Skip to content

NA 10973HDRecalls & reported problems

Karl Storz GmbH & Co. KG · Video Mediastinoscope · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04048551425718), received between Dec 21, 2021 and May 8, 2023 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Sharp Edges” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Sharp Edges5
  • Device Handling Problem1

Patient problems

  • Needle Stick/Puncture4
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2023Device Handling Problem; Sharp Edges; Device Handling Problem; Sharp Edges

    IT WAS REPORTED THAT THERE WAS EVENT WITH A 10973HD VIDEO MEDIASTINOSCOPE. THE SURGEON COMPLAINED THAT THIS HD VERSION FELT TO SHARP AT THE TIP AND COMPLAINED ABOUT THE ANGLE AT THE DISTAL PORTION OF THE SCOPE. SURGEON FELT THERE WAS A REAL RISK OF PERFORATING THE TRACHEA. NO PATIENT INJURY, PATIENT UNDER ANESTHESIA, BUT NO INCISION MADE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16885815 ↗
  • InjuryDec 21, 2021Sharp Edges; Sharp Edges

    AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(4): AS REPORTED: "PROF. LEVENT CANSEVER SAID THAT HE WAS PERFORMING A DIAGNOSTIC MEDIASTINOSCOPY AND CONDUCTING A BIOPSY ON TRACHEAL LYMPE NODES, AS A USUAL PROCEDURE FOR HIM, WHILE HE WAS PUSHING FORWARD THE BLADE IN MID-LINE OF THE TRACHEA, HE FELT STRESS ON THE TISSUE THAN HE PUSHED THE BLADE MORE AND THE TRACHEA WAS INJURED. AFTER THIS INCIDENT, HE HAD TO DO OPEN SURGERY FOR TRACHEA REPAIR." (MANUFACTURER'S INTERNAL COMPLAINT # (B)(4)).

    Patient problems: Needle Stick/Puncture

    MAUDE report 13052553 ↗
  • InjuryDec 21, 2021Sharp Edges; Sharp Edges

    AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(6): AS REPORTED, "TWO INSERTIONS OF THE MEDIASTINOSCOPE EACH RESULTED IN SHEARING OF A LOWER THYROID VEIN WITH SEVERE BLOOD LOSS. " (MANUFACTURER'S INTERNAL COMPLAINT #(B)(4)).

    Patient problems: Needle Stick/Puncture

    MAUDE report 13052548 ↗
  • InjuryDec 21, 2021Sharp Edges; Sharp Edges

    AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(4): "DURING A MEDIA STINOSCOPY WITH OUR NEWLY PURCHASED VIDEO MEDIASTINOSCOP, APPARENTLY A TRACHEA TEAR HAS OCCURRED DURING THE INSERTION OF THE SCOPE." (MANUFACTURER'S INTERNAL COMPLAINT # (B)(4)).

    Patient problems: Needle Stick/Puncture

    MAUDE report 13052546 ↗
  • InjuryDec 21, 2021Sharp Edges; Sharp Edges

    AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(4): AS REPORTED: "DUE TO A SHARP EDGE AT THE TIP OF THE SCOPE, THE BRONCHUS OF THE PATIENT WAS RIPPED UPON INTRODUCTION OF THE SCOPE, DAMAGE TO BRONCHUS HAD TO BE SUTURED/REPAIRED" (MANUFACTURER'S INTERNAL COMPLAINT # (B)(4)).

    Patient problems: Needle Stick/Puncture

    MAUDE report 13052545 ↗

FDA clearance and classification

The FDA regulates product code EWY (mediastinoscope, surgical) under 21 CFR 874.4720 as a Class II device. 21 CFR 874.4720 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the NA 10973HD

Has the NA 10973HD been recalled?

No FDA recall record in our data names this exact device identifier (DI 04048551425718), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 10973HD?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Dec 21, 2021 and May 8, 2023. In those reports, the most common specific device problems are “Sharp Edges” (5) and “Device Handling Problem” (1); the most common patient problem is “Needle Stick/Puncture” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 10973HD?

None of the reports are classified as deaths, 4 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the NA 10973HD, and is it still sold?

The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand NA, model 10973HD). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Mediastinoscope, surgical devices

All Mediastinoscope, surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page