Skip to content

C-MAC 8403AXCRecalls & reported problems

Karl Storz GmbH & Co. KG · C-MAC Video Laryngoscope MAC #3 · Class 1

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04048551389478), received between May 11, 2025 and Sep 17, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Output Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem4
  • Device Reprocessing Problem3
  • No Display/Image3
  • Component Missing1
  • Detachment of Device or Device Component1
  • Device Handling Problem1
  • Loss of or Failure to Bond1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812702 ↗
  • MalfunctionJan 26, 2026Loss of or Failure to Bond; No Display/Image; Component Missing

    ACCORDING TO THE INFORMATION RECEIVED, A PATIENT WAS INTUBATED WITH THE C-MAC DEVICE AS PLANNED. THE DEVICE WAS USED WITHOUT ANY PROBLEMS EARLIER IN THE MORNING. THE DEVICE WAS THEN WASHED IN THE EQUIPMENT MAINTENANCE, AFTER WHICH IT WAS BROUGHT BACK TO THE ROOM TO BE CHARGED. THE DEVICE WAS SWITCHED ON AS NORMAL, AND THE POWER CORD WAS PLUGGED IN. THE IMAGE DISAPPEARED MOMENTARILY A FEW TIMES BUT RETURNED AS NORMAL. JUST BEFORE THE NEXT INTUBATION BEGAN, THE IMAGE ON THE VIDEO LARYNGOSCOPE BECAME VERY DARK, SO-CALLED NEGATIVE, I.E. THE COLORS WERE DARK AND PURPLE. THE DEVICE WAS UNUSABLE AT THAT TIME. IT WAS AN RSI INTUBATION. NO TIME TO CHANGE TO A NEW DEVICE. INTUBATED WITHOUT AN IMAGE LIKE A NORMAL LARYNGOSCOPE. AFTER THE SITUATION, THE DEVICE WAS RESTARTED, SO THE IMAGE WAS STILL UNUSABLE. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24179414 ↗
  • MalfunctionJun 23, 2025No Display/Image; Output Problem; Device Handling Problem; No Display/Image; Output Problem; Device Handling Problem

    IT WAS REPORTED THAT DURING THE PREPARATION PROCESS OF AN INTUBATION THE BLADE WAS NOT BEING RECOGNIZED. LIGHT COMES ON BUT IT DOESN'T SHOW UP ON THE CAMERA. THERE WAS NO SIGNIFICANT DELAY ONLY ABOUT 20 SECONDS TO OBTAIN THE DL EQUIPMENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 22311649 ↗
  • MalfunctionMay 13, 2025Use of Device Problem; Detachment of Device or Device Component; Use of Device Problem; Detachment of Device or Device Component

    IT WAS REPORTED THAT COVER GLASS IS MISSING, BENT BLADE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22015285 ↗
  • MalfunctionMay 11, 2025Device Reprocessing Problem; Output Problem; Device Reprocessing Problem; Output Problem

    IT WAS REPORTED THAT "LIGHT NOT WORKING, UPON RECEIPT NOTICED THAT COVER GLASS MISSING ON BLADES." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21999922 ↗
  • MalfunctionMay 11, 2025Device Reprocessing Problem; Output Problem; Device Reprocessing Problem; Output Problem

    IT WAS REPORTED THAT "LIGHT NOT WORKING, UPON RECEIPT NOTICED THAT COVER GLASS MISSING ON BLADES." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21999917 ↗
  • MalfunctionMay 11, 2025Device Reprocessing Problem; Output Problem; Device Reprocessing Problem; Output Problem

    IT WAS REPORTED THAT "LIGHT NOT WORKING, UPON RECEIPT NOTICED THAT COVER GLASS MISSING ON BLADES." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21999889 ↗

Other LARYNGOSCOPE, RIGID devices

All LARYNGOSCOPE, RIGID devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.