S-LINE 8680205Recalls & reported problems
Richard Wolf GmbH · WORKING ELEMENT ACTIVE BIPO 0/12/30 closed, compatible with telescopes Ø 4mm, bayonet-cap, reusable · Class 2
All models of S-LINE (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 1
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04055207042939), received between May 17, 2018 and Apr 8, 2026 — classified as 1 death, 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent2
- Adverse Event Without Identified Device or Use Problem1
- Electrical /Electronic Property Problem1
- Temperature Problem1
Patient problems
- Insufficient Information2
- Burn(s)1
- Death1
- Electrocution1
- Full thickness (Third Degree) Burn1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryApr 8, 2026Material Twisted/Bent
THE CUSTOMER STATED THAT THE HANDLE ON THE WORKPIECE IS BENT, PREVENTING THE HF LOOP FROM BEING MOVED ALL THE WAY TO THE END. THE SURGERY COULD NOT BE COMPLETED. THE USER DID NOT PROVIDE ANY INFORMATION REGARDING THE PATIENT'S CONDITION.
Patient problems: Insufficient Information
MAUDE report 24821595 ↗InjuryApr 8, 2026Material Twisted/Bent
THE CUSTOMER STATED THAT THE HANDLE ON THE WORK PIECE IS BENT, PREVENTING THE HF LOOP FROM BEING MOVED ALL THE WAY TO THE END. THE SURGERY COULD NOT BE COMPLETED. THE USER DID NOT PROVIDE ANY INFORMATION REGARDING THE PATIENT'S CONDITION.
Patient problems: Insufficient Information
MAUDE report 24821580 ↗DeathSep 13, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON SEPTEMBER 11, 2019, RICHARD WOLF (B)(4) RECEIVED THE FOLLOWING INFORMATION: A PATIENT DEATH WAS REPORTED BY THE HOSPITAL (B)(6) IN (B)(6) WHERE AN S-LINE RESECTOSCOPE WORKING ELEMENT AND ACCESSORIES WERE BEING USED DURING A PROCEDURE, TRANSURETHRAL RESECTION OF THE BLADDER. THE BLADDER OF THE PATIENT WAS REPORTED TO HAVE BEEN FILLED WITH SODIUM CHLORIDE (NACL) SOLUTION. DURING THE RESECTION (NOT IN THE IGNITION PHASE), THE BLADDER BURST AND CAUSED A PELVIC ARTERY TO TEAR. THIS HAPPENED ALMOST AT THE END OF THE SURGICAL PROCEDURE. AN UNSUCCESSFUL EMERGENCY OPERATION WAS CARRIED OUT FOR OVER APPROXIMATELY THREE(3) HOURS, WITH FOUR (4) DOCTORS PRESENT. THE PATIENT EXPIRED DURING THE EMERGENCY OPERATION. THE PATIENT WAS (B)(6) YEARS OLD AND WAS IN NORMAL, GOOD STATE OF HEALTH. THERE WAS NO INFORMATION AVAILABLE REGARDING THE QUANTITY OF NACL USED OR ABOUT THE FLUSHING PRESSURE. THE POSITION OF THE NACL BOTTLE WAS APPROXIMATELY 1M ABOVE THE OPERATING TABLE WITHOUT A PUMP. IT IS CURRENTLY NOT KNOWN HOW LONG THE RESECTION TOOK.
Patient problems: Death
MAUDE report 9050559 ↗InjuryFeb 22, 2019Temperature Problem; Temperature Problem
ON NOVEMBER 07, 2018, RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) RECEIVED A REPORT FROM RICHARD WOLF (B)(4) INVOLVING THE FOLLOWING DEVICES: 815.052 LOT 523/573 (MDR 9611102-2018-00015), 815.032 LOT 006/086 (MDR 9611102-2018-00016), 8674.205 LOT 1259038 (MDR 9611102-2018-00017), 8680.205 LOT 1255367 (THIS MDR), 8654.204 LOT 1202777 (MDR 9611102-2018-00019). IT WAS REPORTED THAT THE PATIENT SUFFERED THIRD DEGREE BURN ON THE RIGHT FINGER DUE TO HF APPLICATION. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE INSPECTION SHOWED THAT THE CAUSE OF THE FAULT DESCRIBED BY THE USER CANNOT BE TRACED BACK TO THE WORKING ELEMENTS OR HF CONNECTION CABLE MONO. PROBABLE ROOT CAUSE WAS DETERMINED TO BE USE ERROR. NO DEFECTS WERE FOUND IN THE MEDICAL DEVICES. IN CONCLUSION, THE COMBUSTION WAS NOT CAUSED BY A FAULTY HF DEVICE OR ACCESSORY.
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 8363751 ↗MalfunctionMay 17, 2018Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
DURING BIPOLAR CUTTING AND COAGULATION DURING A BIPOLAR TUR, THE ANESTHESIOLOGIC SISTER JUST BELOW THE PUNCTURE SITE OF THE VENOUS CATHETER ON THE PATIENT'S LEFT FOREARM BURNED HER FINGERTIP OF THE RING FINGER TO THE RIGHT. ABOUT THIS CATHETER WAS ADMINISTERED A SALINE INFUSION. THE JUMPERS DURING THE PROCEDURE FOUND ONLY A CURRENT DELIVERY AT THE DESCRIBED LOCATION OF THE PATIENT. LEFT ARM OF THE PATIENT, PARALLEL TO RESECTION OF THE LEFT PROSTATE VALVE. IN OTHER PARTS OF THE PATIENT'S BODY NO CURRENT DELIVERY WAS DETECTED. TYPE: 8680.205 WORKING ELEMENT ACTIVE BIPO 0/12/30, MANUFACTURED DATE: 15 DEC 2015, LOT SIZE: (B)(4) UNITS, PURCHASE DATE: 19 DEC 2016.
Patient problems: Burn(s); Electrocution
MAUDE report 7522540 ↗
FDA clearance and classification
The FDA regulates product code FDC (resectoscope, working element) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the S-LINE 8680205
Has the S-LINE 8680205 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04055207042939), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the S-LINE 8680205?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between May 17, 2018 and Apr 8, 2026. In those reports, the most common specific device problems are “Material Twisted/Bent” (2), “Electrical /Electronic Property Problem” (1) and “Temperature Problem” (1); the most common patient problems are “Burn(s)” (1), “Death” (1) and “Electrocution” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the S-LINE 8680205?
1 report is classified as a death, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the S-LINE 8680205, and is it still sold?
The FDA's device database lists Richard Wolf GmbH as the company responsible for it (brand S-LINE, model 8680205). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other RESECTOSCOPE, WORKING ELEMENT devices
- S-LINE 8680224Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT7 reports
- E-LINE 86553841Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT7 reports
- S-LINE 8680204Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT6 reports
- S-LINE 8680225Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT5 reports
- NA 8654282Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT3 reports
- PRINCESS 8653018Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT2 reports
- MIDI-SIZE 86770155Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT2 reports
- SHARK 8674225Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page