Skip to content

Infusomat® 490038Recalls & reported problems

B. Braun Medical Inc. · 60 drops/mL, priming volume: 11mL, length: 129in · Class 2

All models of Infusomat (54)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04046964185946), received between Jan 4, 2024 and Oct 11, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2023Open, Classified

    In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

    Z-0355-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash2
  • Device Alarm System1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 11, 2024Leak/Splash; Leak/Splash

    AS REPORTED, THE BLUE-GREEN VENT COVER FELL OFF THE VENT ON THE SPIKE. THIS MADE SOME OF THE CHEMO MEDICATION TO COME OUT OF THE VENT. THE NURSE QUICKLY PUT THE VENT COVER BACK IN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20427409 ↗
  • MalfunctionJan 29, 2024Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: LEAKING AT THE PUMP SEGMENT. DETAILED INQUIRY DESCRIPTION: LEAKING AT THE PUMP SEGMENT DURING PRIMING BEFORE THE ADMINISTRATION OF CHEMO. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18595672 ↗
  • MalfunctionJan 4, 2024Device Alarm System; Device Alarm System

    AS REPORTED BY THE USER FACILITY: INFUSOMAT SPACE PUMP IV SET PORTLESS TUBING/EPIDURAL ADMIN TUBING SET ALARMS AND DOES NOT ALLOW USER TO INFUSE CRITICAL MED (IE: FENTANYL). NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18443560 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →