Babylog VN600 8422200Recalls & reported problems

Drägerwerk AG & Co. KGaA · Class 2

Quick summary

This Dräger Babylog VN600 ventilator is a continuous ventilator for facility use. Reports mostly describe touchscreen displays not working, making the device hard to program, with some mentions of failure to deliver air and unexpected shutdowns. There are 23 reports that mention no deaths and 1 injury. No recalls name this device, with reports rising to 15 in the last 12 months from 5 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 04048675551713), received between Dec 21, 2023 and Jun 30, 2026 — classified as 1 injury, 22 malfunctions.

The most frequently reported specific device problem is “Device Difficult to Program or Calibrate” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 9 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Difficult to Program or Calibrate14
  • Failure to Deliver6
  • No Display/Image4
  • Unexpected Shutdown3
  • Complete Loss of Power2
  • Failure to Read Input Signal2
  • Gas Output Problem2
  • Activation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • Increased Respiratory Rate1
  • Low Oxygen Saturation1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCHSCREEN OF THE DEVICE WAS MALFUNCTIONING. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25663735 ↗
  • MalfunctionJun 30, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCHSCREEN OF THE DEVICE WAS MALFUNCTIONING. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25663734 ↗
  • MalfunctionJun 30, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCHSCREEN OF THE DEVICE WAS MALFUNCTIONING. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25663732 ↗
  • MalfunctionMay 12, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE WAS NOT FUNCTIONING DURING USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150808 ↗
  • MalfunctionMay 12, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE WAS NOT FUNCTIONING DURING USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150747 ↗
  • MalfunctionMay 12, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE WAS NOT FUNCTIONING DURING USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150743 ↗
  • MalfunctionMay 12, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE WAS NOT FUNCTIONING DURING USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150742 ↗
  • MalfunctionMay 12, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE WAS NOT FUNCTIONING DURING USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150741 ↗
  • MalfunctionMay 11, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE WAS NOT FUNCTIONING DURING USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25135042 ↗
  • MalfunctionMay 11, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE WAS NOT FUNCTIONING DURING USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25134647 ↗
  • MalfunctionMay 11, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE WAS NOT FUNCTIONING DURING USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25134646 ↗
  • MalfunctionMay 11, 2026Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE TOUCH SCREEN OF THE DEVICE WAS NOT FUNCTIONING DURING USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25134643 ↗
  • MalfunctionApr 20, 2026No Display/Image; Failure to Read Input Signal

    IT WAS REPORTED THAT THE DISPLAY OF THE DEVICE SHUT DOWN DURING OPERATION. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24923796 ↗
  • MalfunctionJan 26, 2026Decrease in Pressure; Failure to Deliver; Imprecise or Erratic Results

    IT WAS REPORTED THAT THE DEVICE ISSUED THE ALARM MESSAGE "DEVICE FAILURE 1" AND PERFORMED A REBOOT DURING PATIENT USE. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24176276 ↗
  • MalfunctionNov 26, 2025No Display/Image; Device Difficult to Program or Calibrate; No Display/Image; Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT A 'DEVICE ERROR 14' OCCURRED ON (B)(6) 2025. ON (B)(6) 2025, A 'DEVICE FAILURE 14' WAS POSTED AGAIN. THERE WERE NO PATIENT HEALTH CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23655514 ↗
  • MalfunctionJul 3, 2025Failure to Deliver; Complete Loss of Power; Unexpected Shutdown; Failure to Deliver; Complete Loss of Power; Unexpected Shutdown

    IT WAS REPORTED THAT THE DEVICE SHUT DOWN DURING USE WITH A PATIENT. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22392466 ↗
  • MalfunctionJun 5, 2025Failure to Read Input Signal; Failure to Deliver; Activation Failure; Failure to Read Input Signal; Failure to Deliver; Activation Failure

    THE FOLLOWING ISSUE WAS REPORTED: PROBLEM OF "DEVICE FAILURE 02" ARISES WHILE CONTINUE RUNNING OF VENTILATOR MORE THAN TWO DAYS. BEFORE AND AFTER OF THIS ERROR, SYSTEM TEST IS SUCCESSFUL ALWAYS. ONCE WE RESTART THE VENTILATOR, NO DEVICE FAILURE IS THERE BUT COMES AGAIN WITHIN 2- OR 3-DAYS CONTINUOUS VENTILATION. WE HAVE OBSERVED THIS DEVICE FAILURE 3 TIMES. NO PATIENT HEALTH CONSEQUENCES HAVE BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22148709 ↗
  • MalfunctionMay 8, 2025No Display/Image; Human-Device Interface Problem; Unexpected Shutdown; No Display/Image; Human-Device Interface Problem; Unexpected Shutdown

    IT WAS REPORTED THAT THE DEVICE TURNED OFF WHILE CONNECTED TO A PATIENT. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21983398 ↗
  • MalfunctionMar 7, 2025Failure to Deliver; Complete Loss of Power; Unexpected Shutdown; Failure to Deliver; Complete Loss of Power; Unexpected Shutdown

    IT WAS REPORTED THAT THE DEVICE SUDDENLY SHUT DOWN DURING USE ON A PATIENT. THERE WERE NO HEALTH CONSEQUENCES FOR THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21544696 ↗
  • MalfunctionNov 11, 2024Gas Output Problem; Failure to Deliver; Intermittent Communication Failure; Gas Output Problem; Failure to Deliver; Intermittent Communication Failure

    THE FOLLOWING ISSUE WAS REPORTED: THE CUSTOMER REPORTED AN ISSUE WHERE THE DEVICE SUDDENLY SHUT DOWN DURING OPERATION WHILE CONNECTED TO A PATIENT. THE DEVICE ALARMED. ACCORDING TO THE INITIAL LOG ANALYSIS, THE DEVICE HAD STOPPED UNEXPECTEDLY AT THE REPORTED TIME. NO PAT. HEALTH CONSEQUENCES HAVE BEEN REPORTED. HOWEVER, IN CASE THE BREATHING PRESSURE (INCL. PEEP) DROPS TO AMBIENT, A DETERIORATION IN STATE OF HEALTH CANNOT BE EXCLUDED FOR PATIENTS WITH DIFFICULT LUNG/VENTILATION CONDITIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20650788 ↗
  • MalfunctionAug 12, 2024No Display/Image; Device Difficult to Program or Calibrate; No Display/Image; Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE DEVICE DISPLAYED THE ALARM MESSAGE "DEVICE FAILURE 14" AND ¿ALARM SYSTEM ERROR¿. THE SCREEN OF THE DEVICE BLACKED OUT. NO HEALTH CONSEQUENCES FOR THE PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19961574 ↗
  • InjuryMay 16, 2024Complete Blockage; Gas Output Problem; Obstruction of Flow; Adverse Event Without Identified Device or Use Problem; Complete Blockage; Gas Output Problem; Obstruction of Flow; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PRETERM, SEDATED BABY WAS CONNECTED TO THE VENTILATOR. AT AROUND 03:30 P.M. THE "BABY SUDDENLY DETERIORATED WITH POOR PERFUSION AND DESATURATION. BABY WAS CHECKED AND A CHEST X-RAY WAS DONE, BUT NO PNEUMOTHORAX WAS NOTED. THE VENTILATOR WAS SHOWING A HIGH RESPIRATORY RATE (UP TO 114) AND A PRESSURE AND FLOW RATE CURVE WITH MARKEDLY CHANGED TO FASCICULATION. THE BABY WAS DISCONNECTED FROM THE VENTILATOR AND THE TEST LUNG WAS CONNECTED. THE SAME PROBLEM CONTINUED EVENT WITH THE TEST LUNG.".

    Patient problems: Increased Respiratory Rate; Low Oxygen Saturation

    MAUDE report 19330508 ↗
  • MalfunctionDec 21, 2023Failure to Deliver; Failure to Deliver

    IT WAS REPORTED THAT THE DEVICE SHUT DOWN DURING USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18370903 ↗