n.a. 26159BHWRecalls & reported problems
Karl Storz GmbH & Co. KG · Biopsy Spoon Forceps, 5 Fr. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04048551094242), received between Oct 11, 2024 and Aug 27, 2026 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Detachment of Device or Device Component3
- Material Fragmentation2
- Material Separation2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryAug 27, 2026Material Separation
The following was reported: During hysteroscopic surgery in the Department of Gynecology, forceps model 26159BHW were used for the procedure. The forceps functioned properly for the first several uses, but failed to open upon subsequent use. The forceps were removed from the uterine cavity for examination, and the jaw screw ofthe forceps was found missing. The physician suspected that the fallen screw might remain inside the patient's body; therefore, hysteroscopic exploration was performed, yet the missing screw was not located. Nevertheless, the physician considered there remained a risk that the screw could be retained within the uterine cavity, hence thisreport was submitted.
Patient problems: Foreign Body In Patient
MAUDE report 50487502 ↗InjuryAug 27, 2026Material Separation
The following was reported: During hysteroscopic surgery in the Department of Gynecology, forceps model 26159BHW were used for the procedure. The forceps functioned properly for the first several uses, but failed to open upon subsequent use. The forceps were removed from the uterine cavity for examination, and the jaw screw ofthe forceps was found missing. The physician suspected that the fallen screw might remain inside the patient's body; therefore, hysteroscopic exploration was performed, yet the missing screw was not located. Nevertheless, the physician considered there remained a risk that the screw could be retained within the uterine cavity, hence thisreport was submitted.
Patient problems: Foreign Body In Patient
MAUDE report 50487493 ↗InjuryJun 22, 2026Material Fragmentation
IT WAS REPORTED THAT WHILE IN THE UTERINE CAVITY, THE HINGE SCREW OF THE BIOPSY FORCEPS CAME LOOSE. COULD NOT FIND THE SCREW (VERY MINUSCULE).
Patient problems: Foreign Body In Patient
MAUDE report 25585507 ↗MalfunctionMar 5, 2026Break
IT WAS REPORTED THAT JAW IS BROKEN. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24524050 ↗MalfunctionFeb 10, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT THE DISTAL TIP FALL OFF, ALWAYS. DURING THE TECHNICAL INSPECTION IT WAS DETERMINED THAT THE PIN FROM THE DISTAL JOINT IS MISSING. THERE IS NO INFORMATION THAT THE EVENT CAUSED A HARM OR SERIOUS INJURY TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21345398 ↗MalfunctionNov 8, 2024Break; Material Fragmentation; Break; Material Fragmentation
IT WAS REPORTED THAT DURING A HYSTEROSCOPY PROCEDURE IN THE OBSTETRICS AND GYNECOLOGY DEPARTMENT, AN ISSUE OCCURRED WHILE USING THE DEVICE TO COLLECT TISSUE SAMPLES FROM A LESION. THE CONNECTING PART OF THE JAW EXPERIENCED A FAILURE, CAUSING THE PIN TO DETACH. AS A RESULT, THERE WAS A ROUGHLY ONE-HOUR PROCESS TO LOCATE THE PIN WITHIN THE PATIENT'S BODY, AND IT WAS SUCCESSFULLY RETRIEVED. IT WAS CONFIRMED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. SINCE THE PROCEDURE HAD TO BE PROLONGED BY ROUGHLY ONE-HOUR TO LOCATE THE PIN WITHIN THE PATIENT'S BODY, THIS CASE IS DEEMED REPORTABLE AS A PRECAUTIONARY ACTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20642574 ↗MalfunctionOct 30, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING GYNECOLOGIC HYSTEROSCOPIC SURGERY, THE CONNECTION PART OF THE JAW BROKE WHILE THE PRODUCT WAS BEING USED AS A HYSTERO SHEATH WORKING CHANNEL TO COLLECT A LESION. THE HOSPITAL STAFF DO NOT KNOW WHERE THE MISSING PART HAS FALLEN OR IF X-RAYS WERE TAKEN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SURGICAL DELAY OF LESS THAN 15 MINUTES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20569731 ↗MalfunctionOct 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
"IT WAS REPORTED THAT DURING GYNECOLOGIC HYSTEROSCOPIC SURGERY, THE CONNECTION PART OF THE JAW PART FAILED WHEN THE PRODUCT WAS USED AS HYSTERO SHEATH WORKING CHANNEL TO COLLECT LESIONS. THE SURGERY WAS PROLONGED FOR LESS THAN 15 MINUTES. AFTER THE SURGERY THE PRODUCT WAS RETURNED TO THE LOCAL SUBSIDIARY AND INSPECTED. IT WAS DISCOVERED THAT A BAR-LIKE PART CONNECTING TWO SECTIONS OF THE SQUARE BOX WAS MISSING. NO HARM ON PATIENT OR NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. BECAUSE OF THE UNCERTAIN INFORMATION IF THE FALLEN OFF CONNECTION PART WAS LEFT IN THE PATIENT A REPORT TO THE FDA IS REQUIRED AS A PRECAUTIONARY MEASURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20429794 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page