Skip to content

n.a. 20133720-1Recalls & reported problems

Karl Storz GmbH & Co. KG · Cold Light Fountain D-LIGHT P SCB · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048551004685), received between Oct 3, 2019 and Mar 20, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Fire” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Fire1

Patient problems

  • Burn(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 20, 2023Fire; Fire

    CUSTOMER REPORTED THAT DURING A PROCEDURE, THE DEVICE CAUGHT ON FIRE; NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16573167 ↗
  • InjuryOct 3, 2019Adverse Event Without Identified Device or Use Problem

    ALLEGEDLY, DURING A LAP CHOLE PROCEDURE, WHILE THEY WERE USING FLUOROSCOPY, THE LAPAROSCOPE WAS PULLED BACK AND THE TIP OF THE SCOPE WAS RESTED INSIDE OF THE TROCAR CANNULA (THE SCOPE WAS CONNECTED TO AN ACTIVE LIGHT CABLE THAT WAS PLUGGED INTO THE D-LIGHT P HIGH POWER LIGHT SOURCE; THIS RESULTED IN THE PATIENT BEING BURNED AT THE UMBILICUS TROCAR ENTRANCE SITE. THE BURN WAS NOTICED AFTER THE PROCEDURE. WE HAVE BEEN UNABLE TO OBTAIN ANY FURTHER DETAILS ABOUT THE PATIENT OR THE EVENT.

    Patient problems: Burn(s)

    MAUDE report 9153838 ↗

FDA clearance and classification

The FDA regulates product code OWN (confocal optical imaging) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the n.a. 20133720-1

Has the n.a. 20133720-1 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04048551004685), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the n.a. 20133720-1?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Oct 3, 2019 and Mar 20, 2023. In those reports, the most common specific device problem is “Fire” (1); the most common patient problem is “Burn(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the n.a. 20133720-1?

None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the n.a. 20133720-1, and is it still sold?

The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand n.a., model 20133720-1). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Confocal optical imaging devices

All Confocal optical imaging devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page