n.a. 021199-10Recalls & reported problems
Karl Storz GmbH & Co. KG · Valve Seal · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048551400227), received between Apr 13, 2026 and Apr 13, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Output Problem2
- Physical Resistance/Sticking2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 13, 2026Detachment of Device or Device Component; Output Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT THE VALVES OF THE TROCARS BROKE DURING THEIR FIRST USE THE HANDLE PRESENTED EXCESSIVE RESISTANCE WHEN OPENING/CLOSING AND WHEN ROTATING THE KNOB, THEY WERE IN THE PROCESS OF PERFORMING SURGERY ON A NEWBORN PATIENT WHEN THE PRODUCTS MENTION FAILED. AS REPORTED PATIENT IS IN THE USUAL CONDITION EXPECTED AFTER SURGERY OF THIS TYPE (THERE IS NO WORSENED CONDITION DUE TO THE EQUIPMENT FAILURE). DURING THE PROCEDURE, THE VALVES OF THE TROCARS BROKE DURING THEIR FIRST USE, APPROXIMATELY HALFWAY THROUGH THE SURGERY. AS A CONSEQUENCE, THE SURGICAL TEAM REPEATEDLY LOST PNEUMOTHORAX, WHICH CAUSED RESPIRATORY COMPLICATIONS AND JEOPARDISED THE CONTINUATION OF THE PROCEDURE. ACCORDING TO THE HOSPITAL, THE INSTRUMENTS INSERTED WERE COMPATIBLE AND WERE HANDLED WITH CARE. IT WAS REPORTED THAT THE MONOPOLAR CABLE SUPPLIED IN THE SAME ORDER DID NOT FUNCTION. THE MEDICAL TEAM CONFIRMED THAT THE ISSUE WAS CAUSED BY THE CABLE ITSELF, AS WHEN REPLACED WITH ANOTHER STORZ CABLE AVAILABLE AT THE HOSPITAL, THE SYSTEM WORKED CORRECTLY. ISSUES WERE IDENTIFIED IN THE HANDLE, WHICH WAS BEING USED FOR THE SECOND TIME. THE CLINICIANS REPORTED THAT THE HANDLE PRESENTED EXCESSIVE RESISTANCE WHEN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24853507 ↗InjuryApr 13, 2026Detachment of Device or Device Component; Output Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT THE VALVES OF THE TROCARS BROKE DURING THEIR FIRST USE THE HANDLE PRESENTED EXCESSIVE RESISTANCE WHEN OPENING/CLOSING AND WHEN ROTATING THE KNOB, THEY WERE IN THE PROCESS OF PERFORMING SURGERY ON A NEWBORN PATIENT WHEN THE PRODUCTS MENTION FAILED. AS REPORTED PATIENT IS IN THE USUAL CONDITION EXPECTED AFTER SURGERY OF THIS TYPE (THERE IS NO WORSENED CONDITION DUE TO THE EQUIPMENT FAILURE). DURING THE PROCEDURE, THE VALVES OF THE TROCARS BROKE DURING THEIR FIRST USE, APPROXIMATELY HALFWAY THROUGH THE SURGERY. AS A CONSEQUENCE, THE SURGICAL TEAM REPEATEDLY LOST PNEUMOTHORAX, WHICH CAUSED RESPIRATORY COMPLICATIONS AND JEOPARDISED THE CONTINUATION OF THE PROCEDURE. ACCORDING TO THE HOSPITAL, THE INSTRUMENTS INSERTED WERE COMPATIBLE AND WERE HANDLED WITH CARE. IT WAS REPORTED THAT THE MONOPOLAR CABLE SUPPLIED IN THE SAME ORDER DID NOT FUNCTION. THE MEDICAL TEAM CONFIRMED THAT THE ISSUE WAS CAUSED BY THE CABLE ITSELF, AS WHEN REPLACED WITH ANOTHER STORZ CABLE AVAILABLE AT THE HOSPITAL, THE SYSTEM WORKED CORRECTLY. ISSUES WERE IDENTIFIED IN THE HANDLE, WHICH WAS BEING USED FOR THE SECOND TIME. THE CLINICIANS REPORTED THAT THE HANDLE PRESENTED EXCESSIVE RESISTANCE WHEN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24853485 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page