Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04048438006337), received between Mar 21, 2024 and Mar 22, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Positioning Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Positioning Problem4
- Leak/Splash2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 22, 2024Positioning Problem; Positioning Problem
IT WAS REPORTED THAT TWO USED CASSETTE HOSE SETS FROM TWO DIFFERENT BATCHES WERE LEAKING AT THE PRESSURE DIAPHRAGM AFTER INSERTION INTO THE ENDOMAT SELECT UP210 PUMP. THE RINSING FLUID ESCAPED VIA THE DEFECTIVE MEMBRANE AFTER THE PUMP WAS STARTED UP AND FLOODED THE EQUIPMENT TRAILER WITH THE EQUIPMENT UNDERNEATH. NO HARM TO PATIENT, USER OR THIRD REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18960726 ↗MalfunctionMar 22, 2024Leak/Splash; Positioning Problem; Leak/Splash; Positioning Problem
IT WAS REPORTED THAT TWO USED CASSETTE HOSE SETS FROM TWO DIFFERENT BATCHES WERE LEAKING AT THE PRESSURE DIAPHRAGM AFTER INSERTION INTO THE ENDOMAT SELECT UP210 PUMP. THE RINSING FLUID ESCAPED VIA THE DEFECTIVE MEMBRANE AFTER THE PUMP WAS STARTED UP AND FLOODED THE EQUIPMENT TRAILER WITH THE EQUIPMENT UNDERNEATH. NO HARM TO PATIENT, USER OR THIRD REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18960720 ↗MalfunctionMar 21, 2024Leak/Splash; Positioning Problem; Leak/Splash; Positioning Problem
IT WAS REPORTED THAT TWO USED CASSETTE HOSE SETS FROM TWO DIFFERENT BATCHES WERE LEAKING AT THE PRESSURE DIAPHRAGM AFTER INSERTION INTO THE ENDOMAT SELECT UP210 PUMP. THE RINSING FLUID ESCAPED VIA THE DEFECTIVE MEMBRANE AFTER THE PUMP WAS STARTED UP AND FLOODED THE EQUIPMENT TRAILER WITH THE EQUIPMENT UNDERNEATH. NO HARM TO PATIENT, USER OR THIRD REPORTED. THE RISK IS COVERED IN THE RELATED RISK FILE (SAP-ID: (B)(6) ; RISK-ID: 1.1.05; SEVERITY-LEVEL: 4 (HARM: ELECTRIC SHOCK); PROBABILITY-LEVEL: 1 (IMPROBABLE)). IN ADDITION, NO EVENT (TIMEFRAME: BETWEEN 2022-03-05 AND 2024-03-19) THAT COULD BE CONSIDERED AS SIMILAR TO THIS ONE AND CAUSED OR CONTRIBUTED TO A DEATH OR SERIOUS INJURY COULD BE FOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18952063 ↗InjuryMar 21, 2024Positioning Problem; Positioning Problem
IT WAS REPORTED THAT TWO USED CASSETTE HOSE SETS FROM TWO DIFFERENT BATCHES WERE LEAKING AT THE PRESSURE DIAPHRAGM AFTER INSERTION INTO THE ENDOMAT SELECT UP210 PUMP. THE RINSING FLUID ESCAPED VIA THE DEFECTIVE MEMBRANE AFTER THE PUMP WAS STARTED UP AND FLOODED THE EQUIPMENT TRAILER WITH THE EQUIPMENT UNDERNEATH. NO HARM TO PATIENT, USER OR THIRD REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18952059 ↗
Other TUBE, ASPIRATING, FLEXIBLE, CONNECTING devices
- DeRoyal 71-C506Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall4 reports
- DeRoyal 71-C512Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall2 reports
- DeRoyal 71-C520Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall1 report
- DeRoyal 71-C501Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
- DeRoyal 71-C510Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
- DeRoyal 71-C5100NSDeroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
- DeRoyal 71-C6100NSDeroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
- DeRoyal 71-C7100NSDeroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page