Skip to content

ERAGON MODUBLADES 4393040Recalls & reported problems

Richard Wolf GmbH · METZENBAUM SCISSORS ATTACHMENT MONO PACK=10 PCS, for ERAGONmodublades insert and sheath, ERAGONmodular and ERAGONaxial handles, sterile, for single use · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04055207048658), received between May 22, 2025 and Jul 22, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 22, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT AFTER COMPLETION OF THE LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, PARTS OF THE PLASTIC INSULATION WERE DISCOVERED MISSING FROM THE DISPOSABLE SCISSORS. A NEW LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE WAS PERFORMED AND THE MISSING PARTS WERE FOUND AND REMOVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 22564380 ↗
  • InjuryMay 22, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC PROCEDURE, THE INSULATION ON THE LAPAROSCOPIC SCISSORS BROKE AND A PIECE OF PLASTIC INSULATION CAME LOOSE IN THE PATIENT'S ABDOMEN. THE LOOSE PIECE WAS FOUND AND REMOVED.

    Patient problems: Insufficient Information

    MAUDE report 22070385 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page