Skip to content

THERMOFLATOR 26432020-1Recalls & reported problems

Karl Storz GmbH & Co. KG · THERMOFLATOR® SCB · Class 2

Adverse event reports
17
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 04048551069264), received between Apr 7, 2017 and Jul 14, 2026 — classified as 9 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Improper Flow or Infusion” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event7
  • Improper Flow or Infusion2
  • Inflation Problem2
  • Material Separation2
  • Pressure Problem2
  • Device Operates Differently Than Expected1
  • Insufficient Device Problem Information1
  • No Flow1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient2
  • Pulmonary Emphysema2
  • Cardiac Arrest1
  • Hematoma1
  • Laceration(s)1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 14, 2026Improper Flow or Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25817555 ↗
  • MalfunctionJul 14, 2026Pressure Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25817554 ↗
  • MalfunctionMay 22, 2026Pressure Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC PROCEDURE, THE INSUFFLATOR EXHIBITED ABNORMALLY HIGH OR LOW INFLATION PRESSURE AND FAILED TO AUTOMATICALLY INFLATE OR DEFLATE GAS ACCORDING TO THE SET PRESSURE. DUE TO INACCURATE INFLATION PRESSURE, A BACKUP DEVICE WAS USED AS A REPLACEMENT, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25260834 ↗
  • MalfunctionApr 9, 2026No Flow

    IT WAS REPORTED THAT ON (B)(6) 2025, IN THE OPERATING ROOM, THE DOCTOR USED THE INSUFFLATORTO EXPAND THE SURGICAL FIELD AND FOUND THAT THE INSUFFLATOR FAILED TO SUPPLY GAS, AFFECTING THEOPERATION. THE TECHNICAL STAFF WAS IMMEDIATELY CONTACTED FOR INSPECTION, AND THE EQUIPMENT WAS REPLACED. AFTER REPLACEMENT, THE OPERATION PROCEEDED NORMALLY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24835205 ↗
  • MalfunctionFeb 13, 2026Improper Flow or Infusion; No Apparent Adverse Event

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN ELECTIVE THORACOSCOPIC. A STORZ INSUFFLATOR WAS USED AT THE START OF THE SURGERY, AT WHICH POINT IT WAS FOUND THAT THE PNEUMOPERITONEUM COULD NOT BE MAINTAINED IN THE PATIENT, RESULTING IN A POOR SURGICAL FIELD OF VIEW. AFTER RULING OUT INSUFFICIENT CO2 SUPPLY FROM THE CENTRAL GAS SUPPLY SYSTEM, THE INSUFFLATOR WAS DETERMINED TO MALFUNCTION. THE INSUFFLATOR WAS REPLACED, AFTER WHICH IT FUNCTIONED NORMALLY; THE ENTIRE PROCESS TOOK APPROXIMATELY 20 MINUTES, AND THE SURGERY WAS SUCCESSFULLY COMPLETED IN THE END. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24338597 ↗
  • InjuryApr 20, 2023Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THERE WERE INSUFFLATION PROBLEMS DURING SURGERY AND PRESSURE CHECK WAS NEEDED. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE PATIENT SUFFERED A PULMONARY EMPHYSEMA AND NEEDED ADDITIONAL HOSPITALIZATION AFTER SURGERY.

    Patient problems: Pulmonary Emphysema

    MAUDE report 16776918 ↗
  • MalfunctionApr 20, 2023Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THERE WERE INSUFFLATION PROBLEMS DURING SURGERY AND PRESSURE CHECK WAS NEEDED. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE PATIENT SUFFERED A PULMONARY EMPHYSEMA AND NEEDED ADDITIONAL HOSPITALIZATION AFTER SURGERY.

    Patient problems: Pulmonary Emphysema

    MAUDE report 16776917 ↗
  • InjuryJan 19, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(6) : DURING INSUFFLATION, THE PATIENT HAD A CARDIAC ARREST. THE CUSTOMER REPORTED THAT THE PATIENT HAS RECOVERED. (MFR'S. INTERNAL COMPLAINT # (B)(4)).

    Patient problems: Cardiac Arrest

    MAUDE report 13289321 ↗
  • InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event

    THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6), (B)(6) AND (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12872802 ↗
  • InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event

    THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR 15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO STATED THAT THEY A

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12872798 ↗
  • InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event

    THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6), (B)(6) AND (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12870742 ↗
  • InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event

    THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO STATED THAT THEY AR

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12870740 ↗
  • InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event

    THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR15. SURGEON IN ALL 3 CASES WAS DR. (B)(6) . CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO STATED THAT THEY A

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12870738 ↗
  • InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event

    THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR 15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO STATED THAT THEY A

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12870730 ↗
  • MalfunctionJun 23, 2017Material Separation; Material Separation

    ALLEGEDLY, DURING A LAPAROSCOPIC MYOMECTOMY PROCEDURE THE DOCTOR NOTED THAT BOTH JAWS BROKE OFF THE INSTRUMENT AND FELL INTO THE PATIENT. THE DOCTOR USED ANOTHER BOWEL GRASPER FROM ANOTHER SET AND ONE OF THE JAWS BROKE OFF AS WELL. THE DOCTOR IMMEDIATELY RETRIEVED THE MISSING PIECES USING A STERILE MAGNET. NOTE: THIS REPORT PERTAINS TO THE SECOND INSTRUMENT THAT BROKE; A SEPARATE REPORT WAS FILED FOR THE FIRST INSTRUMENT THAT BROKE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6665621 ↗
  • MalfunctionJun 23, 2017Material Separation; Material Separation

    ALLEGEDLY, DURING A LAPAROSCOPIC MYOMECTOMY PROCEDURE THE DOCTOR NOTED THAT BOTH JAWS BROKE OFF THE INSTRUMENT AND FELL INTO THE PATIENT. THE DOCTOR USED ANOTHER BOWEL GRASPER FROM ANOTHER SET AND ONE OF THE JAWS BROKE OFF AS WELL. THE DOCTOR IMMEDIATELY RETRIEVED THE MISSING PIECES USING A STERILE MAGNET. NOTE: THIS REPORT PERTAINS TO THE FIRST INSTRUMENT THAT BROKE; AS SEPARATE REPORT WILL BE FILED FOR THE SECOND INSTRUMENT THAT BROKE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6665615 ↗
  • InjuryApr 7, 2017Device Operates Differently Than Expected

    ALLEGEDLY, THE PATIENT UNDERWENT A LAPAROSCOPIC PROCEDURE AND WHILE BEING INSUFFLATED, THE HOSPITAL CLAIMS THE THERMOFLATOR UNIT SOMEHOW CREATED A VACUUM SITUATION, CAUSING THE VERESS NEEDLE TO PERFORATE/LACERATE A BLOOD VESSEL, CREATING A HEMATOMA WHICH THEY WERE ABLE TO EVACUATE. THE PATIENT WAS ADMITTED OVERNIGHT AND WAS DISCHARGED THE FOLLOWING DAY.

    Patient problems: Hematoma; Laceration(s)

    MAUDE report 6473587 ↗

Other INSUFFLATOR, LAPAROSCOPIC devices

All INSUFFLATOR, LAPAROSCOPIC devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.