ERAGON MODULAR 8393249Recalls & reported problems
Richard Wolf GmbH · INTESTINE GRASP. FORCEPS INSERT Ø 5MM atraumatic, single fenestration with cross toothing, jaw length 38mm, for SL 330mm, double action, reusable · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04055207055731), received between Aug 28, 2018 and Aug 11, 2026 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation3
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Detachment of Device or Device Component1
- Insufficient Device Problem Information1
Patient problems
- Insufficient Information2
- Bowel Perforation1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- Perforation1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 11, 2026Material Fragmentation
It was reported that during laparoscopic surgery while moving bowel, a small metal piece of the grasper fell off. An X-Ray was taken and the missing piece was visualized. However, it was unable to be removed. The Port site extended, and the small piece of the grasper was located. The Bowel resection completed.
Patient problems: Insufficient Information
MAUDE report 50201456 ↗InjuryAug 11, 2026Material Fragmentation
It was reported that during laparoscopic surgery, the grasper broke inside of patient while grasping a piece of bowel to hold and retract it. Two pieces were located and removed at that time. A third piece was determined to be missing and unable to be located.An X-Ray was taken and the Third piece was visualized. However, it was unable to be removed via the laparoscopic surgery. Therefore, the procedure was converted to an open resection, and the third piece was removed. The bowel resection was completed.
Patient problems: Insufficient Information
MAUDE report 50201451 ↗MalfunctionDec 27, 2022Material Fragmentation; Material Fragmentation
RICHARD WOLF GMBH REFERENCE: (B)(4). RICHARD WOLF GMBH IN KNITTLINGEN /MANUFACTURER WAS INFORMED ABOUT THE FOLLOWING: DESCRIPTION OF INCIDENT/CLAIM: ITEM 8393249 LOT 4500250452, INSTRUMENT WAS BEING USED FOR A LAPAROSCOPIC BOWEL CASE. ONCE THE CASE HAD FINISHED THE SCRUB NURSE REALIZED THAT AN INSTRUMENT WAS DAMAGED AND THAT A PIECE WAS MISSING. AFTER AN XRAY OF THE PATIENT THE MISSING PIECE WAS LOCATED IN THE PATIENT WHO WAS SUBSEQUENTLY RE-OPENED AND THE PIECE WAS REMOVED. PATIENT HAD TO RETURN TO THEATRE AND BE ANESTHETIZED FOR A 2ND TIME TO REMOVE BROKEN ITEM. RWGMBH MDR AWARENESS DATE: (B)(6) 2022
Patient problems: Foreign Body In Patient
MAUDE report 16055635 ↗MalfunctionSep 14, 2020Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
RWMIC EVALUATION 09/24/2020: FINDINGS, INSERT 8393.249: THE WELDING SEAMS AROUND ALL 4 RIVETS ARE CRACKED, BUT THE RIVETS ARE PRESENT. THE PULLROD CONNECTOR IS BROKEN JUST PROXIMAL OF THE SEGMENT CONNECTIONS, BUT ALL PARTS ARE PRESENT. SHEATH 8393.933 / 1341886: DISTAL END OF SHEATH IS DEFORMED AND BENT. THE CONDITION OF THESE 2 COMPONENTS OF THE MODULAR SYSTEM SHOWS EXTREME MIS-USE OF THE BOWEL GRASPER. THE JOINTS WERE OVERLOADED WITH GRASPING FORCE, AND THE WHOLE DISTAL SECTION OF THE INSTRUMENT HAS BEEN BENT. THE CONDITION REPORTED BY THE USER FACILITY IS CONFIRMED. PRODUCT DISPOSITION IS SCRAP. POSSIBLE ROOT CAUSE IS USER ERROR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10532328 ↗InjuryAug 28, 2018Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Insufficient Device Problem Information
RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (HEREINAFTER REFERRED TO AS RWMIC) WAS MADE AWARE OF AN EVENT FROM RICHARD WOLF (B)(4) (HEREINAFTER REFERRED TO AS (B)(4)) ON 8/17/2018. (B)(4) WAS NOTIFIED ON 7/31/2018 BY DISTRIBUTOR IN (B)(6) OF AN EVENT THAT OCCURRED AT THE HOSPITAL ON (B)(6) 2018 DURING A SMALL BOWEL OCCLUSION SURGERY. DURING PROCEDURE 7 PERFORATIONS WERE NOTED AND DETERMINED TO BE LIKELY CAUSED BY MANIPULATION WITH THE FORCEPS. THE BOWEL WAS UNCHARACTERISTIC DURING THE PROCEDURE AND THE MANIPULATION WAS NOT EXCESSIVE. THE DOCTOR HAD TO SET SEVERAL CLAMPS WHICH HAD TO BE REMOVED LATER. THE PATIENT REQUIRED TO STAY IN HOSPITAL LONGER THAN EXPECTED.
Patient problems: Bowel Perforation; Perforation
MAUDE report 7824372 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.