S-LINE 8680224Recalls & reported problems
Richard Wolf GmbH · WORKING ELEMENT PASSIVE BIPO 0/12/30° open, compatible with telescopes Ø 4mm, bayonet-cap, reusable · Class 2
All models of S-LINE (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04055207042946), received between Jun 28, 2017 and Jun 8, 2021 — classified as 3 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Grounding Malfunction” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Grounding Malfunction3
- Unintended Power Up3
- Bent2
- Adverse Event Without Identified Device or Use Problem1
- Unintended Electrical Shock1
Patient problems
- Lead(s), Burn(s) From3
- No Consequences Or Impact To Patient3
- Shock2
- Burn, Thermal1
- Electric Shock1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 8, 2021Unintended Electrical Shock; Unintended Electrical Shock
THE PURPOSE OF THIS REPORT IS TO REPORT THE DEVICE EVALUATION/INVESTIGATION RESULTS. PLEASE SEE MANUFACTURERS NARRATIVE.
Patient problems: Electric Shock
MAUDE report 11957808 ↗InjuryJan 10, 2019Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2018, RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) WAS SERVED WITH A PATIENT LAWSUIT INVOLVING A "BUTTON TYPE BIPOLAR PROSTATE RESECTOR AND/OR ELECTROSURGICAL GENERATOR". ON (B)(6) 2016, THE PATIENT UNDERWENT A PROCEDURE DESCRIBED AS "CYSTOSCOPY WITH URETHRAL DILATION, TRANSURETHRAL RESECTION/ABLATION OF PROSTATE WITH BIPOLAR". DURING SAID PROCEDURE, THE PATIENT WAS UNDER GENERAL ANESTHESIA AND SUFFERED THERMAL INJURY TO THEIR BLADDER AND URETHRA. AT THIS TIME, THE INVOLVEMENT OF RICHARD WOLF PRODUCTS HAS NOT BEEN CONFIRMED. HOWEVER, THE FOLLOWING ARE CONSIDERED SUSPECT PRODUCTS AND MDRS WILL BE SUBMITTED FOR EACH UNTIL THE USER FACILITY IS ABLE TO CONFIRM THAT NO RICHARD WOLF PRODUCTS WERE INVOLVED: 8680.225 WORKING ELEMENT PASSIVE BIPO 0/12/30 (MDR 1418479-2018-00051), 8680.224 WORKING ELEMENT PASSIVE BIPO 0/12/30 (MDR), 46300223 BIVAP ELECTRODE BIPO 22FR 12/30 (MDR 1418479-2018-00053), 46300243 BIVAP ELECTRODE BIPO 24-26FR 12/30 (1418479-2018-00054).
Patient problems: Burn, Thermal
MAUDE report 8237997 ↗InjuryAug 22, 2017Bent
FOLLOW UP REPORT #1. RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) RECEIVED ADDITIONAL INFORMATION FROM FACILITY ON 20SEP2017. THE FOLLOWING BOXES HAVE BEEN UPDATED/COMPLETED: PATIENT INFORMATION. DATE DESCRIPTION. SUSPECT DEVICE -(NOT CHANGED, ON MISSING DATA REPORT). INITIATOR ADDRESS. IMPORTER INFORMATION.
Patient problems: Shock
MAUDE report 6810262 ↗InjuryAug 22, 2017Bent
(B)(4) SALES REPRESENTATIVES WAS NOTIFIED BY FACILITY THAT DURING A BIPOLAR TURBT PROCEDURE TWO NURSES WERE SHOCKED WHEN THEY GRABBED THE BIZ MONITOR DEVICE WHILE REMOVING IT FROM THE PATIENT. NO INJURY TO PATIENT WAS REPORTED. TWO SEPARATE MDRS WILL BE SUBMITTED, ONE FOR EACH NURSE: MDR 1418479-2017-00018, NURSE #1, MDR 1418479-2017-00019, NURSE #2. USER FACILITY WAS CONTACTED IN AN EFFORT TO GATHER ADDITIONAL/MISSING MEDWATCH INFORMATION, NO RESPONSE AS OF 08/22/17. MANUFACTURER / DEVICE RETURNED FOR INVESTIGATION: (B)(4) WORKING ELEMENT (ID #8680.224 LT #1324186), (B)(4) ELECTRODE (ID #4622.1313 LT #51007450), (B)(4) BIPOLAR CABLE (ID #8108.232 LT #07/14), (B)(4) UNIPOLAR CABLE (ID #815.033 LOT #128/180). ADDITIONAL DEVICES USED DURING PROCEDURE: (B)(4) GENERATOR (ID #UNKNOWN), BIZ MONITOR (ID #UNKNOWN). MANUFACTURED DATE: 12-OCT-2016, PURCHASE DATE: (B)(6) 2017, SERVICE DATE: N/A. (B)(4) RECEIVED ACTUAL DEVICE FROM USER FACILITY ON 03-AUG-2017. A VISUAL INSPECTION OF DEVICE FOUND SHAFT WAS BENT. NO PROBLEMS FOUND WHEN FUNCTIONAL TESTING WAS PERFORMED ON DEVICE. (B)(4) COULD NOT REPLICATE EVENT WITH THE DEVICES RETURNED FOR INVESTIGATION. NONE OF THE RETURNED DEVICES CONTAINED D
Patient problems: Shock
MAUDE report 6810257 ↗MalfunctionJun 28, 2017Unintended Power Up; Grounding Malfunction; Unintended Power Up; Grounding Malfunction
RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) WAS NOTIFIED BY USER FACILITY THAT DURING A PROCEDURE THE ANESTHESIOLOGIST TOUCHED AN EKG ELECTRODE AND WAS BURNED ON THE FINGER. NO INJURY TO THE PATIENT WAS REPORTED. OCCURED APPROXIMATELY TWO YEARS AGO. THIS EVENT HAS OCCURRED ON THREE SEPARATE OCCASIONS: 1) (B)(6) 2017 - MDR 1418479-2017-00014. 2) (B)(6) 2017 - MDR 1418479-2017-00015. 3) (B)(6) 2015 - MDR 1418479-2017-00016. USER FACILITY CONTACTED IN AN EFFORT TO GATHER ADDITIONAL/MISSING INFORMATION, NO RESPONSE AS OF 06/28/17. LIST OF RWMIC ITEMS USED DURING PROCEDURE: 1) WORKING ELEMENT (ID #8680.224), 2) ELECTRODE (ID #4622.1333). . LIST OF RWMIC (B)(6) ITEMS USED DURING PROCEDURE: 1) INNER SHEATH (ID #8655.3841), 2) OUTER SHEATH (ID #8655.374), 3) CONNECTOR (ID #8654.3742). LIST OF OTHER KNOWN ITEMS USED DURING PROCEDURE: 1) ERBE VIO GENERATOR (ID #UNKNOWN), 2) ERBE BIPOLAR CABLE (ID #8108.232), 3) EKG MONITOR (ID #UNKNOWN). SUSPECT DEVICE AT THIS TIME IS THE WORKING ELEMENT ID #8680.224.
Patient problems: Lead(s), Burn(s) From; No Consequences Or Impact To Patient
MAUDE report 6675287 ↗MalfunctionJun 28, 2017Unintended Power Up; Grounding Malfunction; Unintended Power Up; Grounding Malfunction
RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) WAS NOTIFIED BY USER FACILITY THAT DURING A PROCEDURE THE ANESTHESIOLOGIST TOUCHED AN EKG ELECTRODE AND WAS BURNED ON THE FINGER. NO INJURY TO THE PATIENT WAS REPORTED. THIS EVENT HAS OCCURRED ON THREE SEPARATE OCCASIONS: ON (B)(6) 2017 - MDR 1418479-2017-00014, ON (B)(6) 2017 - MDR 1418479-2017-00015, ON (B)(6) 2015 - MDR 1418479-2017-00016. USER FACILITY CONTACTED IN AN EFFORT TO GATHER ADDITIONAL/MISSING INFORMATION, NO RESPONSE AS OF 06/23/17. LIST OF RWMIC ITEMS USED DURING PROCEDURE: WORKING ELEMENT (ID #8680.224), ELECTRODE (ID #4622.1333). LIST OF RWMIC (B)(4) ITEMS USED DURING PROCEDURE: INNER SHEATH (ID #8655.3841), OUTER SHEATH (ID #8655.374), CONNECTOR (ID #8654.3742). LIST OF OTHER KNOWN ITEMS USED DURING PROCEDURE: ERBE VIO GENERATOR (ID #UNKNOWN), ERBE BIPOLAR CABLE (ID #8108.232), EKG MONITOR (ID #UNKNOWN). SUSPECT DEVICE AT THIS TIME IS THE WORKING ELEMENT ID #8680.224.
Patient problems: Lead(s), Burn(s) From; No Consequences Or Impact To Patient
MAUDE report 6675285 ↗MalfunctionJun 28, 2017Unintended Power Up; Grounding Malfunction; Unintended Power Up; Grounding Malfunction
RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) WAS NOTIFIED BY USER FACILITY THAT DURING A PROCEDURE THE ANESTHESIOLOGIST TOUCHED AN EKG ELECTRODE AND WAS BURNED ON THE FINGER. NO INJURY TO THE PATIENT WAS REPORTED. THIS EVENT HAS OCCURRED ON THREE SEPARATE OCCASIONS: (B)(6). USER FACILITY CONTACTED IN AN EFFORT TO GATHER ADDITIONAL/MISSING INFORMATION, NO RESPONSE AS OF 06/23/2017. LIST OF RWMIC ITEMS USED DURING PROCEDURE: WORKING ELEMENT (ID #8680.224). ELECTRODE (ID #4622.1333). LIST OF RWMIC (B)(4) ITEMS USED DURING PROCEDURE: INNER SHEATH (ID #8655.3841). OUTER SHEATH (ID #8655.374). CONNECTOR (ID #8654.3742). LIST OF OTHER KNOWN ITEMS USED DURING PROCEDURE: ERBE VIO GENERATOR (ID #UNKNOWN). ERBE BIPOLAR CABLE (ID #8108.232). EKG MONITOR (ID #UNKNOWN). SUSPECT DEVICE AT THIS TIME IS THE WORKING ELEMENT ID #8680.224.
Patient problems: Lead(s), Burn(s) From; No Consequences Or Impact To Patient
MAUDE report 6675277 ↗
FDA clearance and classification
The FDA regulates product code FDC (resectoscope, working element) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the S-LINE 8680224
Has the S-LINE 8680224 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04055207042946), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the S-LINE 8680224?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Jun 28, 2017 and Jun 8, 2021. In those reports, the most common specific device problems are “Grounding Malfunction” (3), “Unintended Power Up” (3) and “Bent” (2); the most common patient problems are “Lead(s), Burn(s) From” (3), “Shock” (2) and “Burn, Thermal” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the S-LINE 8680224?
None of the reports are classified as deaths, 3 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the S-LINE 8680224, and is it still sold?
The FDA's device database lists Richard Wolf GmbH as the company responsible for it (brand S-LINE, model 8680224). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other RESECTOSCOPE, WORKING ELEMENT devices
- E-LINE 86553841Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT7 reports
- S-LINE 8680204Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT6 reports
- S-LINE 8680225Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT5 reports
- S-LINE 8680205Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT1 death5 reports
- NA 8654282Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT3 reports
- PRINCESS 8653018Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT2 reports
- MIDI-SIZE 86770155Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT2 reports
- SHARK 8674225Richard Wolf GmbH · RESECTOSCOPE, WORKING ELEMENT2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page