InfusomatRecalls & reported problems across all models

B. Braun Medical Inc. · 54 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

30 open recalls name at least one Infusomat model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
2,783
Deaths reported
38
Injuries reported
97
Recalls
30

We found 54 models of Infusomat by B. Braun Medical Inc. in the FDA's device registry (GUDID). Together they are named in 2,783 adverse event reports in the FDA's MAUDE database, received between Nov 18, 2016 and Sep 29, 2026 — classified as 38 deaths, 97 injuries, 2,648 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (884 reports).

30 FDA recall records name at least one of these models by its device identifier, 30 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 52 reportsDec 2023: 57 reportsJan 2024: 80 reportsFeb 2024: 81 reportsMar 2024: 32 reportsApr 2024: 74 reportsMay 2024: 83 reportsJun 2024: 70 reportsJul 2024: 50 reportsAug 2024: 19 reportsSep 2024: 50 reportsOct 2024: 81 reportsNov 2024: 69 reportsDec 2024: 96 reportsJan 2025: 78 reportsFeb 2025: 52 reportsMar 2025: 81 reportsApr 2025: 26 reportsMay 2025: 65 reportsJun 2025: 89 reportsJul 2025: 42 reportsAug 2025: 46 reportsSep 2025: 18 reportsOct 2025: 17 reportsNov 2025: 17 reportsDec 2025: 13 reportsJan 2026: 25 reportsFeb 2026: 21 reportsMar 2026: 28 reportsApr 2026: 28 reportsMay 2026: 23 reportsJun 2026: 30 reportsJul 2026: 8 reportsAug 2026: 5 reportsSep 2026: 16 reportsOct 2026: 0 reports20242025202696
Last 36 months. The most recent month may be incomplete.

Infusomat recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Oct 29, 2025Open, Classified

    Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

    Z-0640-2026 · Root cause: Process control · names 32 of these models · Recall details · FDA record ↗

  2. Aug 7, 2024Open, Classified

    Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected.

    Z-2972-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

  3. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2403-2024 · Root cause: Process control · Recall details · FDA record ↗

  4. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2401-2024 · Root cause: Process control · Recall details · FDA record ↗

  5. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2400-2024 · Root cause: Process control · Recall details · FDA record ↗

  6. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2399-2024 · Root cause: Process control · Recall details · FDA record ↗

  7. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2398-2024 · Root cause: Process control · Recall details · FDA record ↗

  8. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2397-2024 · Root cause: Process control · Recall details · FDA record ↗

  9. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2396-2024 · Root cause: Process control · Recall details · FDA record ↗

  10. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2395-2024 · Root cause: Process control · Recall details · FDA record ↗

  11. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2394-2024 · Root cause: Process control · Recall details · FDA record ↗

  12. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2393-2024 · Root cause: Process control · Recall details · FDA record ↗

  13. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2392-2024 · Root cause: Process control · Recall details · FDA record ↗

  14. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2391-2024 · Root cause: Process control · Recall details · FDA record ↗

  15. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2390-2024 · Root cause: Process control · Recall details · FDA record ↗

  16. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2389-2024 · Root cause: Process control · Recall details · FDA record ↗

  17. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2388-2024 · Root cause: Process control · Recall details · FDA record ↗

  18. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2387-2024 · Root cause: Process control · Recall details · FDA record ↗

  19. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2386-2024 · Root cause: Process control · Recall details · FDA record ↗

  20. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2385-2024 · Root cause: Process control · Recall details · FDA record ↗

  21. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2384-2024 · Root cause: Process control · Recall details · FDA record ↗

  22. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2383-2024 · Root cause: Process control · Recall details · FDA record ↗

  23. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2382-2024 · Root cause: Process control · Recall details · FDA record ↗

  24. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2381-2024 · Root cause: Process control · Recall details · FDA record ↗

  25. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2380-2024 · Root cause: Process control · Recall details · FDA record ↗

  26. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2379-2024 · Root cause: Process control · Recall details · FDA record ↗

  27. Jun 17, 2024Open, Classified

    Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

    Z-2378-2024 · Root cause: Process control · Recall details · FDA record ↗

  28. Nov 1, 2023Open, Classified

    In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

    Z-0355-2024 · Root cause: Process control · names 41 of these models · Recall details · FDA record ↗

  29. Sep 21, 2023Open, Classified

    Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.

    Z-0124-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

  30. Sep 21, 2023Open, Classified

    Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.

    Z-0123-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported Infusomat problems

Device problems

  • Air/Gas in Device884
  • Device Alarm System506
  • Leak/Splash353
  • Excess Flow or Over-Infusion332
  • Fluid/Blood Leak240
  • Insufficient Flow or Under Infusion167
  • Device Misassembled During Manufacturing /Shipping86
  • Contamination /Decontamination Problem66

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,115
  • No Known Impact Or Consequence To Patient165
  • No Consequences Or Impact To Patient139
  • Low Blood Pressure/ Hypotension127
  • Cardiac Arrest48
  • Chemical Exposure38
  • Hemodynamic instability35
  • Distress33

Infusomat models

54 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Infusomat adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other B. Braun Medical Inc. products