Richard Wolf GmbH · LASER WORKING INSERT 12/30 round, working channel for instruments up to 6.5Fr, compatible with sheaths 8675322, 8675324, 8675524, 8632030, and telescopes Ø 4mm, with introduction port, bayonet-cap, reusable · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04055207018194), received between Apr 1, 2021 and Jul 1, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Difficult to Remove1
- Fracture1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 1, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
RICHARD WOLF GMBH COMPLAINT REFERENCE (B)(4). ACCORDING TO THE USER FACILITY: "BLADDER PERFORATION WITH THE LASER WORKING INSERT. COULD NOT BE COMPLETED ENDOSCOPICALLY, URGENT LAPAROTOMY WAS PERFORMED TO REPAIR THE PATIENT'S BLADDER." THE USER FACILITY HAS NOT GIVEN INFORMATION ABOUT THE EXACT PRODUCT PROBLEM, THEY HAVE ONLY SHARED THE INFORMATION ABOUT THE HARM OF THE PATIENT. THE INVESTIGATION IS PENDING DUE TO NOT RECEIVING THE DEVICE AT RICHARD WOLF GMBH IN (B)(4). RWGMBH AWARENESS DATE, 13-JUN-2022.
Patient problems: Insufficient Information
MAUDE report 14887201 ↗MalfunctionApr 1, 2021Fracture; Difficult to Remove; Insufficient Device Problem Information; Fracture; Difficult to Remove; Insufficient Device Problem Information
RICHARD WOLF GMBH REPORTS THAT: THE CUSTOMER SENT US THE LASER WORKING INSERT TYPE 8632.284. IN SPITE OF REPEATED REQUESTS, THE ATTACHMENT FOR CYSTOSCOPE TYPE 8650264 WAS NOT SENT TO US, SO THAT ONLY THE LASER WORKING INSERT TYPE 8632.284 COULD BE EXAMINED. THE PRESENT LASER WORKING INSERT TYPE 8632.284 ORIGINATES FROM THE BATCH: 1249546 AND WAS BOOKED TO THE NEW STOCK ON JUNE 10, 2014 WITH AN ORDER SIZE OF 15 PIECES. THE CHECK OF THE FLOW CHART DID NOT SHOW ANY ANOMALIES. THE DELIVERY TO THE CUSTOMER TOOK PLACE ON 05.11.2014. THE INSPECTION OF THE LASER WORK INSERT IN THE SPECIALIST DEPARTMENT DID NOT REVEAL ANY FUNCTIONAL DEFECT. HOWEVER, THE LASER WORK INSERT HAS CLEAR SIGNS OF USE DUE TO ITS AGE AND USE AND IS THERMALLY ATTACKED BY THE USE OF THE LASER. THE CAUSE CAN BE TRACED BACK TO USER ERROR. WITH REGARD TO SEVERAL NUMBER OF PIECES, WORLDWIDE DELIVERED WORKING INSERTS AND NO COMPARABLE COMPLAINT CASE OF THIS TYPE, A PRODUCT PROBLEM OR A MANUFACTURING PRODUCT PROBLEM OR A MANUFACTURING DEFECT CAN BE RULED OUT. THIS IS FURTHERMORE CONFIRMED BY THE FACT THAT THE PRESENT APPROACH HAS ALREADY BEEN DELIVERED FOR 7 YEARS. THUS, NO "EARLY FAILURE" DUE TO A PRODUCT DEFECT IS EVIDENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11603107 ↗
Other CYSTOURETHROSCOPE devices
- n.a. 11272C1Karl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall16 reports
- n.a. 27050CAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE54 reports
- n.a. 11272VNUKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall4 reports
- C-VIEW 11272VUKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- n.a. 11272CU1Karl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VUAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
- C-VIEW 11272VAKarl Storz GmbH & Co. KG · CYSTOURETHROSCOPE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page