N/A CS 3825-01Recalls & reported problems
Ulrich GmbH & Co. KG · Screwdriver, open, cannulated, hex 3.5 mm coupling A · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04052536081202), received between Nov 30, 2016 and Sep 26, 2019 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K123717); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Fragmentation3
- Break1
Patient problems
- No Known Impact Or Consequence To Patient3
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 26, 2019Material Fragmentation; Material Fragmentation
WE RECEIVED A COMPLAINT THAT THE SCREWDRIVER HAD SHEARED OFF DURING THE CASE. WHEN THE PRODUCT WAS RECEIVED BACK WE DISCOVERED THE TIP HAD BROKEN OFF.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9124031 ↗InjuryJun 28, 2019Material Fragmentation; Material Fragmentation
THE SURGEON WAS NOT ABLE TO SUCCESSFULLY IMPLANT THE SCREW; TWO SCREWDRIVERS BROKE WHILE TRYING BUT REMAINED INTACT. HE DECIDED TO REMOVE THE SCREW AND BROKE ANOTHER SCREWDRIVER TRYING TO DO SO AND SEVERAL PIECES OF METAL BROKE OFF FROM THE SCREWDRIVER BUT WERE RECOVERED AND RETURNED ALONG WITH THE DRIVERS. THE SURGEON WAS ABLE TO REMOVE THE SCREW USING A ROD AND LOCKING SCREW. REPORT 3/5.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8745565 ↗InjuryJun 28, 2019Material Fragmentation; Material Fragmentation
THE SURGEON WAS NOT ABLE TO SUCCESSFULLY IMPLANT THE SCREW; TWO SCREWDRIVERS BROKE WHILE TRYING BUT REMAINED INTACT. HE DECIDED TO REMOVE THE SCREW AND BROKE ANOTHER SCREWDRIVER TRYING TO DO SO AND SEVERAL PIECES OF METAL BROKE OFF FROM THE SCREWDRIVER BUT WERE RECOVERED AND RETURNED ALONG WITH THE DRIVERS. THE SURGEON WAS ABLE TO REMOVE THE SCREW USING A ROD AND LOCKING SCREW. REPORT 2/5.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8745550 ↗MalfunctionNov 30, 2016Break; Break
SCREWDRIVER TIP BROKE OFF WHILE IN SURGERY, WAS NOT ABLE TO BE RETRIEVED, AND REMAINS IN THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6135004 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123717. One submission often covers many models, so check the product and lot details in each record.
- Z-1126-2019 · Mar 8, 2019 · TerminatedULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.
Other Orthosis, spondylolisthesis spinal fixation devices
- CD HORIZON® Spinal System 7540020Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation257 reports
- MANTIS Redux 48289999Stryker Corporation · Orthosis, spondylolisthesis spinal fixation76 reports
- EXPEDIUM TOP NOTCH 179771555Depuy Spine, LLC · Orthosis, spondylolisthesis spinal fixation41 reports
- EXPEDIUM 179712650Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation31 reports
- EXPEDIUM VERSE 199721530Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4 5 48130000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4.5 System - XIA CT 48850000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation27 reports
- CD HORIZON® Spinal System 7576545Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation26 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page