NA 2232644Recalls & reported problems
Richard Wolf GmbH · TEM PUMP US version, U: 100-230VAC, 50/60Hz, Dim. (wxhxd): 330x100x390mm · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04055207056462), received between Sep 11, 2017 and Jul 24, 2020 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Display or Visual Feedback Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Display or Visual Feedback Problem1
- Inflation Problem1
- Power Problem1
- Pumping Stopped1
- Unexpected Shutdown1
Patient problems
- No Consequences Or Impact To Patient3
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 24, 2020Pumping Stopped; Pumping Stopped
IT WAS REPORTED: AFTER ABOUT 30 MINUTES DURING THE FIRST CASE THE PUMP WOULD QUIT WORKING AND THE UNIT WOULD BEEP. I HAD TO OPEN THE TRAP DOOR AND CLOSE IT TO MAKE THE BEEPING GO AWAY. THEN I WOULD PRESS THE TEM PUMP BUTTON AGAIN TO GET IT TO WORK. THIS WAS A TEMPORARY FIX SINCE IT KEPT FAULTING MORE OFTEN. IT BECAME SO BAD THAT MY TRICK DIDN'T WORK ANYMORE AND I HAD TO SET UP THE LOANER. WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTING ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? NO. WAS THERE ANY INJURY OR ILLNESS TO PATIENT OR OTHER PERSONNEL DUE TO ISSUE? NO. DID THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED? YES. DID THE DELAY PUT THE PATIENT AT RISK? NO. WAS THERE A SIMILAR BACK-UP DEVICE AVAILABLE FOR USE? YES. WAS THE SCHEDULED PROCEDURE COMPLETED? YES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10321025 ↗MalfunctionFeb 7, 2020Display or Visual Feedback Problem; Display or Visual Feedback Problem
ON (B)(6) 2020, RICHARD WOLF MEDICAL INSTRUMENTS CORP. (RWMIC) RECEIVED THE FOLLOWING INFORMATION: FIRST TIME THE CUSTOMER HAS USED BOTH ITEMS SINCE THEY CAME IN FOR REPAIR IN (B)(6) 2019. DURING THE PROCEDURE, PUMP STOPPED WORKING AND ERROR CODE E37 SHOWED UP ON SCREEN AND THEN THE RED TRIANGLE STARTED TO FLASH. PROCEDURE WAS STILL BEING DONE WHEN CUSTOMER CALLED THIS IN, BUT IT WAS LATER LEARNED THAT THE PROCEDURE WAS NOT COMPLETED DUE TO THE ITEM GOING DOWN. A BACK UP DEVICE WAS NOT AVAILABLE FOR USE. THERE WAS NO INJURY TO PATIENT REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9682543 ↗MalfunctionMar 11, 2019Power Problem; Power Problem
ON (B)(6) 2018, (B)(4) RECEIVED MEDWATCH REPORT # (B)(4). THE FOLLOWING WAS REPORTED: "PATIENT SCHEDULED FOR A TRANSANAL ENDOSCOPIC MICROSURGICAL EXCISION OF THE RECTAL TUMOR USING THE (B)(4) TEM MACHINE. THE PATIENT WAS INTUBATED AND A PROCTOSIGMOIDOSCOPY WAS PERFORMED BY THE SURGEON AND THEN THE PATIENT WAS POSITIONED AND PREPPED FOR SURGERY. WHEN IT CAME TIME TO START THE MACHINE ONLY THE TOP UNIT WAS LIGHTING UP AND WORKING (INSUFFLATOR). THE BOTTOM UNIT DID NOT TURN ON. OTHER STAFF CAME INTO THE ROOM TO TROUBLESHOOTING THE PROBLEM WITH NO LUCK GETTING THE BOTTOM UNIT TO TURN ON. THE (B)(4) REP, AND TEAM LEADER WAS CALLED TO HELP TROUBLESHOOTING. AT THIS TIME THE SURGEON WANTED TO ABORT THE CASE." THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE SALES REPRESENTATIVE WENT TO THE USER FACILITY TO INVESTIGATE THE REPORTED ISSUE AND FOUND THAT THE CAUSE OF THE PROBLEM WAS A DEFECTIVE OUTLET. THE UNIT WAS PLUGGED INTO A DIFFERENT OUTLET AND BEGAN WORKING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8407944 ↗MalfunctionNov 21, 2018Unexpected Shutdown; Unexpected Shutdown
ON (B)(6) 2018, THE USER FACILITY REPORTED THE FOLLOWING TO RICHARD WOLF MEDICAL INSTRUMENTS (RWMIC): DURING THE PROCEDURE THE EQUIPMENT JUST SHUT OFF AND THE DOCTOR HAD TO TURN THE SURGERY INTO AN OPEN SURGERY. WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED DURING A PROCEDURE WHEN THE ISSUE OCCURRED? YES. SPECIFICALLY, WAS THE DEVICE BEING USED ON A PATIENT WHEN THE ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO PATIENT OR OTHER PERSONNEL DUE TO ISSUE? NO. DID THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED THAT PUT THE PATIENT AT RISK? YES. WAS THERE A SIMILAR BACK-UP DEVICE AVAILABLE FOR USE? NO. WAS THE SCHEDULED PROCEDURE COMPLETED? YES. HOW WAS THE PATIENT ANESTHETIZED? GENERAL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8094291 ↗InjurySep 11, 2017Inflation Problem
FOLLOW UP #2 - MANUFACTURER INVESTIGATION RESULTS.
Patient problems: No Information
MAUDE report 6855237 ↗
FDA clearance and classification
The FDA regulates product code HIF (insufflator, laparoscopic) under 21 CFR 884.1730 as a Class II device. 21 CFR 884.1730 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the NA 2232644
Has the NA 2232644 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04055207056462), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 2232644?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Sep 11, 2017 and Jul 24, 2020. In those reports, the most common specific device problems are “Display or Visual Feedback Problem” (1), “Inflation Problem” (1) and “Power Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 2232644?
None of the reports are classified as deaths, 1 as an injury and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the NA 2232644, and is it still sold?
The FDA's device database lists Richard Wolf GmbH as the company responsible for it (brand NA, model 2232644). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other INSUFFLATOR, LAPAROSCOPIC devices
- HIGH FLOW INSUFFLATION UNIT UHI-3Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC1 recall162 reports
- HIGH FLOW INSUFFLATION UNIT UHI-3Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC1 recall48 reports
- AIRSEAL AS-iFS1Conmed Corporation · Insufflator, laparoscopic4 deaths91 reports
- Pneumosure 0620040610Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 death79 reports
- Pneumosure 0620040690Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 recall22 reports
- Pneumosure 0620040660Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 recall18 reports
- CYLINDER HOSE (PIN) FOR UHI-3 MAJ-1080Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC58 reports
- Pneumosure 0620040680Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page