Infusomat® 363904Recalls & reported problems
B. Braun Medical Inc. · UNIV. 15 DROP PUMP SET, ASV · Class 2
All models of Infusomat (54)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This B. Braun Infusomat pump set, an intravenous tubing set used with an infusion pump, has 116 adverse event reports on file. Many describe false air-in-line or bubble alarms that stop infusions. The reports include no deaths and no injuries, and the number of reports fell from 36 to 1 over the last 12 months. There is 1 open recall, begun in November 2023, over tubing that may be the wrong width or too short, causing false air alarms.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 116
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 116 adverse event reports that list this exact device (DI 04046964699061), received between Nov 18, 2019 and Apr 10, 2026 — classified as 116 malfunctions.
The most frequently reported specific device problem is “Air/Gas in Device” (67 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 1, 2023Open, Classified
In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.
Z-0355-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Air/Gas in Device67
- Device Alarm System66
- Leak/Splash2
- Contamination /Decontamination Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions108
- Low Blood Pressure/ Hypotension6
- Insufficient Information1
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 116 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 10, 2026Device Alarm System
AS REPORTED BY THE USER FACILITY: AIR IN LINE; UNABLE TO RESOLVE DESPITE THOROUGH TROUBLESHOOTING. NO VISIBLE AIR IN SET. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24838762 ↗MalfunctionOct 24, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): "A NURSE IN THE SURGICAL MEDICAL INTENSIVE CARE UNIT REPORTED THAT DESPITE CHECKING THE CONNECTIONS AND TUBING AT THE BEGINNING OF HER SHIFT, THE "ANTI-SIPHON" VALVE AUTOMATICALLY LOOSENED FROM THE TUBING 4 HOURS LATER, CAUSING A VASOPRESSOR THAT WAS SCHEDULED TO BE RELEASED AT A HIGH DOSE. FORTUNATELY, THERE WAS NO SEVERE IMPACT ON THE PATIENT." 1. THE REPORT MENTIONS "A VASOPRESSOR THAT WAS SCHEDULED TO BE RELEASED AT A HIGH DOSE." COULD YOU CONFIRM WHICH OF THE FOLLOWING OCCURRED: EXTERNAL LEAKAGE FROM THE INFUSION SET. THE NURSE HAD DONE HER INITIAL CHECK AT THE START OF THE SHIFT, AND CONFIRMED THAT ALL CONNECTIONS, LINES AND TUBING'S WERE PROPERLY SECURED. ABOUT 4H INTO HER SHIFT, SHE NOTICED A LEAK AT THE CONNECTION SITE OF THE ANTI-SIPHON VALVE. IT APPEARED THAT THE VALVE HAD BECOME LOOSE SOMEHOW FROM THE ADJACENT TUBING, WHICH RESULTED IN THE LEAK. THE INFUSION HAD BEEN RUNNING CONTINUOUSLY THE ENTIRE TIME, ALTHOUGH TITRATION OF THE MEDICATION WAS PERFORMED AS REQUIRED. HOWEVER, THE INFUSION WAS NEVER PAUSED OR RESTARTED AT A HIGHER RATE, SO IT IS UNLIKELY THAT THE CAUSE OF THE LEAK WAS A PRESSURE SUR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23385020 ↗MalfunctionMar 13, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: INCIDENT DESCRIPTION: AMIODARONE INFUSION. BUBBLES ALARM- FREQUENT, IV TUBING LESS THAN 24 HOURS USE. PATIENT ON AMIODARONE INFUSION FOR 3 DAYS AND HAVE HAD TO CHANGE THE TUBING EVERY DAY DUE TO BUBBLE ALARM THAT CAN BE FIXED BY FLICKING THE BUBBLES OR REPRIMING THE TUBE AGAIN. FLUID ROOM TEMP DRIP CHAMBER 1/2-2/3 FILLED TUBING PROPERLY LOADED IV LINE PRIMED THROUGH PUMP UNUSED PRIME FUNCTION TO REMOVE AIR IN LINE ALARM RECURRED, REPRIME OPEN DOOR REMOVE AND VISUALIZE TUBING TAP TUBING SECTION AND OBSERVE FOR BUBBLES VISIBLE BUBBLES, USED ALCOHOL PAD AND WIPED SEGMENT OF TUBING BETWEEN THE SILICONE PUMP SEGMENT AND GREEN CLAMP, RELOAD TUBING AND RESTART. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21599039 ↗MalfunctionMar 12, 2025Air/Gas in Device; Air/Gas in Device
ADDITIONAL INFORMATION PROVIDED: ACCORDING TO THE COMPLAINANT, A 6-HOUR PROTAMINE INFUSION WAS BEING ADMINISTERED. THE FLUID WAS AT ROOM TEMPERATURE, WITH THE DRIP CHAMBER ½ TO 2/3 FILLED, AND THE TUBING PROPERLY LOADED. THE INTRAVENOUS (IV) LINE WAS PRIMED THROUGH THE PUMP, WITH THE PRIME FUNCTION BEING USED TO REMOVE AIR FROM THE LINE. DURING THE INFUSION, AN AIR IN LINE ALARM WAS TRIGGERED, PROMPTING THE FOLLOWING TROUBLESHOOTING STEPS. THE USER REPRIMED THE LINE, OPENED THE PUMP DOOR, REMOVED AND VISUALIZED THE TUBING, AND THEN TAPPED THE SECTION IN QUESTION TO CHECK FOR AIR BUBBLES. AIR BUBBLES WERE VISIBLE. THE USER THEN WIPED THE TUBING SEGMENT WITH AN ALCOHOL PAD, SPECIFICALLY IN THE AREA BETWEEN THE SILICONE PUMP SEGMENT AND THE GREEN CLAMP, BEFORE RELOADING THE TUBING AND RESTARTING THE PUMP. THE ALARM RECURRED, PROMPTING REPLACEMENT OF THE IV SET. ONE HOUR BEFORE COMPLETING THE TOTAL 6-HOUR DOSE, AN AIR BUBBLE ALARM WAS TRIGGERED AGAIN. THE USER ATTEMPTED TO FLICK THE BUBBLES, BUT THIS METHOD WAS INEFFECTIVE. ADDITIONALLY, THE USER FELT THAT FLUSHING THE LINE WAS NOT AS OPTION, AS IT WOULD RESULT IN LOST MEDICATION, PREVENTING THE PATIENT FROM RECEIVING THE FULL DOSE. TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21588493 ↗MalfunctionMar 5, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): SODIUM PHOSPHATE BUBBLES ALARM MULTIPLE TIMES EVEN AFTER BUBBLES REMOVED. SODIUM PHOSPHATE BAG HAS BEEN ON ROOM TEMP PRIOR TO HOOKING. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21526409 ↗MalfunctionMar 5, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): VASOPRESSIN DRIP BUBBLES ALARM. TUBING IS LESS THAN 8 IN USE AND HAVE TO CHANGE TUBING.
Patient problems: Low Blood Pressure/ Hypotension
MAUDE report 21526407 ↗MalfunctionMar 5, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): NOTED EXCESS BUBBLES. WASTED MEDS FROM REMOVING BUBBLES AND FLUSHING. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21526102 ↗MalfunctionMar 4, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): CHASER LINE LR. NEW IV STARTED ON (B)(6) AT 21:50. EXCESS BUBBLE ALARM EVEN ALL THROUGH BUBBLES. LINE CHANGED AT 09:45 (B)(6) ONLY 24 HOURS IN USE. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21516076 ↗MalfunctionMar 4, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): TUBE IN USE LESS THAN 80 NEED TO CHANGE TUBINGS D/T BUBBLES ALARM. TUBING FLICKED FOR BUBBLES MULTIPLE TIMES INTERFERING MED INFUSION (NOREPI, VARO, INSULIN, PROPOFUL. NO INJURY REPORTED).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21516027 ↗MalfunctionJan 31, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: EVENT 3: DETAILS OF COMPLAINT (REPORTED ISSUE): "AIR IN LINE" ALARM. AMIODARONE INFUSION. BUBBLES ALARM FREQUENT. IV TUBING LESS THAN 24 USE. PT ON AMINO INFUS X 3 DAYS AND WE HAVE BEEN CHANGING TUBING'S EVERYDAY BUBBLE ALARM THAT CAN BE FIXED BY FLICKING THE BUBBLES ON REPRIMING THE TUBE AGAIN." NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21286993 ↗MalfunctionJan 31, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: EVENT 2: DETAILS OF COMPLAINT (REPORTED ISSUE): "AIR IN LINE" ALARM. AMIODARONE INFUSION. BUBBLES ALARM FREQUENT. IV TUBING LESS THAN 24 USE. PT ON AMINO INFUS X 3 DAYS AND WE HAVE BEEN CHANGING TUBING'S EVERYDAY BUBBLE ALARM THAT CAN BE FIXED BY FLICKING THE BUBBLES ON REPRIMING THE TUBE AGAIN." NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21286992 ↗MalfunctionJan 31, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: EVENT 1: DETAILS OF COMPLAINT (REPORTED ISSUE): "AIR IN LINE" ALARM. AMIODARONE INFUSION. BUBBLES ALARM FREQUENT. IV TUBING LESS THAN 24 USE. PT ON AMINO INFUS X 3 DAYS AND WE HAVE BEEN CHANGING TUBING'S EVERYDAY BUBBLE ALARM THAT CAN BE FIXED BY FLICKING THE BUBBLES ON REPRIMING THE TUBE AGAIN." NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21286991 ↗MalfunctionJan 23, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: EVENT 2: DETAILS OF COMPLAINT (REPORTED ISSUE): PT IN ISOLATION FOR COVID +, MULTIPLE ALARMS FOR AIR IN LINE 0 RESOLVED WITH FLUSH MULTIPLE TIMES PUMP AND PULLING OUT + FLUSHING. ENTIRE BAG OF MED WAS FLUSHED WITHIN 5 HOURS AND WASTED + USED 2 IV LINES TRYING TO RESOLVE ALARMS. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21216596 ↗MalfunctionJan 23, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: EVENT 1: DETAILS OF COMPLAINT (REPORTED ISSUE): PT IN ISOLATION FOR COVID +, MULTIPLE ALARMS FOR AIR IN LINE 0 RESOLVED WITH FLUSH MULTIPLE TIMES PUMP AND PULLING OUT + FLUSHING. ENTIRE BAG OF MED WAS FLUSHED WITHIN 5 HOURS AND WASTED + USED 2 IV LINES TRYING TO RESOLVE ALARMS. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21216595 ↗MalfunctionJan 23, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): "AIR IN LINE" ALARM. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21216544 ↗MalfunctionJan 22, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: INCIDENT_DESCRIPTION: CONTINUOUS AIR IN LINE DRIP CHAMBER 1/2-2/3 FILLED TUBING PROPERLY LOADED IV LINE PRIMED THROUGH PUMP USED PRIME FUNCTION TO REMOVE AIR IN LINE ALARM RECURRED, RE-PRIME OPENED DOOR, REMOVE AND VISUALIZE TUBING TAP TUBING SECTION AND OBSERVE FOR BUBBLES VISIBLE BUBBLES, USED ALCOHOL PAD AND WIPED SEGMENT OF TUBING BETWEEN THE SILICONE PUMP SEGMENT AND GREEN CLAMP, RELOAD TUBING AND RESTART ALARM RECURRED, CHANGED TUBING UPSTREAM AND DOWNSTREAM OCCLUSIONS, NO ISSUES WITH PATIENTS LINE IV SITE, TUBING CLAMPS OPEN, NO KINKS PRESSURE ADJUSTED AND INFUSION RESTARTED, ALARM RECURRED CHANGED THE TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21206602 ↗MalfunctionJan 21, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): "AIR IN LINE" ALARM. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196955 ↗MalfunctionJan 21, 2025Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): "AIR IN LINE" ALARM. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196953 ↗MalfunctionJan 21, 2025Air/Gas in Device; Air/Gas in Device
ADDITIONAL INFORMATION PROVIDED: ACCORDING TO THE COMPLAINANT, A 6-HOUR PROTAMINE INFUSION WAS BEING ADMINISTERED. THE FLUID WAS AT ROOM TEMPERATURE, WITH THE DRIP CHAMBER ½ TO 2/3 FILLED, AND THE TUBING PROPERLY LOADED. THE INTRAVENOUS (IV) LINE WAS PRIMED THROUGH THE PUMP, WITH THE PRIME FUNCTION BEING USED TO REMOVE AIR FROM THE LINE. DURING THE INFUSION, AN AIR IN LINE ALARM WAS TRIGGERED, PROMPTING THE FOLLOWING TROUBLESHOOTING STEPS. THE USER REPRIMED THE LINE, OPENED THE PUMP DOOR, REMOVED AND VISUALIZED THE TUBING, AND THEN TAPPED THE SECTION IN QUESTION TO CHECK FOR AIR BUBBLES. AIR BUBBLES WERE VISIBLE. THE USER THEN WIPED THE TUBING SEGMENT WITH AN ALCOHOL PAD, SPECIFICALLY IN THE AREA BETWEEN THE SILICONE PUMP SEGMENT AND THE GREEN CLAMP, BEFORE RELOADING THE TUBING AND RESTARTING THE PUMP. THE ALARM RECURRED, PROMPTING REPLACEMENT OF THE IV SET. ONE HOUR BEFORE COMPLETING THE TOTAL 6-HOUR DOSE, AN AIR BUBBLE ALARM WAS TRIGGERED AGAIN. THE USER ATTEMPTED TO FLICK THE BUBBLES, BUT THIS METHOD WAS INEFFECTIVE. ADDITIONALLY, THE USER FELT THAT FLUSHING THE LINE WAS NOT AS OPTION, AS IT WOULD RESULT IN LOST MEDICATION, PREVENTING THE PATIENT FROM RECEIVING THE FULL DOSE. TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196952 ↗MalfunctionJan 7, 2025Device Alarm System; Air/Gas in Device; Device Alarm System; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: MEDICATION NEEDED URGENTLY FROM FRIDGE SO WAS NOT AT ROOM TEMPERATURE. AIR IN LINE ALARM. TROUBLESHOOTING COMPLETED. TUBING LOADED PROPERLY, LINE PRIMED, DOOR OPENED, NO VISIBLE AIR BUBBLES. ONGOING ISSUES WITH COLD MEDS FROM FRIDGE. NO SAMPLE AVAILABLE. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21094034 ↗MalfunctionDec 11, 2024Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: INCIDENT DESCRIPTION: AIR INLINE UNABLE TO RESOLVE AFTER PRIMING 3 TIMES AND MANUAL PRIME. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20904599 ↗MalfunctionDec 11, 2024Air/Gas in Device; Air/Gas in Device
AS REPORTED BY THE USER FACILITY: INCIDENT DESCRIPTION: DRIP CHAMBER 1/2-2/3 FILLED TUBING PROPERLY LOADED IV LINE PRIMED THROUGH PUMP UNSED PRIME FUNCTION TO REMOVE AIR IN LINE ALARM RECURRED, REPRIME OPEN DOOR REMOVE AND VISUALIZE TUBING TAP TUBING SECTION AND OBSERVE FOR BUBBLES VISIBLE BUBBLES, USED ALCOHOL PAD AND WIPED SEGMENT OF TUBING BETWEEN THE SILICONE PUMP SEGMENT AND GREEN CLAMP, RELOAD TUBING AND RESTART ALARM RECURRED, CHANGED TUBING COULD NO CLEAR BUBBLES, PRIMED X2 STILL BUBBLE AND AIR BUBBLE ALARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20904541 ↗MalfunctionNov 20, 2024Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED TO B. BRAUN MEDICAL INC. THAT AN INFUSOMAT SPACE PUMP INTRAVENOUS (IV) PUMP SET (MATERIAL 363904, LOT UNKNOWN) WAS BEING USED TO ADMINISTER AMIODARONE AT 44 MILLILITERS (ML) PER HOUR ON (B)(6) 2024. ACCORDING TO THE COMPLAINANT, THERE WAS AIR WITHIN THE IV LINE. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. FROM CHECKLIST, AIR IN LINE. USED PRIME FUNCTION TO REMOVE THE AIR IN THE LINE AND INFUSION RESTARTED. ALARM REOCCURS, USED PRIME FUNCTION AGAIN TO REMOVE THE AIR IN THE LINE AND RESTART INFUSION. OPEN DOOR, REMOVE AND VISUALIZE, TAB THE TUBING SECTION AND OBSERVE FOR BUBBLES. NO VISIBLE BUBBLES, RELOAD THE TUBING AND RESTART INFUSION. ALARM RE-OCCURS. CHANGED THE TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20734607 ↗MalfunctionNov 20, 2024Device Alarm System; Device Alarm System
IT WAS REPORTED TO B. BRAUN MEDICAL INC. THAT AN INFUSOMAT SPACE PUMP INTRAVENOUS (IV) PUMP SET (MATERIAL 363904, LOT 0061939231) WAS BEING USED TO ADMINISTER MILRINONE ON (B)(6) 2024. ACCORDING TO THE COMPLAINANT, THE PUMP ALARMED FOR AIR IN LINE, HOWEVER, THERE WERE NO VISIBLE BUBBLES IN THE IV LINE. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. FROM CHECKLIST: AIR IN LINE. USED PRIME FUNCTION TO REMOVE THE AIR IN THE LINE AND INFUSION RESTARTED. ALARM REOCCURS, USED PRIME FUNCTION AGAIN TO REMOVE THE AIR IN THE LINE AND RESTART INFUSION. OPEN DOOR, REMOVE AND VISUALIZE TUBING, TAP THE TUBING SECTION AND OBSERVE FOR BUBBLES. NO VISIBLE BUBBLES, RELOAD THE TUBING AND RESTART INFUSION. ALARM RE-OCCURS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20734490 ↗MalfunctionNov 18, 2024Device Alarm System; Device Alarm System
IT WAS REPORTED TO B. BRAUN MEDICAL INC. THAT AN INFUSOMAT SPACE INTRAVENOUS (IV) PUMP SET (MATERIAL 363904, LOT UNKNOWN) WAS BEING USED TO ADMINISTER 4 GRAMS (G) MAGNESIUM AT A RATE OF APPROXIMATELY 50 MILLILITERS PER HOUR (ML/HR) ON (B)(6) 2024. ACCORDING TO THE COMPLAINANT, THE PUMP INITIALLY ALARMED FOR AN UPSTREAM OCCLUSION. THE PRIME FUNCTION WAS USED TO REMOVE AIR FROM THE LINE AND THE INFUSION WAS RESTARTED. HOWEVER, AN AIR IN LINE ALARM OCCURRED. THE PRIME FUNCTION WAS USED ONCE AGAIN TO SUCCESSFULLY PRIME THE LINE, BUT THE AIR-IN-LINE ALARM CONTINUED TO OCCUR. AT THIS TIME, THE DOOR WAS OPENED, THE TUBING WAS REMOVED, INSPECTED, AND TAPPED, TO OBSERVE AND CONFIRM THAT NO AIR WAS PRESENT IN THE LINES. THE TUBING WAS RELOADED, AND THE INFUSION WAS RESTARTED, BUT THE ALARM REOCCURRED PROMPTING THE USE OF A NEW PUMP SET. REPORTEDLY, THE CHEMOTHERAPY HAD TO EVENTUALLY BE RUN VIA GRAVITY. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20705579 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page