Quick summary
This Dräger Babylog VN800 ventilator has 23 adverse event reports naming its identifier, with no deaths and one injury. Reports mostly describe the unit failing to deliver ventilation or airway pressure, the display or touch screen going blank or freezing, and the device shutting down or alarming during use. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 04048675542476), received between Sep 28, 2022 and Mar 5, 2026 — classified as 1 injury, 21 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Deliver11
- Device Difficult to Program or Calibrate9
- Output Problem5
- Decrease in Pressure4
- Gas Output Problem4
- Human-Device Interface Problem4
- No Display/Image4
- Imprecise or Erratic Results3
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Respiratory Distress Syndrome of Newborns1
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2026Complete Blockage; Gas Output Problem; Failure to Deliver; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DEVICE STOPPED WORKING WHILE THE PATIENT WAS BEING HANDLED FOR EXAMINATION AND CANNULATION. THE CPAP MASK WAS DISCONNECTED FOR A FEW MINUTES. UPON RECONNECTION, IT WAS NOTED THAT THE DEVICE WAS NOT DELIVERING ANY AIRWAY PRESSURE. THE MONITOR REMAINED FUNCTIONAL AND CONFIRMED THE SAME. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24516669 ↗MalfunctionFeb 19, 2026Gas Output Problem; Failure to Deliver; Complete Loss of Power; Intermittent Communication Failure
IT WAS REPORTED THAT DURING USE, A SUDDEN BEEPING ALARM SOUNDED, THE POWER SHUT DOWN, THE SCREEN WENT BLACK, AND VENTILATION STOPPED. AFTER THAT, TURNING THE POWER OFF AND ON ONLY TRIGGERED THE ALARM SOUND, AND THE UNIT DID NOT START. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24385613 ↗MalfunctionFeb 12, 2026No Display/Image; Display or Visual Feedback Problem; Device Difficult to Program or Calibrate
IT WAS REPORTED THAT DURING TURNING THE MONITOR TO ONE SIDE IT GOES BLANK AND WHEN TURN IT TO THE OTHER THE SIDE IT WILL COME BACK ON. NO PATIENT HEALTH CONSEQUENCES HAVE BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24320449 ↗MalfunctionJan 16, 2026Display or Visual Feedback Problem; Image Display Error/Artifact; Device Difficult to Program or Calibrate
THE FOLLOWING WAS REPORTED: SCREEN WENT BLANK - CAME BACK UP - THEN TOUCH SCREEN FROZE UP DATE/TIME: (B)(6) 2025/TIME UNKNOWN DURING THE NIGHT. SWAPPED VENT TO RESOLVE. NO ADVERSE PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24091610 ↗MalfunctionDec 18, 2025Failure to Deliver; Device Sensing Problem; No Visual Prompts/Feedback; Failure to Deliver; Device Sensing Problem; No Visual Prompts/Feedback
IT WAS REPORTED THAT THE DEVICE STOPPED DELIVERING BREATHS WHILST IN USE ON A PATIENT. THE USER INTERFACE WENT BLANK AT THE SAME TIME. THERE WERE NO HEALTH CONSEQUENCES FOR THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23840872 ↗MalfunctionDec 15, 2025Device Difficult to Program or Calibrate; Human-Device Interface Problem; Output Problem; Device Difficult to Program or Calibrate; Human-Device Interface Problem; Output Problem
IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE STOPPED WORKING DURING USE ON A PATIENT. THE DISPLAY LOCKED AND NO CHANGES COULD BE MADE. REPORTEDLY, THE DEVICE KEPT VENTILATING THE PATIENT. THERE WERE NO HEALTH CONSEQUENCES FOR THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23801257 ↗MalfunctionDec 1, 2025Failure to Deliver; Failure to Deliver
IT WAS REPORTED THAT CUSTOMER STATES UNIT SHUT OFF WITH HIGH PITCHED ALARM WHILE ON A PATIENT. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23678879 ↗MalfunctionOct 28, 2025Failure to Deliver; Complete Loss of Power; Unexpected Shutdown; Failure to Deliver; Complete Loss of Power; Unexpected Shutdown
IT WAS REPORTED THAT THE SCREEN OF THE DEVICE WENT BLACK, AND THE VENTILATION STOPPED DURING USE. THERE WERE NO HEALTH CONSEQUENCES FOR THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23406225 ↗MalfunctionJul 29, 2025No Display/Image; Device Difficult to Program or Calibrate; Unexpected Shutdown; No Display/Image; Device Difficult to Program or Calibrate; Unexpected Shutdown
IT WAS REPORTED THAT THE VENTILATOR SHUT DOWN DURING USE AND COULD NOT BE RESTARTED. IT WAS REPORTED THAT NO PATIENT INJURY OCCURRED. AT THE TIME OF THE INCIDENT, THE DEVICE WAS CONNECTED TO THE MAINS POWER AND THE BATTERY WAS CHARGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22641058 ↗MalfunctionJun 25, 2025Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure; Gas Output Problem; Decrease in Pressure; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT THE UNIT SHUT OFF WHILE IN USE. NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22326779 ↗MalfunctionApr 24, 2025Degraded; Device Difficult to Program or Calibrate; Human-Device Interface Problem; Output Problem; Degraded; Device Difficult to Program or Calibrate; Human-Device Interface Problem; Output Problem
IT WAS REPORTED THAT THE TOUCH PANEL DOES NOT WORK AND CANNOT BE OPERATED AT ALL. TURNING OFF WAS ATTEMPTED BUT FAILED. THERE WAS NO HEALTH CONSEQUENCES FOR THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21897985 ↗InjuryMar 26, 2025Use of Device Problem; No Apparent Adverse Event; Device Handling Problem; Misassembled During Installation; Use of Device Problem; No Apparent Adverse Event; Device Handling Problem; Misassembled During Installation
IT WAS REPORTED FROM CUSTOMER THAT DUE TO THE DESIGN OF THE VENTILATOR IT IS POSSIBLE TO INSERT THE INSPIRATORY TUBING INTO THE EXPIRATORY PORT AND VICE VERSA BETWEEN THE HUMIDITY AND VENTILATOR, AS THE TUBING AND PORTS ARE THE SAME SIZE, MAKING IT AN INHERENTLY UNSAFE DESIGN AND AT RISK OF FUTURE INCIDENTS. THIS OCCURRED ON A DRAGER BABYLOG VN800 VENTILATOR IN USE ON A BABY BETWEEN (B)(6) 2024, RESULTING IN NO HUMIDIFIED GASES GOING TO THE BABY DURING THIS TIME. IN THIS CASE IT LED TO RESPIRATORY SECRETIONS DRYING OUT AND BLOCKING THE ET TUBES, NEEDING X5 ET TUBE REPLACEMENTS IN 11 DAY PERIOD DUE TO OBSTRUCTION WITH THESE THICK SECRETIONS, AS WELL AS REGULAR ET LAVAGE TO REMOVE SECRETIONS. THE BABY POSSIBLY STAYED ON VENTILATOR LONGER AND HAD INCREASED HOSPITAL STAY FOR MORE DAYS THAN WOULD HAVE BEEN. TO CLEAR SECRETIONS REQUIRES HUMIDITY AND THE CILIA HAIRS IN THE LINING OF THE AIRWAY MOVES THESE UPWARDS TO BE CLEARED. THE DRY GASES MAY HAVE AFFECTED THESE CILIA, MAKING IT MORE DIFFICULT FOR SECRETIONS TO BE CLEARED AND MAY HAVE CAUSED LONG TERM DAMAGE. BABY IS NOW IN AIR AND HAS NO INCREASED WORK OF BREATHING OR COUGH AND IT IS THEREFORE UNLIKELY THAT THE LACK OF HUMIDITY HAS HA
Patient problems: Respiratory Distress Syndrome of Newborns
MAUDE report 21695998 ↗MalfunctionDec 18, 2024Decrease in Pressure; Failure to Deliver; Imprecise or Erratic Results; Decrease in Pressure; Failure to Deliver; Imprecise or Erratic Results
IT WAS REPORTED THAT THE DEVICE ISSUED A DEVICE FAILURE ALARM DURING USE. THE DEVICE SHUT DOWN. THERE WERE NO HEALTH CONSEQUENCES FOR THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20966547 ↗MalfunctionDec 16, 2024Decrease in Pressure; Failure to Deliver; Imprecise or Erratic Results; Decrease in Pressure; Failure to Deliver; Imprecise or Erratic Results
IT WAS REPORTED THAT THE DEVICE DISPLAYED A DEVICE FAILURE DURING PATIENT USE. THERE WERE NO HEALTH CONSEQUENCES FOR THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20936226 ↗MalfunctionOct 15, 2024Human-Device Interface Problem; Output Problem; Human-Device Interface Problem; Output Problem
IT WAS REPORTED THAT THE TOUCH SCREEN WAS NOT WORKING. THERE WERE NO HEALTH CONSEQUENCES FOR THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20449710 ↗MalfunctionSep 30, 2024Decrease in Pressure; Failure to Deliver; Imprecise or Erratic Results; Decrease in Pressure; Failure to Deliver; Imprecise or Erratic Results
THE FOLLOWING WAS REPORTED: VENT FAIL WHILE IN USE - NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20331745 ↗MalfunctionSep 9, 2024Device Difficult to Program or Calibrate; Human-Device Interface Problem; Output Problem; Device Difficult to Program or Calibrate; Human-Device Interface Problem; Output Problem
IT WAS REPORTED THAT THE CONTROL PANEL WAS GLITCHING AND FREEZING UP. THE CONTROL PANEL WAS UNRESPONSIVE. THE DEVICE WOULD NOT SHUT DOWN AND THE USER HAD TO USE THE ROCKER SWITCH. THERE WERE NO HEALTH CONSEQUENCES FOR THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20168825 ↗MalfunctionJun 26, 2024Loss of Data; No Apparent Adverse Event; Unexpected Shutdown; Loss of Data; No Apparent Adverse Event; Unexpected Shutdown
IT WAS REPORTED THAT MACHINE HAD DATA LOSS MESSAGE AND THEN SHUT DOWN WHILE IN USE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19612415 ↗MalfunctionMay 15, 2024Failure to Deliver; Output Problem; Failure to Deliver; Output Problem
IT WAS REPORTED THAT THE DEVICE SHUTDOWN INTERMITTENTLY. NO INJURY REPORTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 19318472 ↗MalfunctionMay 8, 2024No Display/Image; Device Difficult to Program or Calibrate; No Display/Image; Device Difficult to Program or Calibrate
IT WAS REPORTED THAT THE DEVICE DISPLAYED THE ALARM MESSAGE "DEVICE FAILURE 14". NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19266390 ↗MalfunctionMay 7, 2024No Display/Image; Device Difficult to Program or Calibrate; No Display/Image; Device Difficult to Program or Calibrate
IT WAS REPORTED THAT THE DEVICE DISPLAYED THE ALARM MESSAGE "DEVICE FAILURE 14". NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19257974 ↗MalfunctionSep 28, 2022Gas Output Problem; Device Difficult to Program or Calibrate; Failure to Deliver; Gas Output Problem; Device Difficult to Program or Calibrate; Failure to Deliver
IT WAS REPORTED THAT THE SCREEN WENT BLANK, AND THE DEVICE STOPPED VENTILATING DURING USE ON PATIENT, THE DEVICE WAS IMMEDIATELY SWAPPED, AND NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15500607 ↗
FDA clearance and classification
Cleared through 510(k) K222207 on May 16, 2023 (decision: Substantially Equivalent). Submitted by Draegerwerk AG & CO Kgaa as “Babylog VN800, Babylog VN600”. The FDA review took 295 days. FDA record ↗
The FDA regulates product code CBK (ventilator, continuous, facility use) under 21 CFR 868.5895 as a Class II device. 21 CFR 868.5895 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Common questions about the Babylog VN800 8422400
Has the Babylog VN800 8422400 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048675542476), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Babylog VN800 8422400?
The FDA's MAUDE database lists 22 adverse event reports for this device, received between Sep 28, 2022 and Mar 5, 2026. In those reports, the most common specific device problems are “Failure to Deliver” (11), “Device Difficult to Program or Calibrate” (9) and “Output Problem” (5); the most common patient problem is “Respiratory Distress Syndrome of Newborns” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Babylog VN800 8422400?
None of the reports are classified as deaths, 1 as an injury and 21 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Babylog VN800 8422400 FDA approved?
It was cleared through 510(k) K222207 on May 16, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Babylog VN800 8422400 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Babylog VN800 8422400, and is it still sold?
The FDA's device database lists Drägerwerk AG & Co. KGaA as the company responsible for it (brand Babylog VN800, model 8422400). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Ventilator, Continuous, Facility Use devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,148 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,384 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,511 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,287 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls10 deaths1,841 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls9 deaths1,166 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page