Skip to content

ELAN 4 GA824Recalls & reported problems

Aesculap AG · ELAN 4 ELECTRO LOW SPEED MOTOR INTRA · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046963793876), received between Jul 20, 2020 and Jul 20, 2020 — classified as 2 injuries.

The most frequently reported specific device problem is “Overheating of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Overheating of Device2
  • Temperature Problem1

Patient problems

  • Burn, Thermal2
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 20, 2020Overheating of Device; Temperature Problem; Temperature Problem

    ASSOCIATED MEDWATCH-REPORTS: 2916714-2020-00243 (400476403 + GA824), AND 2916714-2020-00242 . INVOLVED COMPONENTS: GD450M - MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM - SERIAL NO. 2017 1006319. ADDITIONAL INFORMATION RECEIVED VIA INFORMATION FROM MW5095342: THE PATIENT WAS A SIX YEAR-OLD FEMALE. THE DEVICE WAS USED A PROCEDURE FOR MAXILLARY HYPOPLASIA, MANDIBULAR HYPOPLASIA, APERTOGNATHIA AND ASYMMETRY. THE PATIENT UNDERWENT A LEFORT 1 SEGMENTAL 3-PIECE OSTEOTOMY, BILATERAL SAGITTAL SPLIT OSTEOTOMY AND GENIOPLASTY. UPON USING THE LEDEMANN BUR, IT WAS NOTED THAT THE AESCULAP DRILL BECAME VERY HOT; DURING THE REMOVAL OF THE DRILL FROM THE PATIENT'S MOUTH, IT WAS NOTED THAT THE EXTENSION OF THE HAND PIECE WAS HOT AND HAD BURNED THE PATIENT'S LOWER RIGHT LIP. COLD SALINE WAS PLACED ON THE LIP IMMEDIATELY, AND SILVER SULFADIAZINE WAS PLACED, AND CONTINUED TO BE PLACED THROUGHOUT THE CASE. A NEW DRILL WAS REQUESTED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. AFTER THE PROCEDURE, THE STERILIZATION PROCESS WAS REVIEWED AND NO FAILURES IN THE STERILIZATION OF THE DEVICE WERE NOTED.

    Patient problems: Burn, Thermal; Superficial (First Degree) Burn

    MAUDE report 10297687 ↗
  • InjuryJul 20, 2020Overheating of Device; Overheating of Device

    ASSOCIATED MEDWATCH-REPORTS: 9610612-2020-00267 (400476402 + GD450M) INVOLVED COMPONENTS: GD450M - MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM - SERIAL NO.(B)(6). ADDITIONAL INFORMATION RECEIVED VIA INFORMATION FROM (B)(4): THE PATIENT WAS A SIX YEAR-OLD FEMALE. THE DEVICE WAS USED IN A PROCEDURE FOR MAXILLARY HYPOPLASIA, MANDIBULAR HYPOPLASIA, APERTOGNATHIA AND ASYMMETRY. THE PATIENT UNDERWENT A LEFORT 1 SEGMENTAL 3-PIECE OSTEOTOMY, BILATERAL SAGITTAL SPLIT OSTEOTOMY AND GENIOPLASTY. UPON USING THE LEDEMANN BUR, IT WAS NOTED THAT THE AESCULAP DRILL BECAME VERY HOT; DURING THE REMOVAL OF THE DRILL FROM THE PATIENT'S MOUTH, IT WAS NOTED THAT THE EXTENSION OF THE HAND PIECE WAS HOT AND HAD BURNED THE PATIENT'S LOWER RIGHT LIP. COLD SALINE WAS PLACED ON THE LIP IMMEDIATELY, AND SILVER SULFADIAZINE WAS PLACED, AND CONTINUED TO BE PLACED THROUGHOUT THE CASE. A NEW DRILL WAS REQUESTED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. AFTER THE PROCEDURE, THE STERILIZATION PROCESS WAS REVIEWED AND NO FAILURES IN THE STERILIZATION OF THE DEVICE WERE NOTED.

    Patient problems: Burn, Thermal

    MAUDE report 10297686 ↗

FDA clearance and classification

The FDA regulates product code HBC (motor, drill, electric) under 21 CFR 882.4360 as a Class II device. 21 CFR 882.4360 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ELAN 4 GA824

Has the ELAN 4 GA824 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04046963793876), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ELAN 4 GA824?

The FDA's MAUDE database lists 2 adverse event reports for this device, received on Jul 20, 2020. In those reports, the most common specific device problems are “Overheating of Device” (2) and “Temperature Problem” (1); the most common patient problems are “Burn, Thermal” (2) and “Superficial (First Degree) Burn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ELAN 4 GA824?

None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the ELAN 4 GA824, and is it still sold?

The FDA's device database lists Aesculap AG as the company responsible for it (brand ELAN 4, model GA824). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other MOTOR, DRILL, ELECTRIC devices

All MOTOR, DRILL, ELECTRIC devices →

Sources and updates