Skip to content

n.a. 27178ARecalls & reported problems

Karl Storz GmbH & Co. KG · Scissors, 7 Fr., 40 cm · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048551081419), received between Nov 5, 2024 and Jan 29, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem2
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 29, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE UROLOGIST WAS ATTEMPTING TO CUT A SUTURE ON A PREVIOUS UROLIFT IMPLANT ON (B)(6) 2024. A SNAP WAS HEARD AND A SMALL CRESCENT SHAPED PIECE OF THE MOVING PART ON THE SCISSOR BROKE OFF. IT WAS VERY SMALL AND THE UROLOGIST SPENT AT LEAST AN HOUR INSPECTING ALL AREAS OF THE BLADDER AND URETHRA TO LOCATE THE PIECE THAT BROKE OFF THE FLEXIBLE SCISSORS. HE USED THE 30* LENS, 70* LENS AND A FLEXIBLE CYSTOSCOPE TO THOROUGHLY INSPECT THE BLADDER AND URETHRA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21270383 ↗
  • InjuryNov 5, 2024Break; Mechanical Problem; Break; Mechanical Problem

    IT WAS REPORTED THAT THE UROLOGIST WAS ATTEMPTING TO CUT A SUTURE ON A PREVIOUS UROLIFT IMPLANT ON (B)(6) 2024. A SNAP WAS HEARD AND A SMALL CRESCENT SHAPED PIECE OF THE MOVING PART ON THE SCISSOR BROKE OFF. IT WAS VERY SMALL AND THE UROLOGIST SPENT AT LEAST AN HOUR INSPECTING ALL AREAS OF THE BLADDER AND URETHRA TO LOCATE THE PIECE THAT BROKE OFF THE FLEXIBLE SCISSORS. HE USED THE 30* LENS, 70* LENS AND A FLEXIBLE CYSTOSCOPE TO THOROUGHLY INSPECT THE BLADDER AND URETHRA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20611159 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page