1 report
Recent ALPHAMAXX mobile operating table adverse event reports
The 8 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryMay 14, 2026Use of Device Problem
GETINGE BECAME AWARE OF AN INCIDENT INVOLVING ONE OF OUR OPERATING TABLES. DURING SURGERY, THE TABLE FELL DIRECTLY ONTO NURSE'S FOOT. THIS LED TO THE STAFF MEMBER SUSTAINING AN INJURY WHICH REQUIRED IMMEDIATE MEDICAL INTERVENTION. WE DECIDED TO REPORT THE ISSUE DUE TO THE SERIOUS INJURY OF THE USER. ACCORDING TO THE GETINGE TECHNICIAN WHO EVALUATED THE OR TABLE ON SITE, THERE WAS NO MALFUNCTION OF THE TABLE.
Patient problems: Unspecified Tissue Injury
MAUDE report 25174839 ↗InjuryMay 14, 2026Use of Device Problem
GETINGE BECAME AWARE OF AN INCIDENT INVOLVING ONE OF OUR OPERATING TABLES. DURING SURGERY, THE TABLE TABLE FELL DIRECTLY ONTO NURSE'S FOOT. THIS LED TO THE STAFF MEMBER SUSTAINING AN INJURY WHICH REQUIRED IMMEDIATE MEDICAL INTERVENTION. WE DECIDED TO REPORT THE ISSUE DUE TO THE SERIOUS INJURY OF THE USER. ACCORDING TO THE GETINGE TECHNICIAN WHO EVALUATED THE OR TABLE ON SITE, THERE WAS NO MALFUNCTION OF THE TABLE.
Patient problems: Unspecified Tissue Injury
MAUDE report 25171448 ↗MalfunctionMar 8, 2024Positioning Failure; Structural Problem; Positioning Failure; Structural Problem
ON 29TH FEBRUARY 2024, GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES - 113322F5 - ALPHAMAXX SURGICAL TABLE W/AUTO DRIVE. AS IT WAS STATED, THE TABLE WAS STUCK IN THE UPPERMOST POSITION DURING PROCEDURE AND WOULD NOT MOVE. ACCORDING TO THE CUSTOMER ALLEGATION, THE POSITION PUT THE PATIENT AT RISK OF FALLING OFF TABLE. NO ADDITIONAL DETAILS REGARDING THE PERFORMED PROCEDURE AND PATIENT POSITIONING HAVE BEEN RECEIVED TO DATE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY THE INABILITY TO MOVE THE TABLE RESULTING IN DELAY OR INTERRUPTION IN LIFE-SAVING SURGERY AND POTENTIALLY CREATING RISK OF PATIENT FALLING FROM THE TABLE, WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18864226 ↗MalfunctionMar 7, 2024Positioning Failure; Positioning Failure
ON 28TH FEBRUARY 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES - 113322B4 - ALPHAMAXX (460 MM LONGIT. SHIFT), EU. AS IT WAS STATED, A SURGEON COULD NOT CONTROL THE TABLE DURING ORTHOPEDIC CASE WITH X-RAY C-ARM ON ANESTHETIZED PATIENT. ERROR LIGHT WAS SHOWN ON LOCK AND UNLOCK BUTTON AND TABLE COULD NOT BE ADJUSTED (HEIGHT, ZERO POSITION, LEFT TILT FUNCTIONS). THE SURGERY WAS DELAYED AND SURGEON HAD TO WAIT FOR 30 MINUTES TO TRANSFER THE PATIENT FROM ALPHAMAXX TO OTHER TABLE IN ANOTHER OPERATING ROOM. THE PROCEDURE WAS CONTINUED AND FINISHED IN DIFFERENT OPERATING ROOM. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY THE INABILITY TO POSITION THE PATIENT AS REQUIRED AND A DELAY RESULTING IN PROLONGED ANESTHESIA TIME, WERE TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18855409 ↗MalfunctionJan 5, 2024Failure to Charge; Failure to Charge
ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES - (B)(4)- ALPHAMAXX MOBILE OPERATING TABLE. ACCORDING TO THE INITIALLY PROVIDED INFORMATION, THE PATIENT HAD TO BE TRANSFERRED TO ANOTHER TABLE DUE TO BATTERY CHARGER FAILURE. ON (B)(6) 2023, CLARIFICATION WAS RECEIVED STATING THAT WHEN THE MALFUNCTION OCCURRED THE PATIENT WAS NOT YET ON THE TABLE. HOWEVER, IT WAS REQUIRED TO EXCHANGE THE TABLE WITH A DEVICE FROM ANOTHER THEATRE. IN THAT OPERATING ROOM THE PATIENT WAS ALREADY ANESTHETIZED FOR THE PROCEDURE. AS IT WAS CONFIRMED, THE MALFUNCTION RESULTED IN A DELAY OF LISTS IN BOTH ROOMS. THE INCIDENT WAS INITIALLY ASSESSED AS REPORTABLE DUE TO A DELAY IN SURGERY RESULTING IN PROLONGED ANESTHESIA TIME. RECENTLY, THE INCIDENT HAS BEEN REEVALUATED AS ACCORDING TO THE MEDICAL EXPERT'S ASSESSMENT, THE PROCEDURAL DELAY DID NOT HAVE ANY MAJOR CLINICAL RELEVANCE IN THIS SITUATION. THE SCENARIO DESCRIBED IN THE RECORD IS CONSIDERED AS NON-REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18453172 ↗MalfunctionNov 20, 2023Failure to Charge; Circuit Failure; Positioning Failure; Failure to Run on Battery; Failure to Charge; Circuit Failure; Positioning Failure; Failure to Run on Battery
ON 8TH NOVEMBER 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES ¿ 113322B5 - ALPHAMAXX (460 MM LONGIT. SHIFT), EU. AS IT WAS STATED, THE TABLE COULD NOT BE MOVED AT ALL DURING HIP SURGERY ON THE ANESTHETIZED PATIENT EVEN WITH A USE OF THE OVERRIDE PANEL DUE TO A FAILURE OF CHARGER BOARD AND BATTERY. THE SURGERY WAS CONTINUED ON THE AFFECTED DEVICE, HOWEVER, THE SURGEON HAD TO BE SEATED FOR THE PROCEDURE AND THE STAFF STRUGGLED GETTING THE X-RAY MACHINE UNDER THE TABLE. NO DELAY IN SURGERY OCCURRED DUE TO THE ISSUE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY INABILITY TO MOVE THE TABLE DURING THE SURGERY DUE TO CHARGER BOARD MALFUNCTION AFFECTING PATIENT¿S POSITIONING, WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18169373 ↗MalfunctionOct 11, 2023Break; Circuit Failure; Unintended Movement; Break; Circuit Failure; Unintended Movement
ON 25TH SEPTEMBER 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES ¿ 113322B5 ALPHAMAXX (460 MM LONGIT. SHIFT), EU USED WITH HAND CONTROL. AS IT WAS STATED, UNINTENDED TILT MOVEMENT OF THE TABLE OCCURRED DURING SURGERY WHILE THE PATIENT WAS ON THE TABLE. DURING SERVICE VISIT, INTERNAL DAMAGE OF THE HAND CONTROL¿S CABLE HAS BEEN IDENTIFIED WHICH COULD HAVE CONTRIBUTED TO THE INCIDENT. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY INVOLUNTARY TILT MOVEMENT OF THE TABLE WHICH COULD POTENTIALLY LEAD TO A PATIENT'S FALL, WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17910378 ↗MalfunctionMar 3, 2023Use of Device Problem; Use of Device Problem
MANUFACTURER'S REFERENCE NUMBER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16476452 ↗
Other MAQUET GmbH products
- MEERA mobile operating table3 models · 32 reports · 4 recalls
- MEERA ST mobile operating table3 models · 1 reports · 3 recalls
- MEERA CL mobile operating table2 models · 1 reports · 2 recalls
- MAGNUS operating table system8 models · 39 reports · 1 recall
- CORIN mobile operating table6 models · 9 reports · 1 recall
- OTESUS operating table system3 models · 20 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page