Skip to content

ACCEL® 622275Recalls & reported problems

B. Braun Melsungen Aktiengesellschaft · 1000ML DRAINAGE BOTTLE, PVC-FREE · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04046964613371), received between Oct 17, 2022 and Mar 11, 2026 — classified as 3 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Suction Problem” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Suction Problem10
  • Degraded2
  • Loose or Intermittent Connection2

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Pneumothorax4
  • Ascites2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2026Loose or Intermittent Connection; Suction Problem

    AS REPORTED BY THE USER FACILITY: DETAIL: WHEN ATTACHING THE LUER LOCK TO THE BOTTLE, IT WAS EXTREMELY DIFFICULT TO CONNECT THE TUBING AND ESTABLISH SUCTION. AS A RESULT, THE VACUUM SOMEHOW RELEASED, AND WHEN THIS OCCURRED, AIR FROM THE BOTTLE WAS PUSHED INTO THE PATIENT, CAUSING A PNEUMOTHORAX THAT REQUIRED CHEST TUBE PLACEMENT. AFTER THAT INCIDENT, WE PERFORMED A PARACENTESIS AND AGAIN WERE UNABLE TO SECURELY CONNECT THE LUER LOCK TUBING TO ACHIEVE SUCTION. WE ATTEMPTED THIS WITH MULTIPLE BOTTLES AND WERE UNSUCCESSFUL EACH TIME. THESE WERE THE FIRST TWO PATIENTS FOR WHOM WE HAD USED THESE BOTTLES. WE TRIED TO ESTABLISH A CONNECTION WITH EMPTY BOTTLES NOT ATTACHED TO THE PATIENT AND SUCCESSFULLY GOT THE SUCTION TO RELEASE ONCE OUT OF FIVE.

    Patient problems: Ascites; Pneumothorax

    MAUDE report 24568071 ↗
  • InjuryMar 11, 2026Loose or Intermittent Connection; Suction Problem

    AS REPORTED BY THE USER FACILITY: DETAIL: WHEN ATTACHING THE LUER LOCK TO THE BOTTLE, IT WAS EXTREMELY DIFFICULT TO CONNECT THE TUBING AND ESTABLISH SUCTION. AS A RESULT, THE VACUUM SOMEHOW RELEASED, AND WHEN THIS OCCURRED, AIR FROM THE BOTTLE WAS PUSHED INTO THE PATIENT, CAUSING A PNEUMOTHORAX THAT REQUIRED CHEST TUBE PLACEMENT. AFTER THAT INCIDENT, WE PERFORMED A PARACENTESIS AND AGAIN WERE UNABLE TO SECURELY CONNECT THE LUER LOCK TUBING TO ACHIEVE SUCTION. WE ATTEMPTED THIS WITH MULTIPLE BOTTLES AND WERE UNSUCCESSFUL EACH TIME. THESE WERE THE FIRST TWO PATIENTS FOR WHOM WE HAD USED THESE BOTTLES. WE TRIED TO ESTABLISH A CONNECTION WITH EMPTY BOTTLES NOT ATTACHED TO THE PATIENT AND SUCCESSFULLY GOT THE SUCTION TO RELEASE ONCE OUT OF FIVE.

    Patient problems: Ascites; Pneumothorax

    MAUDE report 24567854 ↗
  • MalfunctionMay 5, 2023Suction Problem; Suction Problem

    AS REPORTED BY THE USER FACILITY: EVENT 2: DETAILED INQUIRY DESCRIPTION: NURSES REALIZED THAT THE BOTTLES HAD NO SUCTION AFTER CONNECTING THE LINE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16875407 ↗
  • MalfunctionMay 5, 2023Suction Problem; Suction Problem

    AS REPORTED BY THE USER FACILITY: EVENT 6: DETAILED INQUIRY DESCRIPTION: NURSES REALIZED THAT THE BOTTLES HAD NO SUCTION AFTER CONNECTING THE LINE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16875405 ↗
  • MalfunctionMay 5, 2023Suction Problem; Suction Problem

    AS REPORTED BY THE USER FACILITY: EVENT 5: DETAILED INQUIRY DESCRIPTION: NURSES REALIZED THAT THE BOTTLES HAD NO SUCTION AFTER CONNECTING THE LINE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16875404 ↗
  • MalfunctionMay 5, 2023Suction Problem; Suction Problem

    AS REPORTED BY THE USER FACILITY: EVENT 1: DETAILED INQUIRY DESCRIPTION: NURSES REALIZED THAT THE BOTTLES HAD NO SUCTION AFTER CONNECTING THE LINE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16875403 ↗
  • MalfunctionMay 5, 2023Suction Problem; Suction Problem

    AS REPORTED BY THE USER FACILITY: EVENT 4: DETAILED INQUIRY DESCRIPTION: NURSES REALIZED THAT THE BOTTLES HAD NO SUCTION AFTER CONNECTING THE LINE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16875402 ↗
  • MalfunctionMay 5, 2023Suction Problem; Suction Problem

    AS REPORTED BY THE USER FACILITY: EVENT 3: DETAILED INQUIRY DESCRIPTION: NURSES REALIZED THAT THE BOTTLES HAD NO SUCTION AFTER CONNECTING THE LINE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16875401 ↗
  • InjuryOct 17, 2022Degraded; Suction Problem; Degraded; Suction Problem

    THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). AS REPORTED BY THE USER FACILITY: EVENT DESCRIPTION: PROVIDER WAS PERFORMING A CENTESIS AND WHEN BOTTLE HAD ABOUT 300ML LEFT TO FILL A HISSING SOUND WAS HEARD AND AIR BEGAN TO FILL PATIENT LUNG. THE GREY RUBBER TOP WAS CRACKED AND BROKE THUS BREAKING THE VACUMM.

    Patient problems: Pneumothorax

    MAUDE report 15616054 ↗
  • InjuryOct 17, 2022Degraded; Suction Problem; Degraded; Suction Problem

    AS REPORTED BY THE USER FACILITY: EVENT DESCRIPTION: PROVIDER WAS PERFORMING A CENTESIS AND WHEN BOTTLE HAD ABOUT 300ML LEFT TO FILL A HISSING SOUND WAS HEARD AND AIR BEGAN TO FILL PATIENT LUNG. THE GREY RUBBER TOP WAS CRACKED AND BROKE THUS BREAKING THE VACUMM.

    Patient problems: Pneumothorax

    MAUDE report 15615968 ↗

Other BOTTLE, COLLECTION, VACUUM devices

All BOTTLE, COLLECTION, VACUUM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page