HOPKINS 26120BARecalls & reported problems
Karl Storz GmbH & Co. KG · HOPKINS® Telescope 30°, 2.9 mm, 30 cm · Class 2
All models of HOPKINS (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 04048551093832), received between Jun 5, 2017 and Sep 9, 2026 — classified as 6 injuries, 13 malfunctions.
The most frequently reported specific device problem is “Poor Quality Image” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Poor Quality Image7
- Optical Problem4
- Break3
- Device Reprocessing Problem3
- Failure to Clean Adequately3
- Corroded2
- Crack2
- Decoupling2
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information9
- Foreign Body In Patient1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2026Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50697620 ↗MalfunctionJul 27, 2026Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50006432 ↗MalfunctionMay 13, 2026Poor Quality Image
IT WAS REPORTED THAT "THE OUTPUT IS BLURRY AND CANNOT BE WHITE-BALANCED".
Patient problems: Insufficient Information
MAUDE report 25163482 ↗MalfunctionApr 7, 2026Device Reprocessing Problem; Optical Distortion
THE FOLLOWING WAS REPORTED: "OPTICAL SYSTEM: FOGGY." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 24801206 ↗MalfunctionApr 2, 2026Optical Distortion
THE FOLLOWING WAS REPORTED: OPTICAL SYSTEM: FOGGY/ALWAYS. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 24769771 ↗MalfunctionMar 17, 2026Poor Quality Image
IT WAS REPORTED THAT THE IMAGE IS FOGGY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24623776 ↗MalfunctionMar 17, 2026Poor Quality Image
IT WAS REPORTED THAT THE IMAGE IS FOGGY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24623775 ↗MalfunctionMar 7, 2026Poor Quality Image; Failure to Clean Adequately
IT WAS REPORTED THAT "BAD IMAGE - BLACK DOTS, BLURRY. SCOPE HAS ONLY BEEN USED ONCE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24540583 ↗MalfunctionMar 2, 2026Poor Quality Image; Insufficient Device Problem Information
IT WAS REPORTED THAT "THE UTERINE OFFICE HYSTEROSCOPE MODEL 26120BA PERFORMS POORLY WHEN IN USE. THE HYSTEROSCOPE HAS CRYSTAL FRACTURE AND BLURRED IMAGE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24491914 ↗MalfunctionNov 10, 2025Break; Decoupling; Inadequate or Insufficient Training; Dent in Material; Improper Chemical Reaction; Break; Decoupling; Inadequate or Insufficient Training; Dent in Material; Improper Chemical Reaction
IT HAS BEEN REPORTED: MAJOR DAMAGE, EXPOSED LIGHT FIBERS ON LIGHT POST. THE LIGHT HOOKS BROKE OFF FROM THE SCOPE NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23514632 ↗InjuryJun 24, 2025Device Reprocessing Problem; Corroded; Detachment of Device or Device Component; Material Deformation; Optical Problem; Failure to Clean Adequately; Poor Visibility; Device Reprocessing Problem; Corroded; Detachment of Device or Device Component; Material Deformation; Optical Problem; Failure to Clean Adequately; Poor Visibility
IT WAS REPORTED THAT THE SCOPE HAD ERROR MESSAGE AND BLURRY IMAGE THAT DELAYED THE CASE FOR 45 MINUTES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22321992 ↗InjuryJun 24, 2025Optical Problem; Optical Problem
IT WAS REPORTED THAT THE SCOPE HAD ERROR MESSAGE AND BLURRY IMAGE THAT DELAYED THE CASE FOR 45 MINUTES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 22321991 ↗InjuryJun 24, 2025Device Reprocessing Problem; Corroded; Detachment of Device or Device Component; Material Deformation; Optical Problem; Failure to Clean Adequately; Poor Visibility; Device Reprocessing Problem; Corroded; Detachment of Device or Device Component; Material Deformation; Optical Problem; Failure to Clean Adequately; Poor Visibility
IT WAS REPORTED THAT THE SCOPE HAD ERROR MESSAGE AND BLURRY IMAGE THAT DELAYED THE CASE FOR 45 MINUTES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22321990 ↗InjuryJun 24, 2025Optical Problem; Optical Problem
IT WAS REPORTED THAT THE SCOPE HAD ERROR MESSAGE AND BLURRY IMAGE THAT DELAYED THE CASE FOR 45 MINUTES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 22321989 ↗MalfunctionMay 15, 2025Crack; Decoupling; Inadequate or Insufficient Training; Dent in Material; Improper Chemical Reaction; Crack; Decoupling; Inadequate or Insufficient Training; Dent in Material; Improper Chemical Reaction
IT WAS REPORTED THE LENS IS CRACKED AND THE VISIBILITY HAS DIMINISHED DURING MEDICAL PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22031469 ↗MalfunctionMay 15, 2025Crack; Material Integrity Problem; Crack; Material Integrity Problem
IT WAS REPORTED THE LENS IS CRACKED AND THE VISIBILITY HAS DIMINISHED DURING MEDICAL PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22031464 ↗InjuryJan 28, 2025Break; Break
IT WAS REPORTED DURING A MEDICAL PROCEDURE ON (B)(6) 2024, THE PHYSICIAN WAS TRYING TO GET THE PRODUCT THROUGH THE CERVIX AND THE SCOPE BROKE. THE HYSTEROSCOPE SHEATH AND HOPKINS TELESCOPE 30°, 2.9 MM, 30 CM WERE ATTACHED (BOTH INSTRUMENTS WERE BEING USED DURING THE PROCEDURE). IT WAS REPORTED DURING THE HYSTEROSCOPY PROCEDURE WAS PROLONGED FOR 30 MINUTES ON (B)(6), 2024, THE SURGEON WAS NOT ABLE TO COMPLETE THE PROCEDURE, THE PATIENT WAS REFERRED TO GYN CROSS REFERENCE MDR: 9610617-2025-00025.
Patient problems: Insufficient Information
MAUDE report 21261286 ↗InjuryJan 14, 2025Break; Improper or Incorrect Procedure or Method; Break; Improper or Incorrect Procedure or Method
IT WAS REPORTED DURING A MEDICAL PROCEDURE ON (B)(6) 2024, THE PHYSICIAN WAS TRYING TO GET THE PRODUCT THROUGH THE CERVIX AND THE SCOPE BROKE. THE HYSTEROSCOPE SHEATH AND HOPKINS TELESCOPE 30°, 2.9 MM, 30 CM WERE ATTACHED (BOTH INSTRUMENTS WERE BEING USED DURING THE PROCEDURE). IT WAS NOT REPORTED IF THERE WAS ANY HARM TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED AFTER MULTIPLE ATTEMPTS. CROSS REFERENCE MDR: 9610617-2025-00025.
Patient problems: Insufficient Information
MAUDE report 21150654 ↗MalfunctionJun 5, 2017Material Fragmentation; Material Fragmentation
THIS IS A SUPPLEMENTAL. I WILL BE ADDING METHOD/RESULTS/CONCLUSION CODES AND EVALUATION OF THE INSTRUMENT.
Patient problems: Foreign Body In Patient
MAUDE report 6613568 ↗
Other HYSTEROSCOPE (AND ACCESSORIES) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)453 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- Myosure Reach Tissue Removal Device 10-401FCHologic, Inc. · HYSTEROSCOPE (AND ACCESSORIES)4 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.