Skip to content

DIPLOMAT TulipRecalls & reported problems

SIGNUS Medizintechnik GmbH · The modular DIPLOMAT® pedicle screw system is a rod-screw system that firmly connects two or more screws via the screw head and a rod using a force-fit or form-fit connection. Additional fixation elements, such as cross bars, hooks, connectors and washers, can also be used. The DIPLOMAT® pedicle screw system is applied temporarily to provide internal posterior stabilization until bone fusion in the lumbar or thoracic spine has taken place. The surgeon makes the final decision as to the indwelling time of the implant in the patient‘s body and the time when the implant is to be explanted. The pedicle screw system is implanted from a posterior approach. The implant is screwed into the vertebral body via the pedicle and it can be applied in a single segment or multiple segments. Implantation is facilitated by use of the specially developed accessories for inserting and positioning the implants. Only these accessories ensure safe use. Our product information provides further system-related information on the surgical method. If there are any preoperative uncertainties relating to the implant, information must be obtained from SIGNUS. Instruments specially developed by SIGNUS to ensure safe application are available for use with the implant system. · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04047844020043), received between Mar 14, 2024 and Mar 14, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 14, 2024Fracture; Fracture

    THE PATIENT WAS OPERATED WITH DIPLOMAT PEDICLE SCREW SYSTEM (B)(6) 2022, TWO LEVEL SURGERY L4- L5- S1. HE NOTICED SOME NOISE IN HIS BACK. A CHECK REVEALED BROKEN PEDICLE SCREWS WHICH WERE SURGICALLY REMOVED (B)(6) 2023. FRACTURE OF IMPLANT COMPONENTS IS A KNOWN RISK INDICATED IN THE IFU.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18901899 ↗

Other Orthosis, Spondylolisthesis Spinal Fixation devices

All Orthosis, Spondylolisthesis Spinal Fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page