ERAGON BIPOLAR 83930031Recalls & reported problems
Richard Wolf GmbH · GRASP. FORCEPS INSERT Ø 5.5MM single fenestration with cross toothing, jaw length 21mm, for SL 320mm, double action, reusable · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04055207026526), received between Jun 29, 2020 and Nov 29, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component4
Patient problems
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 29, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
ON NOVEMBER 18, 2021 RICHARD WOLF (B)(4) WAS INFORMED BY THE DISTRIBUTOR. THE CUSTOMER HAS REPORTED THAT THE JAWS FROM THE INSERT 83930031 HAS BROKEN DURING OPERATION. THE BROKEN PART WAS FOUND ON THE GROUND FROM THE OR, IT WAS NOT IN THE BODY. NO INJURY FOR THE PATIENT. BUT THERE WAS A DELAY OF 30 MINUTES DUE TO THE RESEARCH OF THE BROKEN PIECE AND TO PERFORM RADIOSCOPY OF THE BODY. A SECOND ERAGON BIPOLAR DEVICE WAS TAKEN TO COMPLETE THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12887531 ↗MalfunctionAug 12, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE USER FACILITY, DURING THE OPERATION ON (B)(6) 2020, AGAIN A PART OF THE BRANCH OF A 83930031 GRIPPING PLIERS-INNER PART Ø 5,5MM BROKE OFF DURING AN OPERATION IN THE GYNAECOLOGICAL HOSPITAL ON THE ELBE. THE BROKEN OFF PART WAS RECOVERED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10397484 ↗MalfunctionJun 29, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE OPERATION A PART OF THE FORCEPS BROKE OFF AND REMAINED IN THE PATIENT. THE BROKEN PART COULD ONLY BE FOUND AND RECOVERED IN A 2ND OPERATION WITH X-RAY, FOR WHICH THE PATIENT HAD TO BE TRANSFERRED TO ANOTHER CLINIC. THE USER INFORMED US OF THE FOLLOWING: THE BRANCH OF THE BIPOLAR FORCEPS BROKE OFF WHEN A SECOND FORCEPS WAS GRASPED AND JUMPED IN THE ABDOMEN. WE SEARCHED FOR 40 MINUTES FOR THE BROKEN OFF PART AND DID NOT FIND IT. THE PATIENT WAS LAPAROSCOPED AGAIN THE SAME DAY AT THE (B)(6) UNDER X-RAY CONTROL AND THE BRANCH WAS FOUND. THE PATIENT WAS DISCHARGED ON (B)(6) 2020. CONDITION OF THE PATIENT AFTER THE APPLICATION IS VERY GOOD. THE GYNAECOLOGICAL CLINIC HAS REQUESTED THE BROKEN OFF PART FROM THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10209329 ↗MalfunctionJun 29, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
ON (B)(6) 2020 , A PART OF THE INSERT OF ERAGON BIPOLAR FORCEP (83930031) BROKE OFF DURING AN OPERATION IN THE GYNAECOLOGICAL HOSPITAL ON THE ELBE. THE BROKEN PART WAS RECOVERED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10209320 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗