C-MAC 8403AXCRecalls & reported problems
Karl Storz GmbH & Co. KG · C-MAC Video Laryngoscope MAC #3 · Class 1
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04048551389478), received between May 11, 2025 and Sep 17, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Output Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Output Problem4
- Device Reprocessing Problem3
- No Display/Image3
- Component Missing1
- Detachment of Device or Device Component1
- Device Handling Problem1
- Loss of or Failure to Bond1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50812702 ↗MalfunctionJan 26, 2026Loss of or Failure to Bond; No Display/Image; Component Missing
ACCORDING TO THE INFORMATION RECEIVED, A PATIENT WAS INTUBATED WITH THE C-MAC DEVICE AS PLANNED. THE DEVICE WAS USED WITHOUT ANY PROBLEMS EARLIER IN THE MORNING. THE DEVICE WAS THEN WASHED IN THE EQUIPMENT MAINTENANCE, AFTER WHICH IT WAS BROUGHT BACK TO THE ROOM TO BE CHARGED. THE DEVICE WAS SWITCHED ON AS NORMAL, AND THE POWER CORD WAS PLUGGED IN. THE IMAGE DISAPPEARED MOMENTARILY A FEW TIMES BUT RETURNED AS NORMAL. JUST BEFORE THE NEXT INTUBATION BEGAN, THE IMAGE ON THE VIDEO LARYNGOSCOPE BECAME VERY DARK, SO-CALLED NEGATIVE, I.E. THE COLORS WERE DARK AND PURPLE. THE DEVICE WAS UNUSABLE AT THAT TIME. IT WAS AN RSI INTUBATION. NO TIME TO CHANGE TO A NEW DEVICE. INTUBATED WITHOUT AN IMAGE LIKE A NORMAL LARYNGOSCOPE. AFTER THE SITUATION, THE DEVICE WAS RESTARTED, SO THE IMAGE WAS STILL UNUSABLE. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24179414 ↗MalfunctionJun 23, 2025No Display/Image; Output Problem; Device Handling Problem; No Display/Image; Output Problem; Device Handling Problem
IT WAS REPORTED THAT DURING THE PREPARATION PROCESS OF AN INTUBATION THE BLADE WAS NOT BEING RECOGNIZED. LIGHT COMES ON BUT IT DOESN'T SHOW UP ON THE CAMERA. THERE WAS NO SIGNIFICANT DELAY ONLY ABOUT 20 SECONDS TO OBTAIN THE DL EQUIPMENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 22311649 ↗MalfunctionMay 13, 2025Use of Device Problem; Detachment of Device or Device Component; Use of Device Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT COVER GLASS IS MISSING, BENT BLADE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22015285 ↗MalfunctionMay 11, 2025Device Reprocessing Problem; Output Problem; Device Reprocessing Problem; Output Problem
IT WAS REPORTED THAT "LIGHT NOT WORKING, UPON RECEIPT NOTICED THAT COVER GLASS MISSING ON BLADES." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21999922 ↗MalfunctionMay 11, 2025Device Reprocessing Problem; Output Problem; Device Reprocessing Problem; Output Problem
IT WAS REPORTED THAT "LIGHT NOT WORKING, UPON RECEIPT NOTICED THAT COVER GLASS MISSING ON BLADES." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21999917 ↗MalfunctionMay 11, 2025Device Reprocessing Problem; Output Problem; Device Reprocessing Problem; Output Problem
IT WAS REPORTED THAT "LIGHT NOT WORKING, UPON RECEIPT NOTICED THAT COVER GLASS MISSING ON BLADES." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21999889 ↗
Other LARYNGOSCOPE, RIGID devices
- McGRATH MAC 301-000-000Covidien LP · LARYNGOSCOPE, RIGID2 recalls202 reports
- McGRATH MAC 300-000-000Covidien LP · LARYNGOSCOPE, RIGID1 recall1 death205 reports
- n.a. 8404HXKarl Storz GmbH & Co. KG · LARYNGOSCOPE, RIGID242 reports
- McGRATH 340-000-000Covidien LP · LARYNGOSCOPE, RIGID2 deaths192 reports
- McGRATH MAC 301-000-000Covidien LP · LARYNGOSCOPE, RIGID2 recalls90 reports
- n.a. 8404BXCKarl Storz GmbH & Co. KG · LARYNGOSCOPE, RIGID141 reports
- n/a 8404AXCKarl Storz GmbH & Co. KG · LARYNGOSCOPE, RIGID108 reports
- McGRATH 340-000-000Covidien LP · LARYNGOSCOPE, RIGID84 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.