Skip to content

S-LINE 8680204Recalls & reported problems

Richard Wolf GmbH · WORKING ELEMENT ACTIVE BIPO 0/12/30° open, compatible with telescopes Ø 4mm, bayonet-cap, reusable · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04055207042922), received between Apr 21, 2022 and Feb 19, 2026 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Arcing of Electrodes” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Arcing of Electrodes1
  • Explosion1
  • Material Separation1
  • Thermal Decomposition of Device1
  • Unintended Electrical Shock1
  • Use of Device Problem1

Patient problems

  • Air Embolism1
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • Laceration(s)1
  • No Clinical Signs, Symptoms or Conditions1
  • Perforation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 19, 2026Use of Device Problem; Explosion

    THE USER REPORTS THAT UPON ACTIVATION OF THE CUTTING CURRENT AND CONTACT WITH AN AIR BUBBLE, BLADDER EXPANSION OCCURS FOLLOWED BY BLADDER DETONATION. PLANNED PROCEDURE WAS CYSTOLITHOTRIPSY. DUE TO A LARGE AMOUNT OF INFECTION AND TISSUE, A RESECTOSCOPE SHARK BIPOLAR WAS INITIALLY USED TO REMOVE THIS TISSUE. THE PATIENT SUSTAINED A BLADDER PERFORATION IN THE AREA OF THE BLADDER VERTEX. ON THE FOLLOWING DAY, A LAPAROTOMY WAS PERFORMED. THE PATIENT'S CONDITION HAS BEEN STABILIZED AND THE DAMAGE WAS REPAIRED.

    Patient problems: Perforation

    MAUDE report 24394198 ↗
  • InjuryFeb 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING BIPOLAR PROSTATE RESECTION, THE PATIENT DEVELOPED PULMONARY EMBOLISM. THE OPERATION TIME WAS PROLONGED. HOWEVER, THE SCHEDULED PROCEDURE WAS COMPLETED. THE PATIENT WAS KEPT IN AN EMERGENCY/HOSPITAL AND WAS ABLE TO LEAVE THE HOSPITAL.

    Patient problems: Air Embolism

    MAUDE report 21403369 ↗
  • InjuryOct 25, 2023Material Separation; Material Separation

    THE USER FACILITY HAS INFORMED RICHARD WOLF GMBH OF AN ISSUE REGARDING A WORKING ELEMENT ACTIVE BIPO 0/12/30°, PART ID: 8680.204, LOT # 1476165. ACCORDING TO THE RECEIVED INFORMATION, THE DEVICE WAS BROKEN PROXIMALLY DURING A TRANSURETHRAL RESECTION OF THE PROSTATE (TURP), DURING WHICH A PART OF THE PROSTATE DISEASED TISSUE WAS RESECTED, FROM PROSTATE. IT HAPPENED DURING THE RESECTION OF THE TISSUE. THE PATIENT HAS MODERATE BLEEDING, WHICH IS IMPROVED BY TRACTION WITH A FOLLEY CATHETER. DUE TO THE BREAK OF THE INSTRUMENT, THE OPERATION TIME HAS BEEN PROLONGED BY ONE(1) HOUR. NOW THE PATIENT IS DOING WELL. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE SCHEDULED SURGERY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Laceration(s)

    MAUDE report 18004395 ↗
  • MalfunctionDec 19, 2022Arcing of Electrodes; Unintended Electrical Shock; Arcing of Electrodes; Unintended Electrical Shock

    RICHARD WOLF GMBH REFERENCE: (B)(4). RICHARD WOLF GMBH IN (B)(4) (MANUFACTURER) WAS INFORMED ABOUT THE FOLLOWING: DESCRIPTION OF INCIDENT/CLAIM ACCORDING TO THE CUSTOMER: "ON (B)(6) 2022 SCHEDULED INTERVENTION: ENDOSCOPIC RESECTION OF THE PROSTATE. ALL CONNECTIONS WERE CORRECT, THE RIGHT MODE/PROGRAM "BIPOLAR RTUP". THE SURGEON INDUCES THE RESECTOR, STARTS THE "SECTIONING" OF THE PROSTATE AND THE PATIENT ON JUMPS LIKE AN EXTERNAL SHOCK OR ELECTROSTIMULATION. ALERTING THE REGULATOR AND THE EXECUTIVE WHO CAME DIRECTLY. MOREOVER IT IS THE SECOND UNDESIRABLE EVENT IN THE SAME ROOM, ON THE SAME INTERVENTION THAT OCCURRED ON (B)(6) 2022. YESTERDAY THE INTERCOM WAS TRIGGERED SEVERAL TIMES JUST BEFORE THE ARC AND TODAY WE NOTICE THE GENERAL LIGHTING OF THE ROOM THAT FLASHES IN AN UNSYSTEMATIC WAY." THE USER INFORMED THAT THERE WERE NO NEGATIVE CONSEQUENCES FOR THE PATIENT, USER OR THIRD PARTIES. THE USER STATED TO REPORT IT TO ANSM. RWGMBH MDR AWARENESS DATE: (B)(6) 2022.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16009294 ↗
  • MalfunctionApr 21, 2022Thermal Decomposition of Device; Thermal Decomposition of Device

    THE PURPOSE OF THIS SUBMISSION IS TO REPORT THE RESULTS OF THE DEVICE INVESTIGATION.

    Patient problems: Insufficient Information

    MAUDE report 14166307 ↗

Other RESECTOSCOPE, WORKING ELEMENT devices

All RESECTOSCOPE, WORKING ELEMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.