Skip to content

PANOVIEW 8970402Recalls & reported problems

Richard Wolf GmbH · MORCESCOPE 12° 24.5FR WL 199MM rigid, with parallel ocular lens, working channel for instruments up to 4.8mm, irrigation via 24.5/26Ch sheath and attachement, suction via morcellator, TL 387mm, rod lens system · Class 2

All models of PANOVIEW (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04055207023525), received between Dec 27, 2021 and Sep 10, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent1
  • Optical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING THE LASER ENUCLEATION OF THE PROSTATE (HOLEP) OF THE PATIENT, THE DOCTOR HAD DIFFICULTY IN GETTING TISSUE IN PART OF THE BLADDER NECK OF THE SCOPE. HE WAS WRENCHING HARD AROUND THE PROSTATE AND STILL COULD NOT REACH IT, SO HE PULLED OUT THE SCOPE AND DISCOVERED IT WAS BENT. THERE IS NO REPORT OF INJURY, HOWEVER, THERE IS A REPORT OF A 30-MINUTE DELAY WHILE WAITING FOR THE 2ND SET OF EQUIPMENT. THE DOCTOR PROCEEDED TO PERFORMED PERINEAL URETHROSTOMY (PU) TO CREATE A CHANNEL FOR BETTER ACCESS AND THEN CONTINUED WITH THE BACKUP WOLF SCOPE AND SHEATHS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23017140 ↗
  • MalfunctionDec 27, 2021Optical Problem; Optical Problem

    THE PURPOSE OF THIS SUBMISSION IS TO REPORT THE RESULTS OF THE DEVICE INVESTIGATION/EVALUATION: THE DEVICE WAS RECEIVED BY RICHARD WOLF MEDICAL INSTRUMENT AND EVALUATED USING VISUAL AND FUNCTIONAL MEANS. THE REPORTED CONDITION, UNABLE TO SEE CLEARLY THROUGH SCOPE, WAS NOT CONFIRMED. ACCORDING TO THE TECHNICIAN, "REPORTED PROBLEM NOT JUSTIFIED, SCOPE PASS ALL TESTS". PROBABLE ROOT CAUSE: NO DEVICE PROBLEM FOUND.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13088282 ↗

Other Resectoscope devices

All Resectoscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page