Skip to content

ENDOFLATOR 26430520-1Recalls & reported problems

Karl Storz GmbH & Co. KG · Electronic CO2 ENDOFLATOR® SCB · Class 2

All models of ENDOFLATOR (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04048551068793), received between Mar 11, 2019 and Jan 28, 2026 — classified as 8 injuries.

The most frequently reported specific device problem is “Gas Output Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Gas Output Problem2
  • Gradient Increase2
  • Insufficient Device Problem Information2
  • Gas/Air Leak1
  • Mechanical Problem1

Patient problems

  • Cardiac Arrest2
  • No Clinical Signs, Symptoms or Conditions2
  • Pulmonary Emphysema2
  • Skin Disorders2
  • Insufficient Information1
  • Pneumothorax1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 28, 2026Gas Output Problem

    IT WAS REPORTED THAT THE INSUFFLATOR STOPPED DELIVERING GAS, PREVENTING THE ESTABLISHMENT OF PNEUMOPERITONEUM. THE ISSUE WAS RESOLVED AFTER APPROXIMATELY 20 MINUTES, RESULTING IN PROLONGED ANESTHESIA AND SURGICAL DURATION FOR THE PATIENT. SINCE THE ANESTHESIA WAS PROLONGED, THE CASE IS CONSIDERED REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24201754 ↗
  • InjuryJan 28, 2026Gas Output Problem

    IT WAS REPORTED THAT THE INSUFFLATOR STOPPED DELIVERING GAS, PREVENTING THE ESTABLISHMENT OF PNEUMOPERITONEUM. THE ISSUE WAS RESOLVED AFTER APPROXIMATELY 20 MINUTES, RESULTING IN PROLONGED ANESTHESIA AND SURGICAL DURATION FOR THE PATIENT. SINCE THE ANESTHESIA WAS PROLONGED, THE CASE IS CONSIDERED REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24201753 ↗
  • InjuryJan 9, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENTS DEVELOPED PULMONARY EMPHYSEMA DURING TWO CONSECUTIVE OPERATIONS WITH AN ENDOFLATOR. SINCE THE PATIENT SUFFERED PERMANENT DAMAGE, THE CASE IS DEEMED AS REPORTABLE.

    Patient problems: Pulmonary Emphysema; Skin Disorders

    MAUDE report 21114578 ↗
  • InjuryJan 9, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENTS DEVELOPED PULMONARY EMPHYSEMA DURING TWO CONSECUTIVE OPERATIONS WITH AN ENDOFLATOR. SINCE THE PATIENT SUFFERED PERMANENT DAMAGE, THE CASE IS DEEMED AS REPORTABLE.

    Patient problems: Pulmonary Emphysema; Skin Disorders

    MAUDE report 21113920 ↗
  • InjuryOct 24, 2024Gradient Increase; Gradient Increase

    IT WAS REPORTED THAT FOUR TO FIVE MINUTES AFTER THE START OF THE LAPAROSCOPIC PROSTATECTOMY PROCEDURE A SUDDEN INCREASE IN INSUFFLATION PRESSURE OCCURRED AND WENT UP TO 28MMHG WHILE THE DEVICE WAS SET AT 12MMHG. THE PATIENT HAD A CARDIAC ARREST AND CAME BACK AFTER RESUSCITATION TREATMENT OF 5-10 MINUTES. AFTER THE PATIENT WAS STABILIZED, THE MEDICAL PROCEDURE CONTINUED AND HAD BEEN SUCCESSFULLY COMPLETED.

    Patient problems: Cardiac Arrest

    MAUDE report 20529474 ↗
  • InjuryOct 24, 2024Gradient Increase; Mechanical Problem; Gradient Increase; Mechanical Problem

    IT WAS REPORTED THAT FOUR TO FIVE MINUTES AFTER THE START OF THE LAPAROSCOPIC PROSTATECTOMY PROCEDURE A SUDDEN INCREASE IN INSUFFLATION PRESSURE OCCURRED AND WENT UP TO 28MMHG WHILE THE DEVICE WAS SET AT 12MMHG. THE PATIENT HAD A CARDIAC ARREST AND CAME BACK AFTER RESUSCITATION TREATMENT OF 5-10 MINUTES. AFTER THE PATIENT WAS STABILIZED, THE MEDICAL PROCEDURE CONTINUED AND HAD BEEN SUCCESSFULLY COMPLETED.

    Patient problems: Cardiac Arrest

    MAUDE report 20528686 ↗
  • InjuryNov 5, 2020Adverse Event Without Identified Device or Use Problem

    AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN GERMANY: FOLLOWING INSUFFLATION, CARDIAC OF THE PATIENT (LIVING PATIENT). USE OF THE SAME DEVICE IN THE AFTERNOON, CARDIAC ARREST AVOIDED.

    Patient problems: Insufficient Information

    MAUDE report 10793585 ↗
  • InjuryMar 11, 2019Gas/Air Leak; Gas/Air Leak

    ALLEGEDLY, DURING A GASTRIC SLEEVE PROCEDURE, THE PATIENT DEVELOPED A PNEUMOTHORAX.

    Patient problems: Pneumothorax

    MAUDE report 8410539 ↗

FDA clearance and classification

The FDA regulates product code HIF (insufflator, laparoscopic) under 21 CFR 884.1730 as a Class II device. 21 CFR 884.1730 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the ENDOFLATOR 26430520-1

Has the ENDOFLATOR 26430520-1 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04048551068793), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ENDOFLATOR 26430520-1?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Mar 11, 2019 and Jan 28, 2026. In those reports, the most common specific device problems are “Gas Output Problem” (2), “Gradient Increase” (2) and “Gas/Air Leak” (1); the most common patient problems are “Cardiac Arrest” (2), “Pulmonary Emphysema” (2) and “Skin Disorders” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ENDOFLATOR 26430520-1?

None of the reports are classified as deaths, 8 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the ENDOFLATOR 26430520-1, and is it still sold?

The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand ENDOFLATOR, model 26430520-1). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other INSUFFLATOR, LAPAROSCOPIC devices

All INSUFFLATOR, LAPAROSCOPIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page