n.a. 26050CARecalls & reported problems

Karl Storz GmbH & Co. KG · Inner sheath for Resectoscope · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04048551007228), received between Sep 1, 2023 and Sep 28, 2026 — classified as 5 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Material Separation2
  • Unintended Electrical Shock2
  • Device Handling Problem1
  • Electrical Shorting1
  • Fluid/Blood Leak1
  • Fracture1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information3
  • Convulsion/Seizure2
  • Electric Shock2
  • Internal Organ Perforation1
  • Urethral Stenosis/Stricture1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Electrical Shorting

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924034 ↗
  • MalfunctionJul 27, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50017291 ↗
  • InjuryJul 10, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT AT THE END OF THE PROCEDURE, THE SURGEON OBSERVED THAT THE CERAMIC TIP OF THE INNER SHEATH OF THE RESECTOSCOPE HAD DETACHED AND WAS LOCATED IN THE BLADDER. DUE TO SIGNIFICANT URETHRAL STENOSIS, EXTRACTION VIA THE NATURAL ROUTE PROVED IMPOSSIBLE. AN ATTEMPT WAS MADE TO FRAGMENT IT USING A 30W HOLMIUM LASER, ALLOWING PARTIAL EXTRACTION OF A FRAGMENT; HOWEVER, A RESIDUAL FRAGMENT REMAINED IN PLACE, INACCESSIBLE VIA THE TRANSURETHRAL ROUTE. ADDITIONALLY, PERFORATION OF THE URETHRAL WALL BY THE CERAMIC TIP LED TO SIGNIFICANT EXTRAVASATION OF SALINE INTO THE RETROPERITONEUM, EVIDENCED BY A DISCREPANCY BETWEEN THE IRRIGATED VOLUME AND THE ASPIRATED VOLUME. GIVEN THE IMPOSSIBILITY OF EXTRACTING THE REMAINING FRAGMENT ENDOSCOPICALLY, THE DECISION WAS MADE TO CONVERT TO LAPAROSCOPY. HOWEVER, EXTRACTION PROVED PARTICULARLY DIFFICULT DUE TO THE RETROVESICAL LOCATION OF THE FRAGMENT. THE PATIENT HAD TO BE TRANSFERRED TO INTENSIVE CARE AFTER EXTRACTION OF THE REMAINING FRAGMENT.

    Patient problems: Insufficient Information; Internal Organ Perforation; Urethral Stenosis/Stricture

    MAUDE report 25782708 ↗
  • MalfunctionOct 15, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE SHEATH CERAMIC BROKE WHILST IN USE ON PATIENT, DEBRIS REMOVED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23280740 ↗
  • MalfunctionJun 26, 2025Loss of or Failure to Bond; Detachment of Device or Device Component; Loss of or Failure to Bond; Detachment of Device or Device Component

    THE CUSTOMER DID NOT PROVIDE ANY ERROR DESCRIPTION. DURING INITIAL EVALUATION OF THE PRODUCT (26050CA INNER SHEATH FOR RESECTOSCOPE) IT WAS FOUND THAT THERE IS A LOOSE ADHESIVE BOND AT THE CERAMIC BEAK AND THE BEAK HAS COME OFF FROM THE SHEATH. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22340759 ↗
  • InjuryJan 31, 2025Leak/Splash; Unintended Electrical Shock; Leak/Splash; Unintended Electrical Shock

    IT WAS REPORTED THAT LEAKAGE OCCURRED FOR THE THIRD TIME IN THE USE OF ELECTRIC ENDOSCOPIC INSTRUMENTS IN THE OPERATION BY USER, AND SHOCKED THE PATIENT, BUT ONLY CONVULSED DURING THE OPERATION, CAUSING NO SERIOUS INJURY, AND THERE WAS NO INTERVENTION TREATMENT IN THE LATER STAGE. IT WAS CONFIRMED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WAS NO FURTHER IMPACT OR CONSEQUENCES. HOWEVER, DUE TO THE REPORTED SHOCK AND CONVULSION OF THE PATIENT THIS CASE IS DEEMED REPORTABLE FOR PRECAUTIONARY REASONS.

    Patient problems: Convulsion/Seizure; Electric Shock

    MAUDE report 21287065 ↗
  • InjuryJan 31, 2025Unintended Electrical Shock; Unintended Electrical Shock

    IT WAS REPORTED THAT LEAKAGE OCCURRED FOR THE THIRD TIME IN THE USE OF ELECTRIC ENDOSCOPIC INSTRUMENTS IN THE OPERATION BY USER, AND SHOCKED THE PATIENT, BUT ONLY CONVULSED DURING THE OPERATION, CAUSING NO SERIOUS INJURY, AND THERE WAS NO INTERVENTION TREATMENT IN THE LATER STAGE. IT WAS CONFIRMED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WAS NO FURTHER IMPACT OR CONSEQUENCES. HOWEVER, DUE TO THE REPORTED SHOCK AND CONVULSION OF THE PATIENT THIS CASE IS DEEMED REPORTABLE FOR PRECAUTIONARY REASONS.

    Patient problems: Convulsion/Seizure; Electric Shock

    MAUDE report 21287062 ↗
  • InjurySep 1, 2023Material Separation; Device Handling Problem; Material Separation; Device Handling Problem

    IT WAS REPORTED THAT DURING A LASER AURIGA 30 FRAGMENTATION OF A LARGE BLADDER STONE WITH THE STORZ RESECTOR, THE CERAMIC TIP BROKE OFF AND PASSED INTO THE BLADDER. THIS FORCED THE SURGEON TO DO A LAPAROTOMY TO RECOVER THE PART INSIDE THE LAVA. THE CERAMIC COMPLETELY SEPARATED FROM THE SHEATH AND PASSED INTO THE BLADDER. THE SURGERY WAS PROLONGED FOR 1H 30 MINUTES. THE RESECTION OF THE PROSTATE WAS RESCHEDULED 15 DAYS LATER.

    Patient problems: Insufficient Information

    MAUDE report 17671662 ↗
  • InjurySep 1, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT DURING A LASER AURIGA 30 FRAGMENTATION OF A LARGE BLADDER STONE WITH THE STORZ RESECTOR, THE CERAMIC TIP BROKE OFF AND PASSED INTO THE BLADDER. THIS FORCED THE SURGEON TO DO A LAPAROTOMY TO RECOVER THE PART INSIDE THE LAVA. THE CERAMIC COMPLETELY SEPARATED FROM THE SHEATH AND PASSED INTO THE BLADDER. THE SURGERY WAS PROLONGED FOR 1H 30 MINUTES. THE RESECTION OF THE PROSTATE WAS RESCHEDULED 15 DAYS LATER.

    Patient problems: Insufficient Information

    MAUDE report 17671661 ↗

Other HYSTEROSCOPE (AND ACCESSORIES) devices

All HYSTEROSCOPE (AND ACCESSORIES) devices →