neon³ CT spinal stabilization CS 3902-35-14Recalls & reported problems
Ulrich GmbH & Co. KG · Screw, Ø 3.5 mm, length 14 mm · Class 2 · Implant
All models of neon³ CT spinal stabilization (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04052536084197), received between Jun 21, 2018 and Jun 5, 2025 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K150650, K161032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Fracture2
Patient problems
- Failure of Implant3
- No Known Impact Or Consequence To Patient2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJun 5, 2025Fracture; Fracture
A PEDICLE SCREW BROKE DURING SURGERY; PART OF IT HAD TO BE LEFT IN THE PATIENT. THE RESULT WAS A DELAY IN THE OPERATING TIME.
Patient problems: Failure of Implant
MAUDE report 22154028 ↗MalfunctionMay 14, 2025Fracture; Fracture
"BREAKAGE OF A PEDICLE-SCREW DURING SURGERY. THIS RESULTS IN A DELAY IN THE OPERATION TIME. ACCORDING TO REPORTING PERSON: "WE ARE VERY SORRY TO INFORM YOU THAT THE ITEM BROKE DURING THE SURGERY AND PART OF IT HAS BEEN LEFT INSIDE THE PATIENT.""
Patient problems: Failure of Implant
MAUDE report 22018886 ↗InjuryDec 3, 2018Break; Break
A PEDIATRIC PATIENT WHO HAD AN OCCIPITAL FIXATION SURGERY IN JULY HAD A REVISION SURGERY IN SEPTEMBER. HE CAME BACK TO THE HOSPITAL AND APPARENTLY THE SYSTEM WAS BROKEN. WHEN THE DOCTOR OPENED THE PATIENT IN THE REVISION SURGERY, HE SAW THAT THE RODS AND SCREWS WERE BROKEN, SO THEY REPLACED THE WHOLE SYSTEM AND ADDED MORE LEVELS.
Patient problems: Failure of Implant
MAUDE report 8126367 ↗InjurySep 24, 2018Break; Break
LATERAL MASS SCREW WAS INSERTED AS PER STANDARD TECHNIQUE (PRE-DRILLING). AS THE SCREW WAS DRIVEN INTO THE BONE, THE SCREW SHANK BROKE AWAY FROM THE HEAD OF THE SCREW. THE SCREW SHANK REMAINED IN THE PATIENT. THE REMAINS OF THE SCREW WERE RETURNED.
Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient
MAUDE report 7904266 ↗InjuryJun 21, 2018Break; Break
TULIP HEAD OF SCREW BROKE DURING IMPLANTATION. THE SCREW WAS REMOVED AND REPLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7626972 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K150650, K161032. One submission often covers many models, so check the product and lot details in each record.
- Z-1350-2020 · Dec 31, 2019 · TerminatedUlrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw
- Z-1126-2019 · Mar 8, 2019 · TerminatedULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.
Other Orthosis, cervical pedicle screw spinal fixation devices
- INFINITY™ Occipitocervical Upper Thoracic System 3604026Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation2 recalls1 report
- CONNECTOR 79-2002Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604532Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604024Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- CONNECTOR 79-2003Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 360PT4022Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- OASYS 48551000Stryker Corporation · Orthosis, cervical pedicle screw spinal fixation45 reports
- SYMPHONY 1020-35-112Medos International Sàrl · Posterior cervical screw system6 deaths15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page