Infusomat® 362911Recalls & reported problems

B. Braun Medical Inc. · Priming volume: 14mL, length: 84in, CARESITE®, Spin-Lock® · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046964837845), received between Apr 1, 2025 and Apr 1, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Air/Gas in Device” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2023Open, Classified

    In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

    Z-0355-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Air/Gas in Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 1, 2025Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: DESCRIBE THE EVENT OR PROBLEM (PLEASE PROVIDE AS MUCH DETAIL AS POSSIBLE): RN ENTERED THE ROOM AT 2030 TO BLINA PUMP BEEPING AIR IN LINE. THIS BUBBLE WAS AROUND 3 INCHES OF TUBING LING. THIS RN WAS ABLE TO FLICK THE AIR BUBBLES OUT AND RESTART THE INFUSION. AT 2045, THE PUMP BEGAN ALARMING AGAIN. UPON INSPECTING THE LINE THIS RN NOTICED THAT THE AIR BUBBLE WAS SIGNIFICANTLY LARGE STRETCHING FROM THE CASSETTE THAT CLIPS INTO THE PUMP TO THE VALVE/Y-SITE ABOVE THE PUMP. THIS RM NOTIFIED THE CHARGE NURSES, H2 PHARMACY, AND THE PRACTITIONER. AFTER MANY ATTEMPTS TO GET RID OF THE AIR AND SWITCHING THE PUMP OUT, THE DECISION WAS MADE TO HANG A NEW BAG OF BLINA. UNSURE IF THIS EVENT WAS RELATED TO BBRAUN TUBING/PUMP OF CURLIN TUBING (YELLOW TAB OF CURLIN BROKE OFF PER DAYSHIFT RN). THIS RN HUNG THE NEW BAG OF BLINA WITH ANOTHER RN AT 2109. TUBING AND PUMP SAVED ON CL OFFICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21741450 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates