Karl Storz GmbH & Co. KG · TIPCAM1 RUBINA 0°, OPAL1 NIR/ICG, 4K 3D · Class 2
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04048551426180), received between Nov 5, 2025 and Sep 29, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Display or Visual Feedback Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Display or Visual Feedback Problem3
- Optical Distortion3
- Poor Quality Image3
- Insufficient Device Problem Information2
- Unexpected Shutdown1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936914 ↗MalfunctionSep 23, 2026Unexpected Shutdown
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50876093 ↗MalfunctionAug 20, 2026Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50354328 ↗MalfunctionApr 14, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT THE OPERATING ROOM TEAM INTERFACE (ORTI) SCREEN ON THE TOWER BEGAN FLICKERING AND BECAME FROZEN AFTER AN ENDOSCOPE ERROR MESSAGE WAS TRIGGERED. THE ISSUE WAS RESOLVED BY REPLACING THE ENDOSCOPE WITH A 30 DEGREE ENDOSCOPE. THERE WAS NO PATIENT INJURY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24874687 ↗MalfunctionApr 14, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT THE OPERATING ROOM TEAM INTERFACE (ORTI) SCREEN ON THE TOWER BEGAN FLICKERING AND BECAME FROZEN AFTER AN ENDOSCOPE ERROR MESSAGE WAS TRIGGERED. THE ISSUE WAS RESOLVED BY REPLACING THE ENDOSCOPE WITH A 30 DEGREE ENDOSCOPE. THERE WAS NO PATIENT INJURY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24873974 ↗MalfunctionMar 20, 2026Optical Distortion
IT WAS REPORTED THAT, DURING A ROBOTIC LAPAROSCOPIC PROCEDURE, THE ENDOSCOPE'S LENS WAS INFILTRATED, RESULTING IN A BLURRED FIELD OF VISION ON THE IMAGE. THERE WAS NO PATIENT INJURY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24656030 ↗MalfunctionFeb 16, 2026Optical Distortion
IT WAS REPORTED THAT, DURING A ROBOTIC LAPAROSCOPIC PROSTATECTOMY, THE ENDOSCOPE'S IMAGE WAS BLURRED AND UNFOCUSED. IT WAS EXCHANGED WITH A NEW ENDOSCOPE AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. THE EXCHANGE RESULTED IN A FIVE-MINUTE DELAY. THERE WAS NO PATIENT INJURY. CROSS REFERENCE MDR: 9610617-2026-00311.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24350023 ↗MalfunctionFeb 16, 2026Optical Distortion
IT WAS REPORTED THAT, DURING A ROBOTIC LAPAROSCOPIC PROSTATECTOMY, THE ENDOSCOPE'S IMAGE WAS BLURRED AND UNFOCUSED. IT WAS EXCHANGED WITH A NEW ENDOSCOPE AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. THE EXCHANGE RESULTED IN A FIVE MINUTE DELAY. THERE WAS NO PATIENT INJURY. CROSS REFERENCE MDR: 9610617-2026-00310.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24350022 ↗MalfunctionDec 23, 2025Poor Quality Image; Insufficient Device Problem Information; Poor Quality Image; Insufficient Device Problem Information
IT WAS REPORTED THAT THE ENDOSCOPE IMAGE WAS BLURRY DURING PROCEDURE. IT APPEARED AS IF THE AUTOFOCUS OF THE CAMERA WAS NOT PERFORMING WELL. EXTERNAL MONITORS SHOWED THE SAME BLUR. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23884892 ↗MalfunctionNov 5, 2025Display or Visual Feedback Problem; Insufficient Device Problem Information; Display or Visual Feedback Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT THE ENDOSCOPE CAMERA DID NOT FUNCTION AND THERE WAS NO IMAGE. A NEW ENDOSCOPE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS A 5-MINUTE DELAY. THERE WAS NO PATIENT INJURY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23477135 ↗
Other Confocal optical imaging devices
- HD 3CMOS AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S200-XZ-EAOlympus Medical Systems Corp. · Confocal Optical Imaging1,110 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S200Olympus Medical Systems Corp. · Confocal Optical Imaging870 reports
- HD 3CMOS CAMERA HEAD OLYMPUS CH-S200-XZ-EBOlympus Medical Systems Corp. · Confocal Optical Imaging698 reports
- L11 0220230300Stryker Corporation · Confocal optical imaging233 reports
- L10 0220220300Stryker Corporation · Confocal optical imaging77 reports
- AIM SafeLight 0233050300Stryker Corporation · Confocal optical imaging62 reports
- SafeLight 0233050400Stryker Corporation · Confocal optical imaging56 reports
- 1588 AIM 1588010000Stryker Corporation · Confocal optical imaging35 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.