ALPHAMAXX mobile operating table 113322B5Recalls & reported problems

MAQUET GmbH · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04046768040014), received between Oct 11, 2023 and May 14, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Circuit Failure” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Open, Classified

    The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.

    Z-2272-2024 · Root cause: Component design/selection · FDA record ↗

Most reported problems

Device problems

  • Circuit Failure2
  • Failure to Charge2
  • Use of Device Problem2
  • Break1
  • Failure to Run on Battery1
  • Positioning Failure1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Unspecified Tissue Injury2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 14, 2026Use of Device Problem

    GETINGE BECAME AWARE OF AN INCIDENT INVOLVING ONE OF OUR OPERATING TABLES. DURING SURGERY, THE TABLE FELL DIRECTLY ONTO NURSE'S FOOT. THIS LED TO THE STAFF MEMBER SUSTAINING AN INJURY WHICH REQUIRED IMMEDIATE MEDICAL INTERVENTION. WE DECIDED TO REPORT THE ISSUE DUE TO THE SERIOUS INJURY OF THE USER. ACCORDING TO THE GETINGE TECHNICIAN WHO EVALUATED THE OR TABLE ON SITE, THERE WAS NO MALFUNCTION OF THE TABLE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 25174839 ↗
  • InjuryMay 14, 2026Use of Device Problem

    GETINGE BECAME AWARE OF AN INCIDENT INVOLVING ONE OF OUR OPERATING TABLES. DURING SURGERY, THE TABLE TABLE FELL DIRECTLY ONTO NURSE'S FOOT. THIS LED TO THE STAFF MEMBER SUSTAINING AN INJURY WHICH REQUIRED IMMEDIATE MEDICAL INTERVENTION. WE DECIDED TO REPORT THE ISSUE DUE TO THE SERIOUS INJURY OF THE USER. ACCORDING TO THE GETINGE TECHNICIAN WHO EVALUATED THE OR TABLE ON SITE, THERE WAS NO MALFUNCTION OF THE TABLE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 25171448 ↗
  • MalfunctionJan 5, 2024Failure to Charge; Failure to Charge

    ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES - (B)(4)- ALPHAMAXX MOBILE OPERATING TABLE. ACCORDING TO THE INITIALLY PROVIDED INFORMATION, THE PATIENT HAD TO BE TRANSFERRED TO ANOTHER TABLE DUE TO BATTERY CHARGER FAILURE. ON (B)(6) 2023, CLARIFICATION WAS RECEIVED STATING THAT WHEN THE MALFUNCTION OCCURRED THE PATIENT WAS NOT YET ON THE TABLE. HOWEVER, IT WAS REQUIRED TO EXCHANGE THE TABLE WITH A DEVICE FROM ANOTHER THEATRE. IN THAT OPERATING ROOM THE PATIENT WAS ALREADY ANESTHETIZED FOR THE PROCEDURE. AS IT WAS CONFIRMED, THE MALFUNCTION RESULTED IN A DELAY OF LISTS IN BOTH ROOMS. THE INCIDENT WAS INITIALLY ASSESSED AS REPORTABLE DUE TO A DELAY IN SURGERY RESULTING IN PROLONGED ANESTHESIA TIME. RECENTLY, THE INCIDENT HAS BEEN REEVALUATED AS ACCORDING TO THE MEDICAL EXPERT'S ASSESSMENT, THE PROCEDURAL DELAY DID NOT HAVE ANY MAJOR CLINICAL RELEVANCE IN THIS SITUATION. THE SCENARIO DESCRIBED IN THE RECORD IS CONSIDERED AS NON-REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18453172 ↗
  • MalfunctionNov 20, 2023Failure to Charge; Circuit Failure; Positioning Failure; Failure to Run on Battery; Failure to Charge; Circuit Failure; Positioning Failure; Failure to Run on Battery

    ON 8TH NOVEMBER 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES ¿ 113322B5 - ALPHAMAXX (460 MM LONGIT. SHIFT), EU. AS IT WAS STATED, THE TABLE COULD NOT BE MOVED AT ALL DURING HIP SURGERY ON THE ANESTHETIZED PATIENT EVEN WITH A USE OF THE OVERRIDE PANEL DUE TO A FAILURE OF CHARGER BOARD AND BATTERY. THE SURGERY WAS CONTINUED ON THE AFFECTED DEVICE, HOWEVER, THE SURGEON HAD TO BE SEATED FOR THE PROCEDURE AND THE STAFF STRUGGLED GETTING THE X-RAY MACHINE UNDER THE TABLE. NO DELAY IN SURGERY OCCURRED DUE TO THE ISSUE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY INABILITY TO MOVE THE TABLE DURING THE SURGERY DUE TO CHARGER BOARD MALFUNCTION AFFECTING PATIENT¿S POSITIONING, WAS TO REOCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18169373 ↗
  • MalfunctionOct 11, 2023Break; Circuit Failure; Unintended Movement; Break; Circuit Failure; Unintended Movement

    ON 25TH SEPTEMBER 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR MOBILE TABLES ¿ 113322B5 ALPHAMAXX (460 MM LONGIT. SHIFT), EU USED WITH HAND CONTROL. AS IT WAS STATED, UNINTENDED TILT MOVEMENT OF THE TABLE OCCURRED DURING SURGERY WHILE THE PATIENT WAS ON THE TABLE. DURING SERVICE VISIT, INTERNAL DAMAGE OF THE HAND CONTROL¿S CABLE HAS BEEN IDENTIFIED WHICH COULD HAVE CONTRIBUTED TO THE INCIDENT. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION, NAMELY INVOLUNTARY TILT MOVEMENT OF THE TABLE WHICH COULD POTENTIALLY LEAD TO A PATIENT'S FALL, WAS TO REOCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17910378 ↗

Other TABLE, OPERATING-ROOM, AC-POWERED devices