ENNOVATE SZ378RRecalls & reported problems
Aesculap AG · ENNOVATE MIS DOWNTUBE SHORT · Class 1
All models of ENNOVATE (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04046964719561), received between Sep 4, 2020 and Sep 30, 2021 — classified as 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Fragmentation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Fragmentation3
- Detachment of Device or Device Component2
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2021Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH SZ378R - ENNOVATE MIS DOWNTUBE SHORT. ACCORDING TO THE COMPLAINT DESCRIPTION, THE RETAINING TABS ARE BROKEN. THE BROKEN PIECES WERE RETRIEVED & DISCARDED. AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. ADDITIONAL INFORMATION WAS NOT PROVIDED. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4). ASSOCIATED MEDWATCH-REPORTS: 9610612-2021-00360 (400510390 - SZ378R).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12555439 ↗InjuryApr 16, 2021Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO AESCULAP INC. THAT AN ENNOVATE MIS DOWNTUBE SHORT (PART # SZ378R) WAS USED DURING A MINIMALLY INVASIVE TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE SURGICAL PROCEDURE, THE ATTACHMENT MECHANISM OF THE DEVICE DISLODGED AND STUCK TO THE PEDICLE SCREW. AN X-RAY WAS PERFORMED TO CONFIRM THE PART LOCATION, AND THEN THE SURGEON RETRIEVED THE FRAGMENT FROM THE PATIENT. A SURGICAL DELAY OF APPROXIMATELY 10 MINUTES WAS REPORTED. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE INTERVENTION PERFORMED. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE XC (B)(4); CC (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11682152 ↗InjuryApr 7, 2021Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH SZ378R - ENNOVATE MIS DOWNTUBE SHORT. ACCORDING TO THE COMPLAINT DESCRIPTION, THE MECHANISM DISLODGED DURING MINIMALLY INVASIVE TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) SURGERY. ATTACHMENT MECHANISM OF THE DOWNTUBE DISLODGED AND STUCK TO THE PEDICLE SCREW, SURGEON REQUIRED ADDITIONAL INTERVENTION TO REMOVE THE PART (X-RAY AND PIECE RETRIEVAL). AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. THERE WAS A DELAY OF 10 MINUTES. ADDITIONAL INFORMATION WAS NOT PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED AS OF THIS REPORT. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11629030 ↗InjuryMar 12, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO AESCULAP INC. THAT A ENNOVATE MIS DOWNTUBE SHORT (PART # SZ378R) WAS USED DURING A MINIMALLY INVASIVE SPINE (MIS) FIXATION PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE PHYSICIAN ASSISTANT OVER ROTATED THE DOWN-TUBE REMOVAL KEY, WHICH LEFT THE TIP STILL SLIGHTLY IN THE SCREW. REPORTEDLY, THE TIP DETACHED, AFTER ROCKING AND TWISTING THE TUBE. THE LOOSE TIP WAS RETRIEVED FROM THE PROXIMAL AREA OF THE WOUND. THE WOUND WAS IRRIGATED, AND THEN VISUALLY INSPECTED. ADDITIONALLY, THE USER STATED THAT THE FIRST GENERATION DOWNTUBE REMOVAL KEY WAS USED. THE ACCOUNT HAS SINCE BEEN TRAINED AND SWITCHED OVER TO THE NEWER VERSION. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. ADDITIONAL INFORMATION WAS RECEIVED WHICH REVEALED THAT THE PROCEDURE WAS DELAYED APPROXIMATELY 5-7 MINUTES. THE PATIENT WAS UNDER ANESTHESIA. THE FINAL OUTCOME WAS NOT AFFECTED BY THE EVENT. THE ADVERSE EVENT / MALFUNCTION WAS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11474643 ↗InjuryMar 10, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
NO CHANGES REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11457272 ↗MalfunctionSep 4, 2020Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH A ENNOVATE MIS DOWNTUBE. ACCORDING TO THE COMPLAINT DESCRIPTION THE RETAINING LUG IS BROKEN OFF INTRAOPERATIVELY BUT DID NOT REMAIN IN SITU. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10495986 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page