Skip to content

VICTUS™ Femtosecond Laser Platform 3.2Recalls & reported problems

Technolas Perfect Vision GmbH · The laser system is an ophthalmologic system based on femtosecond technology. It is used to treat a wide variety of ophthalmologic indications. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04050555002895), received between Sep 22, 2026 and Sep 22, 2026 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Capsular Bag Tear1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem

    On (b)(6) 2026 the surgeon reported a Capsulotomy tag that lead to a tear. On (b)(6) 2026 additional information was received. The surgeon confirmed that parts of the crystalline lens dropped into the Vitreous, and that the patient has been referred to another clinic (this is the awareness date). After further inquiry on (b)(6) 2026 we received the information that the patient had the lens fragments removed and a posterior vitrectomy was performed.

    Patient problems: Capsular Bag Tear

    MAUDE report 50851380 ↗

Other Ophthalmic Femtosecond Laser devices

All Ophthalmic Femtosecond Laser devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page