Skip to content

PIRANHA 49700103Recalls & reported problems

Richard Wolf GmbH · ROTATIONS-MORCELLATOR Ø 4.8MM WL 385MM for MORCESCOPE Storz, for morcellation following laser prostate enucleation, sterile · Class 2

All models of PIRANHA (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04055207063415), received between Sep 9, 2025 and Sep 9, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Suction Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Suction Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 9, 2025Suction Failure; Suction Failure

    IT WAS REPORTED THAT DURING THE HOLEP SURGERY, FOUR (4) MORCELLATOR BLADES FAILED TO ROTATE AND THEREFORE, UNABLE TO PROVIDE SUCTION. THE FIRST BLADE WORKED FOR 10 MINUTES THEN LOCKED UP AND THE BLADES AFTER THAT LOCKED UP AS WELL UPON TESTING THEM BEFORE INSERTING THEM INTO THE PATIENT. THE USER TRIED TO REPLACED THE TUBING AND THE CONTAINER, HOWEVER, FAILED TO GET THE BLADES TO FUNCTION. DUE TO THE REPORTED ISSUE, THE SCHEDULED PROCEDURE WAS CANCELLED AS WELL AS THE SUCCEEDING CASES. THERE WERE NO REPORTS OF INJURIES TO THE PATIENT OR THIRD PARTY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23006566 ↗
  • InjurySep 9, 2025Suction Failure; Suction Failure

    IT WAS REPORTED THAT DURING THE HOLEP SURGERY, FOUR (4) MORCELLATOR BLADES FAILED TO ROTATE AND THEREFORE, UNABLE TO PROVIDE SUCTION. THE FIRST BLADE WORKED FOR 10 MINUTES THEN LOCKED UP AND THE BLADES AFTER THAT LOCKED UP AS WELL UPON TESTING THEM BEFORE INSERTING THEM INTO THE PATIENT. THE USER TRIED TO REPLACE THE TUBING AND THE CONTAINER, HOWEVER, FAILED TO GET THE BLADES TO FUNCTION. DUE TO THE REPORTED ISSUE, THE SCHEDULED PROCEDURE WAS CANCELLED AS WELL AS THE SUCCEEDING CASES. THERE WERE NO REPORTS OF INJURIES TO THE PATIENT OR THIRD PARTY.

    Patient problems: No Clinical Signs, Symptoms or Conditions