SHARK 8675324Recalls & reported problems
Richard Wolf GmbH · INTERNAL SHEATH RESECTOSCOPE 24FR color code yellow, SL 205mm, round, rotatable, slanted distal end, ceramic, continuous irrigation, compatible with Shark outer sheath 8675426, and telescopes Ø 4mm, 0°/12°/30°, Snap-On lock, reusable · Class 2
All models of SHARK (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Richard Wolf Shark resectoscope sheath is a reusable inner sheath used in prostate and bladder procedures. Reports mostly describe the ceramic tip breaking or detaching inside the patient, with pieces retrieved during surgery and procedures extended or interrupted. There are 25 reports that name its identifier, with no deaths and 8 injuries. No recalls name this device, and reports fell to 1 in the last year from 7 the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 27
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 04055207019658), received between Jun 17, 2019 and Sep 16, 2026 — classified as 8 injuries, 19 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (17 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation17
- Break6
- Detachment of Device or Device Component5
- Insufficient Device Problem Information3
- Material Separation3
Patient problems
- Insufficient Information9
- No Clinical Signs, Symptoms or Conditions9
- Foreign Body In Patient7
- Hemorrhage/Blood Loss/Bleeding1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Peritoneal Laceration(s)1
Recent adverse event reports
The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Material Fragmentation
Patient problems: Insufficient Information
MAUDE report 50794841 ↗MalfunctionSep 16, 2026Material Fragmentation
Patient problems: Insufficient Information
MAUDE report 50794812 ↗MalfunctionDec 11, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING THE HOLEP (HOLMIUM LASER ENUCLEATION OF THE PROSTATE) PROCEDURE, THE CERAMIC TIP BROKE OFF INSIDE THE PATIENT. THE SURGERY WAS ONLY EXTENDED BY 20 MINUTES BY RETRIEVING THE BROKEN PIECE, AND THE BROKEN CERAMIC TIP WAS RECOVERED SUCCESSFULLY. THE SCHEDULED PROCEDURE WAS COMPLETED.
Patient problems: Insufficient Information
MAUDE report 23780063 ↗MalfunctionJul 18, 2025Break; Break
IT WAS REPORTED THAT DURING A LASER ENUCLEATION OF THE PROSTATE, THE INNER SHAFT BROKE OFF. THE OPERATION WAS COMPLETED WITHOUT HARM TO THE PATIENT AND WITHOUT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22542248 ↗MalfunctionJul 18, 2025Break; Break
IT WAS REPORTED THAT DURING A LASER ENUCLEATION OF THE PROSTATE, THE INNER SHAFT BROKE OFF. THE OPERATION WAS COMPLETED WITHOUT HARM TO THE PATIENT AND WITHOUT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22541806 ↗InjuryJul 14, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A MONOPOLAR TRANSURETHRAL RESECTION OF PROSTATE, THE CERAMIC TIP DETACHED. THE SURGERY WAS UNABLE TO BE COMPLETED.
Patient problems: Foreign Body In Patient
MAUDE report 22495864 ↗MalfunctionJun 25, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING THE URETEROSCOPY WITH TUMOR ABLATION, THE CERAMIC TIP BROKE OFF INSIDE THE PATIENT'S BLADDER. THE REPORTED ISSUE CAUSED AN HOUR DELAY IN THE PROCEDURE WHILE TROUBLESHOOTING HOW TO RETRIEVE THE PIECE OUT OF THE PATIENT. THE PHYSICIAN USED A LASER AND STONE BASKET TO REMOVE THE FRAGMENTS. THE DELAY DID NOT CAUSE ANY INJURIES OR UNACCEPTABLE CLINICAL RISKS TO THE PATIENT OR OTHER PERSONNEL, AND THE SCHEDULED PROCEDURE WAS SUCCESSFULLY COMPLETED.
Patient problems: Foreign Body In Patient
MAUDE report 22330542 ↗MalfunctionMar 19, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING THE ENDOSCOPIC ENUCLEATION OF A PROSTATIC ADENOMA, ONCE THE RESECTOSCOPE WAS INSERTED THROUGH THE VESICAL NECK, THE DISTAL PORTION OF THE URETHRAL MEMBRANE IN CERAMIC WAS DETACHED. THERE WAS A BRIEF INTERRUPTION IN THE PROCEDURE FOR THE REMOVAL OF THE DISTAL DETACHMENT WHICH WAS EXTRACTED WITH DIFFICULTY BY ENDOSCOPIC FORCEPS." ACCORDING TO THE USER, THE BROKEN PIECE WAS COMPLETELY REMOVED, AND THE PROCEDURE COULD BE COMPLETED WITHOUT ANY CLINICALLY RELEVANT DELAY IN SURGERY. THE MEDICAL PURPOSE OF THE APPLICATION WAS FULFILLED AND THERE WAS NO DANGER FOR THE PATIENT, THE USER OR THIRD PARTIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21644464 ↗MalfunctionMar 3, 2025Material Fragmentation; Material Fragmentation
ACCORDING TO THE RECEIVED INFORMATION, "DR. WAS USING THE RESECTION SCOPE FOR A TRANSURETHRAL RESECTION OF PROSTATE PROCEDURE. ABOUT 5 MINUTES INTO RESECTING, THE TIP OF THE SHEATH BROKE OFF INTO THE PATIENT'S BLADDER. THE SURGEON SAFELY REMOVED THE TIP FROM THE BLADDER AND NO INJURY WAS NOTED.".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21509615 ↗InjuryFeb 10, 2025Material Fragmentation; Material Fragmentation
ACCORDING TO THE RECEIVED INFORMATION, "DURING THE OPERATION, PART OF THE INSERTION TIP OF THE OBTURATOR BROKE OFF / SHATTERED IN THE URINARY BLADDER. " DURING THE RETRIEVAL, THE TIP PARTS WERE CUT / DIVIDED AGAIN BY THE SURGEON DUE TO THEIR SIZE, SO THAT RETRIEVAL WAS POSSIBLE. PRESUMABLY, ALL PARTS WERE FULLY RETRIEVED. THE FRAGMENTATION AND RETRIEVAL OF THE FRAGMENT PROLONGED THE PROCEDURE BY ABOUT 40 MINUTES. THE SCHEDULED SURGICAL PROCEDURE WAS COMPLETED, AND NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21345083 ↗MalfunctionSep 20, 2024Material Fragmentation; Material Fragmentation
RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) RECEIVED A MEDWATCH REPORT, MW5158447, REGARDING AN ISSUE WITH AN INTERNAL SHEATH RESECTOSCOPE 24FR, PART ID: 8675324, BATCH# 1562220. ACCORDING TO THE RECEIVED INFORMATION, "DURING THE PROCEDURE, THE DOCTOR IDENTIFIED THE CERAMIC TIP HAD BROKEN OFF AND WAS IN THE BLADDER. THE BROKEN ITEM WAS REMOVED AND THE OR MANAGER WAS NOTIFIED. THE DOCTOR VIEWED THE REMOVED CERAMIC TIP WITH THE ORIGINAL PART AND DETERMINED IT WAS COMPLETE. NO EVIDENCE OF ANY OTHER FOREIGN OBJECT WAS IDENTIFIED. THE BLADDER WAS IRRIGATED AND DRAINED PRIOR TO THE COMPLETION OF THE CASE." LATER INFORMATION OBTAINED, THAT THE REPORTED ISSUE CAUSED A 5-10 MINUTES DELAY DURING THE PROCEDURE WHILE THE SURGEON RETRIEVED THE BROKEN PIECE. HOWEVER, NO APPARENT HARM OCCURED IN RELATION TO THE ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20270725 ↗MalfunctionSep 20, 2024Material Fragmentation; Material Fragmentation
RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) RECEIVED A MEDWATCH REPORT, MW5158444, REGARDING AN ISSUE WITH AN INTERNAL SHEATH RESECTOSCOPE 24FR, PART ID: 8675324, BATCH# 1562242. ACCORDING TO THE RECEIVED INFORMATION, "SURGEON JUST STARTED TO PROCEED WITH THE PROCEDURE. THE STAFF IN THE ROOM NOTICED THAT THE END OF THE RESECTION SCOPE BROKE OFF INTO THE PATIENT'S BLADDER. THE PIECE OF THE SCOPE WAS REMOVED AND NO TISSUE TRAUMA WAS PRESENT TO THE PATIENT." LATER INFORMATION OBTAINED, THAT THE REPORTED ISSUE CAUSED A 5-10 MINUTES DELAY DURING THE PROCEDURE WHILE THE SURGEON RETRIEVED THE BROKEN PIECE. HOWEVER, NO APPARENT HARM OCCURED IN RELATION TO THE ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20269090 ↗InjuryAug 22, 2024Material Fragmentation; Material Fragmentation
RICHARD WOLF GMBH HAS BEEN INFORMED OF AN ISSUE REGARDING AN INTERNAL SHEATH RESECTOSCOPE 24FR, PART ID: 8675324, BATCH# 1533973. ACCORDING TO THE RECEIVED INFORMATION, DURING A HOLMIUM LASER ENUCLEATION OF THE PROSTATE (HOLEP) PROCEDURE, THE CERAMIC TIP ON THE INTERNAL SHEATH RESECTOSCOPE 24FR, PART ID: 8675324 WITH BATCH 1533973 BECAME LOOSE OR BROKE OFF INTRAOPERATIVELY. HOWEVER, THIS WAS ONLY NOTICED AFTER THE PROCEDURE. THE CLINIC WILL ARRANGE FOR THE PATIENT TO HAVE X-RAYS TAKEN AGAIN.
Patient problems: Insufficient Information
MAUDE report 20027561 ↗MalfunctionDec 12, 2023Material Fragmentation; Material Fragmentation
THE USER FACILITY HAS INFORMED RICHARD WOLF GMBH (RWGMBH) OF AN ISSUE REGARDING AN INTERNAL SHEATH RESECTOSCOPE 24FR, PART ID: 8675324, LOT # 1304359. ACCORDING TO THE RECEIVED INFORMATION, "AT THE BEGINNING OF THE HOLMIUM LASER ENUCLEATION OF THE PROSTATE, THE SURGEON NOTICED DURING CYSTOSCOPY THAT THE CERAMIC TIP OF THE INNER SHAFT WAS FLOATING IN THE BLADDER. CONSIDERATION WAS GIVEN TO HOW THIS TIP COULD BE REMOVED FROM THE BLADDER WITHOUT DAMAGING THE URETHRA. THE CERAMIC TIP WAS GRASPED AT THE BLUNTER END AND EXTRACTED FROM THE BLADDER. THE PATIENT WAS NOT HARMED."
Patient problems: Insufficient Information
MAUDE report 18312813 ↗InjuryJul 26, 2023Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component
ACCORDING TO THE RECEIVED INFORMATION FROM THE CUSTOMER, THERE WAS A "BREAKAGE OF THE BAKELITE INNER JACKET AT THE JUNCTION OF THE INSTRUMENT (THE METAL TUBE SEPARATED FROM THE CIRCULAR LOCK WHEN CHANGING THE INSTRUMENTATION AFTER ENUCLEATION TO SWITCH TO MORCELLATION. APRIORI NO FORCEFUL GESTURE AT THE INTRODUCTION OF THE CAMERA." THE DETACHED PART OF THE INNER SHEATH (METAL TUBE) FELL INTO THE BLADDER AND PERFORATED THE BOTTOM OF THE BLADDER INTO THE PATIENT'S PERITONEUM. LT WAS NECESSARY TO COMPLETE THE OPERATION BY LAPAROTOMY FOR ABDOMINAL EXPLORATION AND SUTURE THE BLADDER. ADDITIONALLY, THERE WAS NO REPORTED DIGESTIVE PERFORATION. THE LAPAROTOMY TOOK LONGER THAN EXPECTED AND AS A CONSEQUENCE, THERE WAS A REPORTED DELAY OF BETWEEN 15 AND 30 MINUTES. PATIENT MEDICAL TREATMENT REQUIRED AN EXTENSION OF STAY AT THE HOSPITAL FOR 6 DAYS. THIS CASE WAS REPORTED BY THE CUSTOMER TO THE FRENCH NATIONAL AGENCY FOR THE SAFETY OF MEDICINES AND HEALTH PRODUCTS (ANSM).
Patient problems: Peritoneal Laceration(s)
MAUDE report 17396928 ↗InjuryJun 30, 2023Material Separation; Material Separation
A USER FACILITY HAS INFORMED RICHARD WOLF GMBH AN ISSUE REGARDING A INTERNAL SHEATH RESECTOSCOPE 24FR, PART ID: 8675324, SERIAL # (B)(6). ACCORDING TO THE RECEIVED INFORMATION, "DURING SURGERY, SUDDEN SHEARING OFF OF AN APPROX. 1 CM LONG CYLINDRICAL FOREIGN BODY CORRESPONDING TO THE INSIDE OF THE RESECTOSCOPE SHAFT. DUE TO INCORRECT USE OF THE SNARE RESECTION, THIS DEFECT WAS ONLY DETECTED OPTICALLY AFTERWARDS. THE FOREIGN BODY WAS THEN REMOVED ENDOSCOPICALLY WITH GRASPING FORCEPS. BY CHANGING THE SURGICAL TECHNIQUE, THE PATIENT COULD BE TREATED WITH A PROLONGED SURGICAL TIME AS INITIALLY."
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 17240803 ↗MalfunctionJan 24, 2023Break; Material Separation; Break; Material Separation
RWGMBH REFERENCE COMPLAINT NO. (B)(4). THE PURPOSE OF THIS SUBMISSION IS TO PROVIDE A CORRECTION ON G3: DATE RECEIVED BY MANUFACTURER. SEE H10 FOR FURTHER EXPLANATION.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 16235640 ↗InjuryOct 12, 2022Material Fragmentation; Material Fragmentation
RWGMBH REFERENCE COMPLAINT NO. (B)(4). RICHARD WOLF GMBH IN KNITTLINGEN (MANUFACTURER) WAS INFORMED ABOUT A USER NOTIFICATION TO THE ITALIAN COMPETENT AUTHORITY VIA DISTRIBUTOR. DESCRIPTION OF INCIDENT/CLAIM: DURING A THULEP PROCEDURE, A COMPONENT OF THE BIPOLAR RESECTOSCOPE WAS APPARENTLY LOST IN THE PATIENT, BUT THIS WAS ONLY NOTICED AFTER THE PROCEDURE WAS COMPLETED. THE PATIENT THEREFORE HAD TO UNDERGO ANOTHER MINIMALLY INVASIVE SURGERY 3 DAYS LATER TO HAVE THE FOREIGN BODY REMOVED. THIS CASE WAS REPORTED BY THE USER TO THE ITALIAN AUTHORITIES. RWGMBH MDR AWARENESS DATE: 23-SEPT-2022
Patient problems: Foreign Body In Patient
MAUDE report 15587820 ↗InjuryJul 22, 2022Break; Material Fragmentation; Break; Material Fragmentation
RWMIC REFERENCE COMPLAINT NO. (B)(4). IT WAS REPORTED TO RICHARD WOLF MEDICAL INSTRUMENTS CORP. (RWMIC) THAT "TIP BROKE OFF DURING A PROCEDURE. DELAY TIME WAS 15 MINUTES." ADDITIONAL INFORMATION REPORTED BY THE INITIAL REPORTER: WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTING ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? NO. WAS THERE ANY INJURY OR ILLNESS TO ANY OTHER PERSONNEL DUE TO THE REPORTED ISSUE? NO. DID THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED? YES. DID THE DELAY PUT THE PATIENT AT RISK? NO. WAS THERE A SIMILAR BACK-UP DEVICE AVAILABLE FOR USE? YES. WAS THE SCHEDULED PROCEDURE COMPLETED? YES. ADDITIONAL DETAILS ABOUT THE ADVERSE EVENT; (B)(6) 2022. WHAT WAS THE SCHEDULED PROCEDURE BEING PERFORMED WHEN THE TIP BROKE? TRANSURETHRAL INCISION OF THE PROSTATE (TUIP). DID THE TIP BREAK INSIDE OR OUTSIDE OF THE PATIENT? YES. WHAT ACTIONS WERE BEING DONE DURING THE DELAY? REMOVE THE TIP OF THE SHEATH FROM THE PATIENT. DID THE PATIENT REQUIRE ANY ADDITIONAL TREATMENT DURING THE DELAY OR WILL REQUIRE ANOTHER PROCEDURE AS THE RESULT OF THE DELAY? NO. WHAT WAS THE PATIENT'S OU
Patient problems: Foreign Body In Patient
MAUDE report 15084792 ↗MalfunctionJul 22, 2022Break; Material Fragmentation; Break; Material Fragmentation
RICHARD WOLF MEDICAL INSTRUMENTS CORP. (RWMIC) REFERENCE (B)(4). IT WAS REPORTED TO RICHARD WOLF MEDICAL INSTRUMENTS CORP. (RWMIC) THAT "TIP BROKE OFF DURING A PROCEDURE. DELAY TIME WAS 15 MINUTES." ADDITIONAL INFORMATION REPORTED BY THE INITIAL REPORTER: WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTING ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? NO. WAS THERE ANY INJURY OR ILLNESS TO ANY OTHER PERSONNEL DUE TO THE REPORTED ISSUE? NO. DID THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED? YES. DID THE DELAY PUT THE PATIENT AT RISK? NO. WAS THERE A SIMILAR BACK-UP DEVICE AVAILABLE FOR USE? YES. WAS THE SCHEDULED PROCEDURE COMPLETED? YES. ADDITIONAL DETAILS ABOUT THE ADVERSE EVENT; (B)(6) 2022. (1) WHAT WAS THE SCHEDULED PROCEDURE BEING PERFORMED WHEN THE TIP BROKE? TRANSURETHRAL INCISION OF THE PROSTATE (TUIP). (2) DID THE TIP BREAK INSIDE OR OUTSIDE OF THE PATIENT? YES. (3) WHAT ACTIONS WERE BEING DONE DURING THE DELAY? REMOVE THE TIP OF THE SHEATH FROM THE PATIENT. (4) DID THE PATIENT REQUIRE ANY ADDITIONAL TREATMENT DURING THE DELAY OR WILL REQUIRE ANOTHER PROCEDURE AS THE RESULT
Patient problems: Foreign Body In Patient
MAUDE report 15084780 ↗MalfunctionJul 6, 2022Detachment of Device or Device Component; Insufficient Device Problem Information; Detachment of Device or Device Component; Insufficient Device Problem Information
RICHARD WOLF GMBH COMPLAINT REFERENCE (B)(4). PLANNED SURGERY: VIDEO MINIMALLY INVASIVE RTU BIPOLAR PROCEDURE. DURING A VIDEO MINIMALLY INVASIVE RTU BIPOLAR PROCEDURE, THE SHARK RESECTOSCOPE SHEATH 8675324 BROKE THE CERAMIC TIP, WHICH WAS LOST IN THE PATIENT'S ABDOMINAL CAVITY. IT WAS NECESSARY TO PERFORM AN OPEN AND MORE COMPLEX PROCEDURE TO REMOVE THE LOOSE PART. RWGMBH AWARENESS DATE: 23-JUN-2022.
Patient problems: Insufficient Information
MAUDE report 14932666 ↗MalfunctionJun 23, 2022Material Fragmentation; Material Fragmentation
RICHARD WOLF GMBH COMPLAINT REFERENCE (B)(4). PLANNED SURGERY: HOLEP. IMMEDIATELY AFTER INSERTION OF THE INSTRUMENT INTO THE BLADDER, A PART OF THE CERAMIC TIP BROKE OFF. AFTER ABOUT 30 MINUTES, THE REST OF THE CERAMIC TIP BROKE OFF. DUE TO THE RETRIEVAL AND REPLACEMENT OF THE RESECTOSCOPE SHAFT, THE SURGERY TIME WAS EXTENDED BY APPROXIMATELY 45 MINUTES. THE PATIENT SUFFERED AN INJURY TO THE URETHRA AND BLADDER MUCOSA. THERE WAS PROLONGED POSTOPERATIVE BLEEDING. RWGMBH AWARENESS DATE, 20-JUN-2022.
Patient problems: Foreign Body In Patient
MAUDE report 14791629 ↗MalfunctionMay 4, 2022Detachment of Device or Device Component; Insufficient Device Problem Information; Detachment of Device or Device Component; Insufficient Device Problem Information
RICHARD WOLF MEDICAL INSTRUMENTS RECEIVED A CALL ON (B)(6) 2022 AT 11:15 AM FROM THE WIFE (CC) OF A PATIENT (WC) UNDERGOING A PROCEDURE. HERE ARE THE NOTES FROM THAT CALL: DR. (B)(6) PERFORMED A SURGICAL PROCEDURE ON (B)(6) 2022 TO ADDRESS WC'S BLADDER PROBLEM. A RICHARD WOLF RESECTOSCOPE WAS BEING USED TO BIOPSY TISSUE THAT WAS SUSPECT TO BE CANCEROUS. THE PATIENT WAS INFORMED THAT THE TIP HAD BROKEN OFF DURING THE PROCEDURE AND IT TOOK THE SURGEON ~ 1 HOUR TO SUCCESSFULLY RETRIEVE THE PIECE. CC DID NOT KNOW THE SPECIFIC MODEL ONLY THAT IT WAS MADE BY RICHARD WOLF. THE PROCEDURE WAS UNSUCCESSFUL TO RETRIEVE A VIABLE BIOPSY. THE PATIENT REPORTEDLY BLED FOR 3 WEEKS POST-SURGERY. 2ND PROCEDURE WAS PERFORMED ON (B)(6) 2022 BY THE SAME SURGEON. SPECIAL INSTRUMENTATION THE SURGEON HAD USED IN ANOTHER FACILITY WAS USED. CC DID NOT KNOW WHAT INSTRUMENTATION WAS USED NOR THE MANUFACTURER. PROCEDURE WAS SUCCESSFUL. PATIENT IS RECOVERING AND DOING WELL. ADDITIONAL INFORMATION PROVIDED BY THE HOSPITAL: WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTING ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? UNKNOW
Patient problems: Insufficient Information
MAUDE report 14282160 ↗InjuryMay 4, 2022Detachment of Device or Device Component; Insufficient Device Problem Information; Detachment of Device or Device Component; Insufficient Device Problem Information
RICHARD WOLF MEDICAL INSTRUMENTS RECEIVED A CALL ON (B)(6) 2022 AT 11:15 AM FROM THE WIFE (B)(6) OF A PATIENT (B)(6) UNDERGOING A PROCEDURE. HERE ARE THE NOTES FROM THAT CALL: DR. (B)(6) PERFORMED A SURGICAL PROCEDURE ON (B)(6) 2022 TO ADDRESS (B)(6) BLADDER PROBLEM. A RICHARD WOLF RESECTOSCOPE WAS BEING USED TO BIOPSY TISSUE THAT WAS SUSPECT TO BE CANCEROUS. THE PATIENT WAS INFORMED THAT THE TIP HAD BROKEN OFF DURING THE PROCEDURE AND IT TOOK THE SURGEON ~ 1 HOUR TO SUCCESSFULLY RETRIEVE THE PIECE. (B)(6) DID NOT KNOW THE SPECIFIC MODEL ONLY THAT IT WAS MADE BY RICHARD WOLF. THE PROCEDURE WAS UNSUCCESSFUL TO RETRIEVE A VIABLE BIOPSY. THE PATIENT REPORTEDLY BLED FOR 3 WEEKS POST-SURGERY. 2ND PROCEDURE WAS PERFORMED ON (B)(6) 2022 BY THE SAME SURGEON. SPECIAL INSTRUMENTATION THE SURGEON HAD USED IN ANOTHER FACILITY WAS USED. (B)(6) DID NOT KNOW WHAT INSTRUMENTATION WAS USED NOR THE MANUFACTURER. PROCEDURE WAS SUCCESSFUL. PATIENT IS RECOVERING AND DOING WELL. ADDITIONAL INFORMATION PROVIDED BY THE HOSPITAL: WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTING ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED
Patient problems: Insufficient Information
MAUDE report 14274135 ↗MalfunctionJul 14, 2021Material Fragmentation; Material Fragmentation
AFTER AN HOUR TURP BIPOLAR SURGERY, THE CASE WAS ALMOST FINISHED. WHEN THEY WERE GOING TO USE ELICK TO REMOVE THE RESIDUE OF PROSTATE, THEY FOUND A BLACK METAL CRACK INSIDE THE PROSTATE. THEY REMOVED THE INNER SHEATH AND CHECKED, THEY FOUND THE WHOLE TIP OF THE INNER SHEATH WAS GONE. THEY EXAMINED INSIDE THE BLADDER, THE TIP HAS BEEN BROKEN INTO 2 PIECES, THEY USED A FORCEPS TO GRASP AND TRY TO REMOVE FROM PATIENT, HOWEVER, AS THE URETHRA OF PATIENT WAS VERY NARROW, THE GRASPER DID NOT HAVE ENOUGH FORCE TO REMOVE IT. FINALLY, IT TOOK ALMOST 1.5 HOURS TO REMOVE 2 PIECES OF METAL FROM PATIENT AND IT SHOULD BE A COMPLETE BROKEN PARTS REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12166582 ↗
FDA clearance and classification
The FDA regulates product code FJL (resectoscope) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the SHARK 8675324
Has the SHARK 8675324 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04055207019658), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SHARK 8675324?
The FDA's MAUDE database lists 27 adverse event reports for this device, received between Jun 17, 2019 and Sep 16, 2026. In those reports, the most common specific device problems are “Material Fragmentation” (17), “Break” (6) and “Detachment of Device or Device Component” (5); the most common patient problems are “Foreign Body In Patient” (7), “Hemorrhage/Blood Loss/Bleeding” (1) and “Peritoneal Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SHARK 8675324?
None of the reports are classified as deaths, 8 as injuries and 19 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the SHARK 8675324, and is it still sold?
The FDA's device database lists Richard Wolf GmbH as the company responsible for it (brand SHARK, model 8675324). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other RESECTOSCOPE devices
- Olympus A22040AOLYMPUS Winter & Ibe GmbH · Resectoscope1 recall1,755 reports
- Olympus A42011AOLYMPUS Winter & Ibe GmbH · Resectoscope1 recall960 reports
- Olympus A22041AOLYMPUS Winter & Ibe GmbH · Resectoscope1 recall654 reports
- Olympus A22042AOLYMPUS Winter & Ibe GmbH · Resectoscope1 recall55 reports
- Olmypus WA22017AOLYMPUS Winter & Ibe GmbH · Resectoscope1 recall31 reports
- Olympus A22040TOLYMPUS Winter & Ibe GmbH · Resectoscope1 recall23 reports
- Olympus A22041TOLYMPUS Winter & Ibe GmbH · Resectoscope1 recall11 reports
- Olympus A22014AOLYMPUS Winter & Ibe GmbH · Resectoscope1 recall11 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page