ELAN 4 GA824Recalls & reported problems
Aesculap AG · ELAN 4 ELECTRO LOW SPEED MOTOR INTRA · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046963793876), received between Jul 20, 2020 and Jul 20, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Overheating of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Overheating of Device2
- Temperature Problem1
Patient problems
- Burn, Thermal2
- Superficial (First Degree) Burn1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 20, 2020Overheating of Device; Temperature Problem; Temperature Problem
ASSOCIATED MEDWATCH-REPORTS: 2916714-2020-00243 (400476403 + GA824), AND 2916714-2020-00242 . INVOLVED COMPONENTS: GD450M - MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM - SERIAL NO. 2017 1006319. ADDITIONAL INFORMATION RECEIVED VIA INFORMATION FROM MW5095342: THE PATIENT WAS A SIX YEAR-OLD FEMALE. THE DEVICE WAS USED A PROCEDURE FOR MAXILLARY HYPOPLASIA, MANDIBULAR HYPOPLASIA, APERTOGNATHIA AND ASYMMETRY. THE PATIENT UNDERWENT A LEFORT 1 SEGMENTAL 3-PIECE OSTEOTOMY, BILATERAL SAGITTAL SPLIT OSTEOTOMY AND GENIOPLASTY. UPON USING THE LEDEMANN BUR, IT WAS NOTED THAT THE AESCULAP DRILL BECAME VERY HOT; DURING THE REMOVAL OF THE DRILL FROM THE PATIENT'S MOUTH, IT WAS NOTED THAT THE EXTENSION OF THE HAND PIECE WAS HOT AND HAD BURNED THE PATIENT'S LOWER RIGHT LIP. COLD SALINE WAS PLACED ON THE LIP IMMEDIATELY, AND SILVER SULFADIAZINE WAS PLACED, AND CONTINUED TO BE PLACED THROUGHOUT THE CASE. A NEW DRILL WAS REQUESTED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. AFTER THE PROCEDURE, THE STERILIZATION PROCESS WAS REVIEWED AND NO FAILURES IN THE STERILIZATION OF THE DEVICE WERE NOTED.
Patient problems: Burn, Thermal; Superficial (First Degree) Burn
MAUDE report 10297687 ↗InjuryJul 20, 2020Overheating of Device; Overheating of Device
ASSOCIATED MEDWATCH-REPORTS: 9610612-2020-00267 (400476402 + GD450M) INVOLVED COMPONENTS: GD450M - MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM - SERIAL NO.(B)(6). ADDITIONAL INFORMATION RECEIVED VIA INFORMATION FROM (B)(4): THE PATIENT WAS A SIX YEAR-OLD FEMALE. THE DEVICE WAS USED IN A PROCEDURE FOR MAXILLARY HYPOPLASIA, MANDIBULAR HYPOPLASIA, APERTOGNATHIA AND ASYMMETRY. THE PATIENT UNDERWENT A LEFORT 1 SEGMENTAL 3-PIECE OSTEOTOMY, BILATERAL SAGITTAL SPLIT OSTEOTOMY AND GENIOPLASTY. UPON USING THE LEDEMANN BUR, IT WAS NOTED THAT THE AESCULAP DRILL BECAME VERY HOT; DURING THE REMOVAL OF THE DRILL FROM THE PATIENT'S MOUTH, IT WAS NOTED THAT THE EXTENSION OF THE HAND PIECE WAS HOT AND HAD BURNED THE PATIENT'S LOWER RIGHT LIP. COLD SALINE WAS PLACED ON THE LIP IMMEDIATELY, AND SILVER SULFADIAZINE WAS PLACED, AND CONTINUED TO BE PLACED THROUGHOUT THE CASE. A NEW DRILL WAS REQUESTED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. AFTER THE PROCEDURE, THE STERILIZATION PROCESS WAS REVIEWED AND NO FAILURES IN THE STERILIZATION OF THE DEVICE WERE NOTED.
Patient problems: Burn, Thermal
MAUDE report 10297686 ↗
Other MOTOR, DRILL, ELECTRIC devices
- NA CRANI-A-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC1 recall358 reports
- NA QD8Synthes GmbH · MOTOR, DRILL, ELECTRIC386 reports
- NA QD11-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC312 reports
- Midas Rex EM200Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC254 reports
- MIDAS REX MR8 EM800Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC186 reports
- Midas Rex MR8 EM800Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC158 reports
- NA SHORT-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC134 reports
- Midas Rex® EM200Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC123 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.