Skip to content

Optitherm 20432030Recalls & reported problems

Karl Storz GmbH & Co. KG · Optitherm® Heating Element · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04048551068663), received between Oct 15, 2021 and Oct 6, 2023 — classified as 3 injuries.

The most frequently reported specific device problem is “Temperature Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Temperature Problem2
  • Excessive Heating1
  • Overheating of Device1

Patient problems

  • Burn(s)3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 6, 2023Temperature Problem; Excessive Heating; Temperature Problem; Excessive Heating

    IT WAS REPORTED THAT DURING A SURGERY, THE HEATING ELEMENT (OPTITHERM 20432030) BURNED THE PATIENT'S SKIN. THE HEATING ELEMENT WAS CONNECTED TO THE THERMOFLATOR AND PROBABLY OVERHEATED.

    Patient problems: Burn(s)

    MAUDE report 17887261 ↗
  • InjuryOct 6, 2023Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT DURING A SURGERY, THE HEATING ELEMENT (OPTITHERM 20432030) BURNED THE PATIENT'S SKIN. THE HEATING ELEMENT WAS CONNECTED TO THE THERMOFLATOR AND PROBABLY OVERHEATED.

    Patient problems: Burn(s)

    MAUDE report 17886674 ↗
  • InjuryOct 15, 2021Overheating of Device; Overheating of Device

    AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(6): THE "OPTITHERM" ACCESSORY, WHICH WAS INTENDED FOR PREHEATING THE GAS TO BE INSUFFLATED HAS OVERHEATED FAR BEYOND THE INTENDED ALARM LIMIT VALUE OF 41 °C. THIS SITUATION WAS NOT DETECTED BY THE MAIN UNIT. THIS CIRCUMSTANCE WAS NOT DETECTED BY THE MAIN UNIT, SO NO ALARM WAS TRIGGERED. DUE TO THE DIRECT SKIN CONTACT, A LARGE BURN CORRESPONDING TO THE SIZE OF THE HOUSING BURN OCCURRED. TWO DEVICES WERE INVOLVED (OPTITHERM AND THERMOFLATOR). THIS REPORT IS FOR DEVICE 1 (OPTITHERM). (MFR'S. INTERNAL COMPLAINT (B)(4)).

    Patient problems: Burn(s)

    MAUDE report 12642687 ↗

Other INSUFFLATOR, LAPAROSCOPIC devices

All INSUFFLATOR, LAPAROSCOPIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page