Karl Storz GmbH & Co. KG · HOPKINS RUBINA 30°, NIR/ICG, Ø 10 mm · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04048551436240), received between Apr 29, 2025 and Apr 7, 2026 — classified as 3 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • No Display/Image2
  • Poor Quality Image2
  • Use of Device Problem2
  • Material Deformation1
  • Optical Distortion1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Burn(s)2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 7, 2026Optical Distortion

    THE FOLLOWING WAS REPORTED: "OPTICAL SYSTEM: FOGGY." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24801233 ↗
  • MalfunctionJan 21, 2026Poor Quality Image; Material Deformation; Scratched Material

    IT WAS REPORTED THAT THE OPTIC IS FOGGY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24141591 ↗
  • MalfunctionDec 3, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT "OPTICAL SYSTEM IS FOGGY". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23706736 ↗
  • InjuryJun 16, 2025No Display/Image; No Display/Image

    ACCORDING TO THE INFORMATION RECEIVED FROM THE FIRST MOMENT IT HAS SUFFERED FOGGING, BUT THE SITUATION GOT WORSE IN SURGERY. DURING SURGERY THEY LOST TOTAL VISION. THE SURGERY WAS SCHEDULED FOR LAPAROSCOPY, BUT THEY WERE NOT ABLE TO SEE ANYTHING THROUGH OPTICS, IN THE END THEY HAD TO CHANGE THE METHOD OF INTERVENTION TO OPEN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22211041 ↗
  • InjuryJun 16, 2025No Display/Image; Use of Device Problem; No Display/Image; Use of Device Problem

    ACCORDING TO THE INFORMATION RECEIVED FROM THE FIRST MOMENT IT HAS SUFFERED FOGGING, BUT THE SITUATION GOT WORSE IN SURGERY. DURING SURGERY THEY LOST TOTAL VISION. THE SURGERY WAS SCHEDULED FOR LAPAROSCOPY, BUT THEY WERE NOT ABLE TO SEE ANYTHING THROUGH OPTICS, IN THE END THEY HAD TO CHANGE THE METHOD OF INTERVENTION TO OPEN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22210989 ↗
  • InjuryApr 30, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST SURGERY USING A KARL STORZ 10 MM LAPAROSCOPE (MODEL: 26003BRA) IN CONJUNCTION WITH A TL400 COLD LIGHT SOURCE (WITH THE AUTOMATIC BRIGHTNESS MODE ON). AN UNEXPECTED INCIDENT OCCURRED DURING THE OPERATION, RESULTING IN THE PATIENT'S SKIN BEING BURNED BY THE LENS[?] AFTER THE PRODUCE, THE PATIENT DID NOT REQUIRE ADDITIONAL INTERVENTION TREATMENT AND THE CONDITION WAS CONTROLLABLE.

    Patient problems: Burn(s)

    MAUDE report 21936322 ↗
  • MalfunctionApr 29, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST SURGERY USING A KARL STORZ 10MM LAPAROSCOPE (MODEL: 26003BRA) IN CONJUNCTION WITH A TL400 COLD LIGHT SOURCE (WITH THE AUTOMATIC BRIGHTNESS MODE ON). AN UNEXPECTED INCIDENT OCCURRED DURING THE OPERATION, RESULTING IN THE PATIENT'S SKIN BEING BURNED BY THE LENS[?] AFTER THE PRODUCE, THE PATIENT DID NOT REQUIRE ADDITIONAL INTERVENTION TREATMENT AND THE CONDITION WAS CONTROLLABLE.

    Patient problems: Burn(s)

    MAUDE report 21930564 ↗