THERMOFLATOR 26432020-1Recalls & reported problems
Karl Storz GmbH & Co. KG · THERMOFLATOR® SCB · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 04048551069264), received between Apr 7, 2017 and Jul 14, 2026 — classified as 9 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Improper Flow or Infusion” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- No Apparent Adverse Event7
- Improper Flow or Infusion2
- Inflation Problem2
- Material Separation2
- Pressure Problem2
- Device Operates Differently Than Expected1
- Insufficient Device Problem Information1
- No Flow1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
- Pulmonary Emphysema2
- Cardiac Arrest1
- Hematoma1
- Laceration(s)1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 14, 2026Improper Flow or Infusion
A CORRECTED EVENT DESCRIPTION WAS RECEIVED ON (b)(6) 2026. IT WAS REPORTED THAT DURING THE PATIENT'S CYSTOSCOPIC RESECTION, ABNORMAL FLOW RATE OF THE INSUFFLATOR CAUSED RAPID ABDOMINAL DISTENSION. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25817555 ↗MalfunctionJul 14, 2026Pressure Problem
IT WAS REPORTED THAT ABNORMAL CARBON DIOXIDE PRESSURE WAS FOUND DURING EQUIPMENT USE ON (b)(6) 2026. INSPECTION OF THE CARBON DIOXIDE PIPELINE REVEALED AGING AND DAMAGE. NORMAL OPERATION WAS RESTORED AFTER REPLACEMENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25817554 ↗MalfunctionMay 22, 2026Pressure Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC PROCEDURE, THE INSUFFLATOR EXHIBITED ABNORMALLY HIGH OR LOW INFLATION PRESSURE AND FAILED TO AUTOMATICALLY INFLATE OR DEFLATE GAS ACCORDING TO THE SET PRESSURE. DUE TO INACCURATE INFLATION PRESSURE, A BACKUP DEVICE WAS USED AS A REPLACEMENT, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25260834 ↗MalfunctionApr 9, 2026No Flow
IT WAS REPORTED THAT ON (B)(6) 2025, IN THE OPERATING ROOM, THE DOCTOR USED THE INSUFFLATORTO EXPAND THE SURGICAL FIELD AND FOUND THAT THE INSUFFLATOR FAILED TO SUPPLY GAS, AFFECTING THEOPERATION. THE TECHNICAL STAFF WAS IMMEDIATELY CONTACTED FOR INSPECTION, AND THE EQUIPMENT WAS REPLACED. AFTER REPLACEMENT, THE OPERATION PROCEEDED NORMALLY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24835205 ↗MalfunctionFeb 13, 2026Improper Flow or Infusion; No Apparent Adverse Event
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN ELECTIVE THORACOSCOPIC. A STORZ INSUFFLATOR WAS USED AT THE START OF THE SURGERY, AT WHICH POINT IT WAS FOUND THAT THE PNEUMOPERITONEUM COULD NOT BE MAINTAINED IN THE PATIENT, RESULTING IN A POOR SURGICAL FIELD OF VIEW. AFTER RULING OUT INSUFFICIENT CO2 SUPPLY FROM THE CENTRAL GAS SUPPLY SYSTEM, THE INSUFFLATOR WAS DETERMINED TO MALFUNCTION. THE INSUFFLATOR WAS REPLACED, AFTER WHICH IT FUNCTIONED NORMALLY; THE ENTIRE PROCESS TOOK APPROXIMATELY 20 MINUTES, AND THE SURGERY WAS SUCCESSFULLY COMPLETED IN THE END. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24338597 ↗InjuryApr 20, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THERE WERE INSUFFLATION PROBLEMS DURING SURGERY AND PRESSURE CHECK WAS NEEDED. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE PATIENT SUFFERED A PULMONARY EMPHYSEMA AND NEEDED ADDITIONAL HOSPITALIZATION AFTER SURGERY.
Patient problems: Pulmonary Emphysema
MAUDE report 16776918 ↗MalfunctionApr 20, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THERE WERE INSUFFLATION PROBLEMS DURING SURGERY AND PRESSURE CHECK WAS NEEDED. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE PATIENT SUFFERED A PULMONARY EMPHYSEMA AND NEEDED ADDITIONAL HOSPITALIZATION AFTER SURGERY.
Patient problems: Pulmonary Emphysema
MAUDE report 16776917 ↗InjuryJan 19, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(6) : DURING INSUFFLATION, THE PATIENT HAD A CARDIAC ARREST. THE CUSTOMER REPORTED THAT THE PATIENT HAS RECOVERED. (MFR'S. INTERNAL COMPLAINT # (B)(4)).
Patient problems: Cardiac Arrest
MAUDE report 13289321 ↗InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event
THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6), (B)(6) AND (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12872802 ↗InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event
THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR 15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO STATED THAT THEY A
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12872798 ↗InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event
THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6), (B)(6) AND (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12870742 ↗InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event
THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO STATED THAT THEY AR
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12870740 ↗InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event
THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR15. SURGEON IN ALL 3 CASES WAS DR. (B)(6) . CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO STATED THAT THEY A
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12870738 ↗InjuryNov 24, 2021No Apparent Adverse Event; No Apparent Adverse Event
THE INITIAL COMPLAINT REPORT NOTED THE FOLLOWING: OUR THERMOFLATOR UNIT WAS USED FOR VEIN HARVESTING PROCEDURES ON (B)(6) OF 2021 (AND POTENTIALLY TWO MORE). THE UNIT IS IN OR 15. SURGEON IN ALL 3 CASES WAS DR. (B)(6). CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION ON THE INCIDENT. HOWEVER, PER THE CUSTOMER, THE INSUFFLATOR IS WORKING ACCORDING TO THE EQUIPMENT'S SPECIFICATIONS. THE ISSUE THAT THEY HAVE IDENTIFIED IS A POTENTIAL USER ERROR WITH THE INSUFFLATOR. THE USER HAD THE PRESSURE AND THE FLOW SETTINGS MUCH HIGHER THAN HE SHOULD HAVE DURING THE ENDOSCOPIC VEIN HARVESTING. PER OUR COMPLAINT INVESTIGATION, THE CUSTOMER STATED: AFTER ENDOSCOPIC VEIN HARVESTING STARTED, THE PATIENT'S END TIDAL CO2 STARTED TO RISE. ANESTHESIA MACHINE SHOWED INCREASE IN ETCO2. PATIENT'S BLOOD GAS SHOWED AN INCREASE IN PCO2. EFFORTS WERE MADE TO CORRECT THE CO2, BUT IT DID NOT DECREASE UNTIL ABOUT 1.5 HOUR WHEN ENDOSCOPIC VEIN HARVESTING WAS FINISHING. ANESTHESIA MACHINE SETTINGS WERE CHANGED IN ORDER TO LOWER ETCO2 (INCREASE RESPIRATORY RATE AND CHANGED OUT ALL HOSES AND FILTERS ON MACHINE). INTERVENTIONS DIDN'T HELP INITIALLY. NO PATIENT IMPACT OR INJURY IDENTIFIED. CUSTOMER ALSO STATED THAT THEY A
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12870730 ↗MalfunctionJun 23, 2017Material Separation; Material Separation
ALLEGEDLY, DURING A LAPAROSCOPIC MYOMECTOMY PROCEDURE THE DOCTOR NOTED THAT BOTH JAWS BROKE OFF THE INSTRUMENT AND FELL INTO THE PATIENT. THE DOCTOR USED ANOTHER BOWEL GRASPER FROM ANOTHER SET AND ONE OF THE JAWS BROKE OFF AS WELL. THE DOCTOR IMMEDIATELY RETRIEVED THE MISSING PIECES USING A STERILE MAGNET. NOTE: THIS REPORT PERTAINS TO THE SECOND INSTRUMENT THAT BROKE; A SEPARATE REPORT WAS FILED FOR THE FIRST INSTRUMENT THAT BROKE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6665621 ↗MalfunctionJun 23, 2017Material Separation; Material Separation
ALLEGEDLY, DURING A LAPAROSCOPIC MYOMECTOMY PROCEDURE THE DOCTOR NOTED THAT BOTH JAWS BROKE OFF THE INSTRUMENT AND FELL INTO THE PATIENT. THE DOCTOR USED ANOTHER BOWEL GRASPER FROM ANOTHER SET AND ONE OF THE JAWS BROKE OFF AS WELL. THE DOCTOR IMMEDIATELY RETRIEVED THE MISSING PIECES USING A STERILE MAGNET. NOTE: THIS REPORT PERTAINS TO THE FIRST INSTRUMENT THAT BROKE; AS SEPARATE REPORT WILL BE FILED FOR THE SECOND INSTRUMENT THAT BROKE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6665615 ↗InjuryApr 7, 2017Device Operates Differently Than Expected
ALLEGEDLY, THE PATIENT UNDERWENT A LAPAROSCOPIC PROCEDURE AND WHILE BEING INSUFFLATED, THE HOSPITAL CLAIMS THE THERMOFLATOR UNIT SOMEHOW CREATED A VACUUM SITUATION, CAUSING THE VERESS NEEDLE TO PERFORATE/LACERATE A BLOOD VESSEL, CREATING A HEMATOMA WHICH THEY WERE ABLE TO EVACUATE. THE PATIENT WAS ADMITTED OVERNIGHT AND WAS DISCHARGED THE FOLLOWING DAY.
Patient problems: Hematoma; Laceration(s)
MAUDE report 6473587 ↗
FDA clearance and classification
The FDA regulates product code HIF (insufflator, laparoscopic) under 21 CFR 884.1730 as a Class II device. 21 CFR 884.1730 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the THERMOFLATOR 26432020-1
Has the THERMOFLATOR 26432020-1 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048551069264), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the THERMOFLATOR 26432020-1?
The FDA's MAUDE database lists 17 adverse event reports for this device, received between Apr 7, 2017 and Jul 14, 2026. In those reports, the most common specific device problems are “Improper Flow or Infusion” (2), “Inflation Problem” (2) and “Material Separation” (2); the most common patient problems are “Pulmonary Emphysema” (2), “Cardiac Arrest” (1) and “Hematoma” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the THERMOFLATOR 26432020-1?
None of the reports are classified as deaths, 9 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the THERMOFLATOR 26432020-1, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand THERMOFLATOR, model 26432020-1). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other INSUFFLATOR, LAPAROSCOPIC devices
- HIGH FLOW INSUFFLATION UNIT UHI-3Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC1 recall162 reports
- HIGH FLOW INSUFFLATION UNIT UHI-3Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC1 recall48 reports
- AIRSEAL AS-iFS1Conmed Corporation · Insufflator, laparoscopic4 deaths91 reports
- Pneumosure 0620040610Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 death79 reports
- Pneumosure 0620040690Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 recall22 reports
- Pneumosure 0620040660Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 recall18 reports
- CYLINDER HOSE (PIN) FOR UHI-3 MAJ-1080Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC58 reports
- Pneumosure 0620040680Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page