n.a. 26055CBRecalls & reported problems
Karl Storz GmbH & Co. KG · Inner Sheath Resectoscope Sheath,22Charr · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04048551275597), received between Jul 23, 2020 and Aug 5, 2026 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Material Fragmentation2
- Fracture1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50146858 ↗MalfunctionMar 30, 2026Break
THE FOLLOWING INFORMATION WAS RECEIVED: DURING TCRF LOOP BROKE INSIDE PATIENT, THE BIPOLAR CUT WAS THEN ACTIVATED CAUSING A SPARK AND BURN/CRACK TO THE CERAMIC BEAK ON SHEATH. DARK CHARRED MARKS WERE NOTED ON FIBROIDS HOWEVER CUSTOMER SWAPPED SET AND CARRIED ON RESECTION AS NORMAL, AT THE END OF THE PROCEDURE THE DARK MARKS WERE NO LONGER PRESENT. THE BROKEN PART WAS RETRIEVED FROM THE PATIENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. THE FOLLOWING INFORMATION WAS RECEIVED: DURING TCRF LOOP BROKE INSIDE PATIENT, THE BIPOLAR CUT WAS THEN ACTIVATED CAUSING A SPARK AND BURN/CRACK TO THE CERAMIC BEAK ON SHEATH. DARK CHARRED MARKS WERE NOTED ON FIBROIDS HOWEVER CUSTOMER SWAPPED SET AND CARRIED ON RESECTION AS NORMAL, AT THE END OF THE PROCEDURE THE DARK MARKS WERE NO LONGER PRESENT. THE BROKEN PART WAS RETRIEVED FROM THE PATIENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR 9610617-2026-00674.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24708742 ↗MalfunctionFeb 18, 2026Break
ACCORDING TO THE INFORMATION RECEIVED, DURING RESECTION OF FIBROID, LOOP BECAME DAMAGED AND THE INNER SHEATH CERAMIC BEAK WAS SUBSEQUENTLY DAMAGED, HAS DARK MARKINGS ON INSIDE, THE CERAMIC BEAK CRACKED BUT DID NOT DETACH INSIDE THE PATIENT. THUS, THEY DID NOT X-RAY. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24378985 ↗InjuryJun 14, 2022Material Fragmentation; Break; Material Fragmentation
AS PER MANUFACTURER INCIDENT REPORT #(B)(4) RECEIVED FROM THE FACTORY IN GERMANY: THE REPORTED EVENT HAPPENED IN SPAIN. ACCORDING TO THE INFORMATION AVAILABLE, THE CERAMIC TIP BROKE OFF INSIDE THE PATIENT, WITH A KIND OF EXPLOSION. THE PHYSICIAN HAD TO REMOVE THE PARTS FROM INSIDE THE PATIENT WITHOUT BEING SURE THAT ALL PARTS WERE COLLECTED. THERE IS NO FURTHER INFORMATION RELATED TO THE DESCRIBED EXPLOSION.
Patient problems: Insufficient Information
MAUDE report 14685225 ↗MalfunctionMay 6, 2022Material Fragmentation; Material Fragmentation
AS PER MANUFACTURER INCIDENT REPORT #(B)(4) RECEIVED FROM THE FACTORY IN (B)(4): THE EVENT OCCURRED IN (B)(6). ACCORDING TO THE INFORMATION AVAILABLE, THE CERAMIC TIP OF THE 26055CB BROKE IN THE PATIENT'S UTERUS. A FEW OF THE BROKEN PARTS ARE STILL IN THE PATIENT'S UTERUS.
Patient problems: Foreign Body In Patient
MAUDE report 14311358 ↗MalfunctionJul 23, 2020Fracture; Fracture
AS PER A REPORT WE RECEIVED FROM THE FACTORY IN (B)(6) IN REGARD TO AN INCIDENT THAT OCCURRED IN (B)(6): WHEN SURGEON WAS PERFORMING A TCR, AT AROUND THREE MINUTES, THEY FOUND THE CUTTING LOOP BURNING AND THE SCOPE WAS DAMAGED AND THE WHITE CERAMIC OF SHEATH ALSO BROKEN. AN X-RAY CONFIRMED EVERYTHING WAS RETRIEVED; THERE WAS NO HARM TO PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10314000 ↗
Other HYSTEROSCOPE (AND ACCESSORIES) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)453 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- Myosure Reach Tissue Removal Device 10-401FCHologic, Inc. · HYSTEROSCOPE (AND ACCESSORIES)4 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page