NA TH121Recalls & reported problems

Karl Storz GmbH & Co. KG · IMAGE1 S 4U RUBINA, OPAL1 NIR/ICG · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
22
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 04048551430569), received between Jul 11, 2025 and Sep 17, 2026 — classified as 2 injuries, 20 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 4 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 4 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 5 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image8
  • No Display/Image5
  • Erratic or Intermittent Display3
  • Image Display Error/Artifact3
  • Optical Distortion2
  • Adverse Event Without Identified Device or Use Problem1
  • Circuit Failure1
  • Excessive Heating1

Patient problems

  • No Clinical Signs, Symptoms or Conditions20
  • Insufficient Information2

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026No Display/Image

    It was reported that "Camera head failed during a case-image went blank. Patient involved.". No negative impact in state of health reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50805450 ↗
  • MalfunctionSep 16, 2026Poor Quality Image

    According to the information received, the device just received back from repair and same issue of Blurry image quality to send back to have it re-evaluated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50789755 ↗
  • MalfunctionSep 10, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722729 ↗
  • MalfunctionSep 10, 2026Optical Distortion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722719 ↗
  • MalfunctionSep 4, 2026Intermittent Continuity; No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50616732 ↗
  • MalfunctionAug 18, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50310113 ↗
  • MalfunctionAug 14, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50260358 ↗
  • MalfunctionAug 11, 2026Excessive Heating

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50216343 ↗
  • MalfunctionAug 7, 2026Erratic or Intermittent Display

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50177960 ↗
  • MalfunctionAug 6, 2026Circuit Failure; No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50164326 ↗
  • MalfunctionMar 3, 2026No Display/Image

    IT WAS REPORTED THAT THE "IMAGE DISAPPEARS DURING USE." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24496871 ↗
  • MalfunctionJan 21, 2026Poor Quality Image

    ACCORDING TO THE INFORMATION RECEIVED, THE IMAGE IS BLURRY AND WILL NOT FOCUS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A BACK-UP DEVICE AND A 5 MIN DELAY. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24133921 ↗
  • MalfunctionDec 19, 2025Image Display Error/Artifact; Image Display Error/Artifact

    IT WAS REPORTED: "OPERATING ROOM MINIMAL MOVEMENT WOULD CAUSE DISTORTION ON THE SCREEN INCONVENIENCE FOR DOCTOR DID NOT LIKE IMAGE. NO PATIENT WAS INJURED OR HARMED DID SWITCH TO ANOTHER KARL STORZ CAMERA WITH BETTER QUALITY". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23853579 ↗
  • MalfunctionDec 19, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THEATRE 5 NCEPOD LIST OUT OF HOURS AT 1930 WHILE COMPLETING SURGERY FOR +/- LAPAROSCOPY THE IMAGE BECAME FOGGY ON THE SCREEN. THE THEATRE STAFF SWAPPED THE SCOPE FOR ANOTHER ONE, BUT DID NOT REPLACE THE LIGHT LEAD THE IMAGE IMPROVED, BUT ONLY FOR A SHORT WHILE. THEY THEN REPLACED THE CAMERA HEAD FOR ANOTHER ONE AND THE PROBLEM WAS RESOLVED. IT IS REPORTED THAT THE CASE WAS PROLONGED FOR APPROXIMATELY 15 MINUTES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23851365 ↗
  • MalfunctionDec 12, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED A MISTY AND BLURRED IMAGE ON THE CAMERA STACK SCREEN. THE SCOPE HAD BEEN CHANGED BUT DID NOT SOLVE THE PROBLEM. THE CAMERA HEAD LENS HAD BEEN CLEANED BUT THIS ALSO DID NOT RESOLVE THE ISSUE. THEN THE CAMERA HEAD HAD BEEN CHANGED FOR ANOTHER SIMILAR UNIT, THIS TOOK A FEW MINUTES TO ACHIEVE. THIS RESOLVED THE PROBLEM AND THE SURGERY WAS ABLE TO PROCEED AFTER APPROXIMATELY A 10 MINUTE DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23791641 ↗
  • MalfunctionDec 8, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE DEVICE HAS A FOGGY IMAGE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23746888 ↗
  • MalfunctionNov 26, 2025Optical Distortion; Optical Distortion

    IT WAS REPORTED: "THE IMAGE WAS OFF CENTER; THE CONNECTION WAS POOR TO THE TOWER BY LOSING SIGNAL AND THE IMAGE WAS BLURRY". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23660632 ↗
  • MalfunctionNov 21, 2025Poor Quality Image; No Apparent Adverse Event; Poor Quality Image; No Apparent Adverse Event

    IT WAS REPORTED THAT POOR IMAGE QUALITY IS OBSERVED DURING SURGERY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23622516 ↗
  • InjuryJul 18, 2025Image Display Error/Artifact; Image Display Error/Artifact

    IT WAS REPORTED THAT THE CAMERA HAD A HAZY IMAGE AND HALO EFFECT AROUND THE IMAGE. AFTER SWITCHING CAMERAS THERE WAS IMPROVEMENT BUT DUE TO BLEEDING AND SOME CONTINUED VISUALIZATION ISSUES THE CASE HAD TO BE CANCELLED. CROSS REFERENCE MDR: 2027009-2025-01229.

    Patient problems: Insufficient Information

    MAUDE report 22546403 ↗
  • InjuryJul 18, 2025Image Display Error/Artifact; Image Display Error/Artifact

    IT WAS REPORTED THAT THE CAMERA HAD A HAZY IMAGE AND HALO EFFECT AROUND THE IMAGE. AFTER SWITCHING CAMERAS THERE WAS IMPROVEMENT BUT DUE TO BLEEDING AND SOME CONTINUED VISUALIZATION ISSUES THE CASE HAD TO BE CANCELLED. CROSS REFERENCE MDR: 2027009-2025-01230.

    Patient problems: Insufficient Information

    MAUDE report 22546402 ↗
  • MalfunctionJul 18, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT FLICKERING AND CAMERA GETTING DISCONNECTED DURING PROCEDURE A FEW TIMES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22542667 ↗
  • MalfunctionJul 11, 2025Erratic or Intermittent Display; Adverse Event Without Identified Device or Use Problem; Erratic or Intermittent Display; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT "UNIT BLANKS OUT AND IMAGE FLICKERING OCCURRED". IT WAS CONFIRMED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT, USER OR THIRD PARTY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDRS: 9610617-2025-01210, 9610617-2025-01212.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22482863 ↗

Other Confocal optical imaging devices

All Confocal optical imaging devices →